Image‐based modeling of blood flow in cerebral aneurysms treated with intrasaccular flow diverting devices. (11th April 2019)
- Record Type:
- Journal Article
- Title:
- Image‐based modeling of blood flow in cerebral aneurysms treated with intrasaccular flow diverting devices. (11th April 2019)
- Main Title:
- Image‐based modeling of blood flow in cerebral aneurysms treated with intrasaccular flow diverting devices
- Authors:
- Mut, Fernando
Chung, Bong Jae
Chudyk, Jorge
Lylyk, Pedro
Kadirvel, Ramanathan
Kallmes, David F.
Cebral, Juan R. - Abstract:
- Abstract: Modeling the flow dynamics in cerebral aneurysms after the implantation of intrasaccular devices is important for understanding the relationship between flow conditions created immediately posttreatment and the subsequent outcomes. This information, ideally available a priori based on computational modeling prior to implantation, is valuable to identify which aneurysms will occlude immediately and which aneurysms will likely remain patent and would benefit from a different procedure or device. In this report, a methodology for modeling the hemodynamics in intracranial aneurysms treated with intrasaccular flow diverting devices is described. This approach combines an image‐guided, virtual device deployment within patient‐specific vascular models with an immersed boundary method on adaptive unstructured grids. A partial mesh refinement strategy that reduces the number of mesh elements near the aneurysm dome where the flow conditions are largely stagnant was compared with the full refinement strategy that refines the mesh everywhere around the device wires. The results indicate that using the partial mesh refinement approach is adequate for analyzing the posttreatment hemodynamics, at a reduced computational cost. The results obtained on a series of four cerebral aneurysms treated with different intrasaccular devices were in good qualitative agreement with angiographic observations. Promising results were obtained relating posttreatment flow conditions and outcomes ofAbstract: Modeling the flow dynamics in cerebral aneurysms after the implantation of intrasaccular devices is important for understanding the relationship between flow conditions created immediately posttreatment and the subsequent outcomes. This information, ideally available a priori based on computational modeling prior to implantation, is valuable to identify which aneurysms will occlude immediately and which aneurysms will likely remain patent and would benefit from a different procedure or device. In this report, a methodology for modeling the hemodynamics in intracranial aneurysms treated with intrasaccular flow diverting devices is described. This approach combines an image‐guided, virtual device deployment within patient‐specific vascular models with an immersed boundary method on adaptive unstructured grids. A partial mesh refinement strategy that reduces the number of mesh elements near the aneurysm dome where the flow conditions are largely stagnant was compared with the full refinement strategy that refines the mesh everywhere around the device wires. The results indicate that using the partial mesh refinement approach is adequate for analyzing the posttreatment hemodynamics, at a reduced computational cost. The results obtained on a series of four cerebral aneurysms treated with different intrasaccular devices were in good qualitative agreement with angiographic observations. Promising results were obtained relating posttreatment flow conditions and outcomes of treatments with intrasaccular devices, which need to be confirmed on larger series. Abstract : This paper describes a methodology for modeling the hemodynamics in intracranial aneurysms treated with intrasaccular flow diverting devices. This approach combines an image‐guided, virtual device deployment within patient‐specific vascular models with an immersed boundary method on adaptive unstructured grids. The results obtained on a series of four cerebral aneurysms treated with different intrasaccular devices were in good qualitative agreement with angiographic observations. … (more)
- Is Part Of:
- International journal for numerical methods in biomedical engineering. Volume 35:Number 6(2019)
- Journal:
- International journal for numerical methods in biomedical engineering
- Issue:
- Volume 35:Number 6(2019)
- Issue Display:
- Volume 35, Issue 6 (2019)
- Year:
- 2019
- Volume:
- 35
- Issue:
- 6
- Issue Sort Value:
- 2019-0035-0006-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2019-04-11
- Subjects:
- cerebral aneurysms -- computational hemodynamics -- flow diversion -- intrasaccular flow diverter
Biomedical engineering -- Periodicals
Imaging systems in medicine -- Periodicals
Numerical analysis -- Periodicals
Engineering mathematics -- Periodicals
610.28 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)2040-7947 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/cnm.3202 ↗
- Languages:
- English
- ISSNs:
- 2040-7939
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4542.403550
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 10863.xml