Evaluating the critical strain energy release rate of bioactive glass coatings on Ti6Al4V substrates after degradation. (February 2018)
- Record Type:
- Journal Article
- Title:
- Evaluating the critical strain energy release rate of bioactive glass coatings on Ti6Al4V substrates after degradation. (February 2018)
- Main Title:
- Evaluating the critical strain energy release rate of bioactive glass coatings on Ti6Al4V substrates after degradation
- Authors:
- Matinmanesh, A.
Li, Y.
Nouhi, A.
Zalzal, P.
Schemitsch, E.H.
Towler, M.R.
Papini, M. - Abstract:
- Abstract: It has been reported that the adhesion of bioactive glass coatings to Ti6Al4V reduces after degradation, however, this effect has not been quantified. This paper uses bilayer double cantilever (DCB) specimens to determine G IC and G IIC, the critical mode I and mode II strain energy release rates, respectively, of bioactive coating/Ti6Al4V substrate systems degraded to different extents. Three borate-based bioactive glass coatings with increasing amounts of incorporated SrO (0, 15 and 25 mol%) were enamelled onto Ti6Al4V substrates and then immersed in de-ionized water for 2, 6 and 24 h. The weight loss of each glass composition was measured and it was found that the dissolution rate significantly decreased with increasing SrO content. The extent of dissolution was consistent with the hypothesis that the compressive residual stress tends to reduce the dissolution rate of bioactive glasses. After drying, the bilayer DCB specimens were created and subjected to nearly mode I and mode II fracture tests. The toughest coating/substrate system (one composed of the glass containing 25 mol% SrO) lost 80% and 85% of its G IC and G IIC, respectively, in less than 24 h of degradation. The drop in G IC and G IIC occurred even more rapidly for other coating/substrate systems. Therefore, degradation of borate bioactive glass coatings is inversely related to their fracture toughness when coated onto Ti6A4V substrates. Finally, roughening the substrate was found to beAbstract: It has been reported that the adhesion of bioactive glass coatings to Ti6Al4V reduces after degradation, however, this effect has not been quantified. This paper uses bilayer double cantilever (DCB) specimens to determine G IC and G IIC, the critical mode I and mode II strain energy release rates, respectively, of bioactive coating/Ti6Al4V substrate systems degraded to different extents. Three borate-based bioactive glass coatings with increasing amounts of incorporated SrO (0, 15 and 25 mol%) were enamelled onto Ti6Al4V substrates and then immersed in de-ionized water for 2, 6 and 24 h. The weight loss of each glass composition was measured and it was found that the dissolution rate significantly decreased with increasing SrO content. The extent of dissolution was consistent with the hypothesis that the compressive residual stress tends to reduce the dissolution rate of bioactive glasses. After drying, the bilayer DCB specimens were created and subjected to nearly mode I and mode II fracture tests. The toughest coating/substrate system (one composed of the glass containing 25 mol% SrO) lost 80% and 85% of its G IC and G IIC, respectively, in less than 24 h of degradation. The drop in G IC and G IIC occurred even more rapidly for other coating/substrate systems. Therefore, degradation of borate bioactive glass coatings is inversely related to their fracture toughness when coated onto Ti6A4V substrates. Finally, roughening the substrate was found to be inconsequential in increasing the toughness of the system as the fracture toughness was limited by the cohesive toughness of the glass itself. Graphical abstract: … (more)
- Is Part Of:
- Journal of the mechanical behavior of biomedical materials. Volume 78(2018)
- Journal:
- Journal of the mechanical behavior of biomedical materials
- Issue:
- Volume 78(2018)
- Issue Display:
- Volume 78, Issue 2018 (2018)
- Year:
- 2018
- Volume:
- 78
- Issue:
- 2018
- Issue Sort Value:
- 2018-0078-2018-0000
- Page Start:
- 273
- Page End:
- 281
- Publication Date:
- 2018-02
- Subjects:
- Borate glass coating -- Ti6Al4V implants -- Degradation -- Fracture toughness
Biomedical materials -- Periodicals
Biomedical materials -- Mechanical properties -- Periodicals
Biomedical materials
Biomedical materials -- Mechanical properties
Periodicals
Electronic journals
610.28 - Journal URLs:
- http://www.sciencedirect.com/science/journal/17516161 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.jmbbm.2017.11.015 ↗
- Languages:
- English
- ISSNs:
- 1751-6161
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5015.809000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 10784.xml