Log file-based patient dose calculations of double-arc VMAT for head-and-neck radiotherapy. (April 2018)
- Record Type:
- Journal Article
- Title:
- Log file-based patient dose calculations of double-arc VMAT for head-and-neck radiotherapy. (April 2018)
- Main Title:
- Log file-based patient dose calculations of double-arc VMAT for head-and-neck radiotherapy
- Authors:
- Katsuta, Yoshiyuki
Kadoya, Noriyuki
Fujita, Yukio
Shimizu, Eiji
Majima, Kazuhiro
Matsushita, Haruo
Takeda, Ken
Jingu, Keiichi - Abstract:
- Highlights: Dosimetric changes on double-arc VMAT in log file-based method were explored. Changes in Dmean were 0.9 Gy on PTV and <0.7 Gy on OARs. Changes in TCP/NTCP were 1.4% on PTV and <1.8% on OARs. Abstract: Purpose: The log file-based method cannot display dosimetric changes due to linac component miscalibration because of the insensitivity of log files to linac component miscalibration. The purpose of this study was to supply dosimetric changes in log file-based patient dose calculations for double-arc volumetric-modulated arc therapy (VMAT) in head-and-neck cases. Methods and materials: Fifteen head-and-neck cases participated in this study. For each case, treatment planning system (TPS) doses were produced by double-arc and single-arc VMAT. Miscalibration-simulated log files were generated by inducing a leaf miscalibration of ±0.5 mm into the log files that were acquired during VMAT irradiation. Subsequently, patient doses were estimated using the miscalibration-simulated log files. Results: For double-arc VMAT, regarding planning target volume (PTV), the change from TPS dose to miscalibration-simulated log file dose in Dmean was 0.9 Gy and that for tumor control probability was 1.4%. As for organ-at-risks (OARs), the change in Dmean was <0.7 Gy and normal tissue complication probability was <1.8%. A comparison between double-arc and single-arc VMAT for PTV showed statistically significant differences in the changes evaluated by Dmean and radiobiological metrics ( PHighlights: Dosimetric changes on double-arc VMAT in log file-based method were explored. Changes in Dmean were 0.9 Gy on PTV and <0.7 Gy on OARs. Changes in TCP/NTCP were 1.4% on PTV and <1.8% on OARs. Abstract: Purpose: The log file-based method cannot display dosimetric changes due to linac component miscalibration because of the insensitivity of log files to linac component miscalibration. The purpose of this study was to supply dosimetric changes in log file-based patient dose calculations for double-arc volumetric-modulated arc therapy (VMAT) in head-and-neck cases. Methods and materials: Fifteen head-and-neck cases participated in this study. For each case, treatment planning system (TPS) doses were produced by double-arc and single-arc VMAT. Miscalibration-simulated log files were generated by inducing a leaf miscalibration of ±0.5 mm into the log files that were acquired during VMAT irradiation. Subsequently, patient doses were estimated using the miscalibration-simulated log files. Results: For double-arc VMAT, regarding planning target volume (PTV), the change from TPS dose to miscalibration-simulated log file dose in Dmean was 0.9 Gy and that for tumor control probability was 1.4%. As for organ-at-risks (OARs), the change in Dmean was <0.7 Gy and normal tissue complication probability was <1.8%. A comparison between double-arc and single-arc VMAT for PTV showed statistically significant differences in the changes evaluated by Dmean and radiobiological metrics ( P < 0.01), even though the magnitude of these differences was small. Similarly, for OARs, the magnitude of these changes was found to be small. Conclusions: Using the log file-based method for PTV and OARs, the log file-based method estimate of patient dose using the double-arc VMAT has accuracy comparable to that obtained using the single-arc VMAT. … (more)
- Is Part Of:
- Physica medica. Volume 48(2018)
- Journal:
- Physica medica
- Issue:
- Volume 48(2018)
- Issue Display:
- Volume 48, Issue 2018 (2018)
- Year:
- 2018
- Volume:
- 48
- Issue:
- 2018
- Issue Sort Value:
- 2018-0048-2018-0000
- Page Start:
- 6
- Page End:
- 10
- Publication Date:
- 2018-04
- Subjects:
- Radiotherapy -- Head-and-neck -- VMAT -- Log file -- Patient-specific QA
Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2018.03.006 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 10724.xml