Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation. (9th May 2019)
- Record Type:
- Journal Article
- Title:
- Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation. (9th May 2019)
- Main Title:
- Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation
- Authors:
- Rateb, Mahmoud
Abdel-Radi, Mahmoud
Eldaly, Zeiad
Elmohamady, Mohamed Nagy
Noor El Din, Asaad - Other Names:
- Mazzotta Cosimo Academic Editor.
- Abstract:
- Abstract : Purpose . To evaluate the different IOP readings by Goldmann applanation tonometer (GAT), ICare rebound tonometer, and Tono-Pen in keratoconus patients after MyoRing implantation. To assess the influence of central corneal thickness (CCT) and thinnest corneal location (TCL) on IOP measurements by different tonometers. Setting . Prospective observational study was conducted in two private centers in Egypt from February 2015 to November 2016. Methods . Seventeen eyes of 10 patients suffering from keratoconus and who underwent MyoRing implantation were recruited. All subjects underwent GAT, ICare, and Tono-Pen IOP measurements in random order. Central corneal thickness and thinnest corneal location were assessed by Pentacam. Difference in mean in IOP readings was assessed by T -test. Correlation between each pair of devices was evaluated by Pearson correlation coefficient. The Bland–Altman analysis was used to assess intertonometer agreement. Results . Seventeen eyes (10 patients) were evaluated. The mean IOP reading was 13.9 ± 3.68, 12.41 ± 2.87, and 14.29 ± 1.31 mmHg in GAT, ICare, and Tono-Pen group, respectively. There was a significant difference between IOP readings by GAT/ICare and Tono-Pen/ICare (p value: 0.032 and 0.002, respectively) with no significant difference between GAT/Tono-Pen (p value: 0.554). Mean difference in IOP measurements between GAT/ICare was 1.49 ± 2.61 mmHg, Tono-Pen/ICare was 1.89 ± 2.15 mmHg, and GAT/Tono-Pen was −0.39 ± 2.59 mmHg.Abstract : Purpose . To evaluate the different IOP readings by Goldmann applanation tonometer (GAT), ICare rebound tonometer, and Tono-Pen in keratoconus patients after MyoRing implantation. To assess the influence of central corneal thickness (CCT) and thinnest corneal location (TCL) on IOP measurements by different tonometers. Setting . Prospective observational study was conducted in two private centers in Egypt from February 2015 to November 2016. Methods . Seventeen eyes of 10 patients suffering from keratoconus and who underwent MyoRing implantation were recruited. All subjects underwent GAT, ICare, and Tono-Pen IOP measurements in random order. Central corneal thickness and thinnest corneal location were assessed by Pentacam. Difference in mean in IOP readings was assessed by T -test. Correlation between each pair of devices was evaluated by Pearson correlation coefficient. The Bland–Altman analysis was used to assess intertonometer agreement. Results . Seventeen eyes (10 patients) were evaluated. The mean IOP reading was 13.9 ± 3.68, 12.41 ± 2.87, and 14.29 ± 1.31 mmHg in GAT, ICare, and Tono-Pen group, respectively. There was a significant difference between IOP readings by GAT/ICare and Tono-Pen/ICare (p value: 0.032 and 0.002, respectively) with no significant difference between GAT/Tono-Pen (p value: 0.554). Mean difference in IOP measurements between GAT/ICare was 1.49 ± 2.61 mmHg, Tono-Pen/ICare was 1.89 ± 2.15 mmHg, and GAT/Tono-Pen was −0.39 ± 2.59 mmHg. There was no significant correlation between the difference in IOP readings among any pair of devices and CCC or TCL. The Bland–Altman analysis showed a reasonable agreement between any pair of tonometers. … (more)
- Is Part Of:
- Journal of ophthalmology. Volume 2019(2019)
- Journal:
- Journal of ophthalmology
- Issue:
- Volume 2019(2019)
- Issue Display:
- Volume 2019, Issue 2019 (2019)
- Year:
- 2019
- Volume:
- 2019
- Issue:
- 2019
- Issue Sort Value:
- 2019-2019-2019-0000
- Page Start:
- Page End:
- Publication Date:
- 2019-05-09
- Subjects:
- Ophthalmology -- Periodicals
Eye Diseases
Ophthalmology
Ophthalmology
Electronic journals
Periodicals
Periodicals
Fulltext
Internet Resources
Periodicals
617.7 - Journal URLs:
- https://www.hindawi.com/journals/joph/ ↗
http://www.ncbi.nlm.nih.gov/pmc/journals/1195/ ↗
http://bibpurl.oclc.org/web/46495 ↗
http://search.ebscohost.com/direct.asp?db=a9h&jid=%229038%22&scope=site ↗ - DOI:
- 10.1155/2019/1964107 ↗
- Languages:
- English
- ISSNs:
- 2090-004X
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library HMNTS - ELD Digital store
- Ingest File:
- 10836.xml