Comparison of microsphere penetration with LC Bead LUMI™ versus other commercial microspheres. (February 2018)
- Record Type:
- Journal Article
- Title:
- Comparison of microsphere penetration with LC Bead LUMI™ versus other commercial microspheres. (February 2018)
- Main Title:
- Comparison of microsphere penetration with LC Bead LUMI™ versus other commercial microspheres
- Authors:
- Caine, Marcus
Zhang, Xunli
Hill, Martyn
Guo, Wei
Ashrafi, Koorosh
Bascal, Zainab
Kilpatrick, Hugh
Dunn, Anthony
Grey, David
Bushby, Rosemary
Bushby, Andrew
Willis, Sean L.
Dreher, Matthew R.
Lewis, Andrew L. - Abstract:
- Abstract: The purpose of this study was to evaluate LC Bead LUMI™ (40–90 µm and 70–150 µm) in order to determine if their increased resistance to compression influences microsphere penetration and distribution compared to more compressible commercial microspheres. LC Bead LUMI™ 40–90 µm and 70–150 µm, LC Bead M1 ® 70–150 µm, Embozene™ 40 µm and Embozene™ 100 µm size and distributions were measured using optical microscopy. Penetration in vitro was evaluated using an established 'plate model', consisting of a calibrated tapered gap between a glass plate and plastic housing to allow visual observation of microsphere penetration depth. Behaviour in vivo was assessed using a rabbit renal embolization model with histopathologic confirmation of vessel penetration depth. Penetration behaviour in vitro was reproducible and commensurate with the measured microsphere size, the smaller the microsphere the deeper the penetration. Comparison of the microsphere diameter measured on the 2D plate model versus the corresponding average microsphere size measured by histopathology in the kidney showed no significant differences ( p = > 0.05 Mann-Whitney, demonstrating good in vitro - in vivo predictive capabilities of the plate model) confirming predictable performance for LC Bead LUMI™ (40–90 µm and 70–150 µm) based on microsphere size, their increased rigidity having no bearing on their depth of penetration and distribution. An assessment of a LC Bead LUMI™ (40–90 µm and 70–150 µm) has shownAbstract: The purpose of this study was to evaluate LC Bead LUMI™ (40–90 µm and 70–150 µm) in order to determine if their increased resistance to compression influences microsphere penetration and distribution compared to more compressible commercial microspheres. LC Bead LUMI™ 40–90 µm and 70–150 µm, LC Bead M1 ® 70–150 µm, Embozene™ 40 µm and Embozene™ 100 µm size and distributions were measured using optical microscopy. Penetration in vitro was evaluated using an established 'plate model', consisting of a calibrated tapered gap between a glass plate and plastic housing to allow visual observation of microsphere penetration depth. Behaviour in vivo was assessed using a rabbit renal embolization model with histopathologic confirmation of vessel penetration depth. Penetration behaviour in vitro was reproducible and commensurate with the measured microsphere size, the smaller the microsphere the deeper the penetration. Comparison of the microsphere diameter measured on the 2D plate model versus the corresponding average microsphere size measured by histopathology in the kidney showed no significant differences ( p = > 0.05 Mann-Whitney, demonstrating good in vitro - in vivo predictive capabilities of the plate model) confirming predictable performance for LC Bead LUMI™ (40–90 µm and 70–150 µm) based on microsphere size, their increased rigidity having no bearing on their depth of penetration and distribution. An assessment of a LC Bead LUMI™ (40–90 µm and 70–150 µm) has shown that despite having greater resistance to compression, these microspheres behave in a predictable manner within in vitro and in vivo models comparable with more compressible microspheres of similar sizes. Graphical abstract: … (more)
- Is Part Of:
- Journal of the mechanical behavior of biomedical materials. Volume 78(2018)
- Journal:
- Journal of the mechanical behavior of biomedical materials
- Issue:
- Volume 78(2018)
- Issue Display:
- Volume 78, Issue 2018 (2018)
- Year:
- 2018
- Volume:
- 78
- Issue:
- 2018
- Issue Sort Value:
- 2018-0078-2018-0000
- Page Start:
- 46
- Page End:
- 55
- Publication Date:
- 2018-02
- Subjects:
- Embolization microspheres -- Penetration plate model -- Renal embolization model -- Compressive modulus
Biomedical materials -- Periodicals
Biomedical materials -- Mechanical properties -- Periodicals
Biomedical materials
Biomedical materials -- Mechanical properties
Periodicals
Electronic journals
610.28 - Journal URLs:
- http://www.sciencedirect.com/science/journal/17516161 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.jmbbm.2017.10.034 ↗
- Languages:
- English
- ISSNs:
- 1751-6161
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5015.809000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 10784.xml