Novel light-guide-PMT geometries to reduce dead edges of a scintillation camera. (April 2018)
- Record Type:
- Journal Article
- Title:
- Novel light-guide-PMT geometries to reduce dead edges of a scintillation camera. (April 2018)
- Main Title:
- Novel light-guide-PMT geometries to reduce dead edges of a scintillation camera
- Authors:
- Wang, Beien
Kreuger, Rob
Beekman, Freek J.
Goorden, Marlies C. - Abstract:
- Highlights: Cost-effective solution to reduce dead edges for Anger camera. Unconventional PMT-lightguide layout, absorptive edges, and maximum likelihood positioning. Evaluated with Monte Carlo simulation toolkit GATE. Abstract: Anger cameras based on monolithic NaI scintillators read out by an array of PMTs are predominant in planar gamma imaging and SPECT. However, position estimation of gamma interactions is usually severely degraded near the edges of the scintillator which can be extremely undesirable for applications like breast imaging. Here we propose a relatively cost-effective solution based on the use of scintillators with absorptive edges with an unconventional light-guide-PMT layout employing a maximum likelihood positioning algorithm. The basic design on which we aim to improve consists of a monolithic NaI(Tl) scintillator read out by 3 × 5 square PMTs (conventional layout, CL) that could be suitable for molecular breast imaging. To better detect gamma interactions near the crystal's critical edge, we tried different set-ups: we replaced the 5 large PMTs near the edge by 11 smaller PMTs (small-sensor layout, SSL); we emulated rectangular PMTs along the critical edge by inserting a row of 5 rectangular light-guides that direct the light toward square PMTs placed behind (shifted layout, SL); we inserted rectangular light-guides alternatingly, such that the PMTs are in an interlocking pattern (alternating shifted layout, ASL). The performance of our designs wasHighlights: Cost-effective solution to reduce dead edges for Anger camera. Unconventional PMT-lightguide layout, absorptive edges, and maximum likelihood positioning. Evaluated with Monte Carlo simulation toolkit GATE. Abstract: Anger cameras based on monolithic NaI scintillators read out by an array of PMTs are predominant in planar gamma imaging and SPECT. However, position estimation of gamma interactions is usually severely degraded near the edges of the scintillator which can be extremely undesirable for applications like breast imaging. Here we propose a relatively cost-effective solution based on the use of scintillators with absorptive edges with an unconventional light-guide-PMT layout employing a maximum likelihood positioning algorithm. The basic design on which we aim to improve consists of a monolithic NaI(Tl) scintillator read out by 3 × 5 square PMTs (conventional layout, CL) that could be suitable for molecular breast imaging. To better detect gamma interactions near the crystal's critical edge, we tried different set-ups: we replaced the 5 large PMTs near the edge by 11 smaller PMTs (small-sensor layout, SSL); we emulated rectangular PMTs along the critical edge by inserting a row of 5 rectangular light-guides that direct the light toward square PMTs placed behind (shifted layout, SL); we inserted rectangular light-guides alternatingly, such that the PMTs are in an interlocking pattern (alternating shifted layout, ASL). The performance of our designs was tested with Monte Carlo simulations. Results showed that SSL, SL, and ASL gave better spatial resolution near the critical edge than CL (3.4, 3.6, and 4.1 mm near the edge compared with 5.3 mm for CL), and thus resulted in a larger usable detector area. To conclude, for applications where small dead edges are crucial, our designs may be cost-effective solutions. … (more)
- Is Part Of:
- Physica medica. Volume 48(2018)
- Journal:
- Physica medica
- Issue:
- Volume 48(2018)
- Issue Display:
- Volume 48, Issue 2018 (2018)
- Year:
- 2018
- Volume:
- 48
- Issue:
- 2018
- Issue Sort Value:
- 2018-0048-2018-0000
- Page Start:
- 84
- Page End:
- 90
- Publication Date:
- 2018-04
- Subjects:
- Gamma-ray detectors -- Maximum likelihood estimation -- Photomultiplier tube -- Dead edge
Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2018.04.004 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 10724.xml