A multicenter study to evaluate automated platelet aggregometry on Sysmex CS‐series coagulation analyzers—preliminary findings. Issue 4 (30th August 2018)
- Record Type:
- Journal Article
- Title:
- A multicenter study to evaluate automated platelet aggregometry on Sysmex CS‐series coagulation analyzers—preliminary findings. Issue 4 (30th August 2018)
- Main Title:
- A multicenter study to evaluate automated platelet aggregometry on Sysmex CS‐series coagulation analyzers—preliminary findings
- Authors:
- Platton, Sean
McCormick, Áine
Bukht, Musfira
Gurney, David
Holding, Ian
Moore, Gary W. - Abstract:
- Abstract: Background and Objectives: Investigations of platelet function by light transmission aggregometry (LTA) using a dedicated aggregometer is time consuming and labor intensive. This multicenter study evaluated an automated LTA method using a coagulation analyzer to establish reference ranges and ideal testing regimen. Methods: Sysmex CS‐2x00 series analyzers were used to measure aggregation using a range of agonists and concentrations: ADP (1‐20 μM); arachidonic acid (0.5‐1.5 mM); collagen (1.25‐5 μg/mL); ristocetin (0.5‐1.5 g/L); epinephrine (5‐10 μM); TRAP (1‐20 μM); U46619 (1 μM); and saline. Maximum and final aggregation, disaggregation, slope, and acquisition time were compared for each. Results: For 42 normal subjects there was no significant difference in aggregation parameters for: 10 μM and 20 μm ADP; 2 and 2.5 μM ADP; 1 and 1.5 mM arachidonic acid; 2.5 and 5 μg/mL collagen; 1 and 1.25 μg/mL collagen; 1.25 and 1.5 g/L ristocetin; 5 and 10 μM epinephrine; 5 and 10 μM or 20 μM TRAP. Maximum aggregation was reached by 300 seconds with 20 and 10 μM ADP, 1 μM U46619, 1 and 1.25 μg/mL collagen, 1.5 g/L ristocetin and 5, 10, and 20 μM TRAP: all others agonists required 600s. Conclusions: A standard panel of agonists can be used on the Sysmex CS‐2x00 series analyzers: ADP (10, 5, 2.5, and 1.25 μM); 1 mM arachidonic acid; 1 μM U46619; 2.5 and 1.25 μg/mL collagen; 1.25 and 0.5 g/L ristocetin; 5 μM epinephrine; 5 and 10 μM TRAP; and saline. Aggregation should beAbstract: Background and Objectives: Investigations of platelet function by light transmission aggregometry (LTA) using a dedicated aggregometer is time consuming and labor intensive. This multicenter study evaluated an automated LTA method using a coagulation analyzer to establish reference ranges and ideal testing regimen. Methods: Sysmex CS‐2x00 series analyzers were used to measure aggregation using a range of agonists and concentrations: ADP (1‐20 μM); arachidonic acid (0.5‐1.5 mM); collagen (1.25‐5 μg/mL); ristocetin (0.5‐1.5 g/L); epinephrine (5‐10 μM); TRAP (1‐20 μM); U46619 (1 μM); and saline. Maximum and final aggregation, disaggregation, slope, and acquisition time were compared for each. Results: For 42 normal subjects there was no significant difference in aggregation parameters for: 10 μM and 20 μm ADP; 2 and 2.5 μM ADP; 1 and 1.5 mM arachidonic acid; 2.5 and 5 μg/mL collagen; 1 and 1.25 μg/mL collagen; 1.25 and 1.5 g/L ristocetin; 5 and 10 μM epinephrine; 5 and 10 μM or 20 μM TRAP. Maximum aggregation was reached by 300 seconds with 20 and 10 μM ADP, 1 μM U46619, 1 and 1.25 μg/mL collagen, 1.5 g/L ristocetin and 5, 10, and 20 μM TRAP: all others agonists required 600s. Conclusions: A standard panel of agonists can be used on the Sysmex CS‐2x00 series analyzers: ADP (10, 5, 2.5, and 1.25 μM); 1 mM arachidonic acid; 1 μM U46619; 2.5 and 1.25 μg/mL collagen; 1.25 and 0.5 g/L ristocetin; 5 μM epinephrine; 5 and 10 μM TRAP; and saline. Aggregation should be observed for 600 seconds for all agonists except TRAP and U46619, which require 300 seconds. If further studies confirm these concentrations detect platelet disorders then Sysmex CS‐series analyzers could replace dedicated aggregometers, or perform LTA where it is currently not available. … (more)
- Is Part Of:
- Research and practice in thrombosis and haemostasis. Volume 2:Issue 4(2018)
- Journal:
- Research and practice in thrombosis and haemostasis
- Issue:
- Volume 2:Issue 4(2018)
- Issue Display:
- Volume 2, Issue 4 (2018)
- Year:
- 2018
- Volume:
- 2
- Issue:
- 4
- Issue Sort Value:
- 2018-0002-0004-0000
- Page Start:
- 778
- Page End:
- 789
- Publication Date:
- 2018-08-30
- Subjects:
- blood platelets -- hemostasis -- laboratory diagnosis -- platelet aggregation -- standardization
Thrombosis -- Periodicals
Hemostasis -- Periodicals
616.135005 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)2475-0379 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/rth2.12140 ↗
- Languages:
- English
- ISSNs:
- 2475-0379
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 10798.xml