Comparison of matrix-assisted laser desorption ionization time of flight mass spectrometry (MALDI-TOF MS) systems for the identification of moulds in the routine microbiology laboratory. (July 2019)
- Record Type:
- Journal Article
- Title:
- Comparison of matrix-assisted laser desorption ionization time of flight mass spectrometry (MALDI-TOF MS) systems for the identification of moulds in the routine microbiology laboratory. (July 2019)
- Main Title:
- Comparison of matrix-assisted laser desorption ionization time of flight mass spectrometry (MALDI-TOF MS) systems for the identification of moulds in the routine microbiology laboratory
- Authors:
- Dupont, D.
Normand, A.-C.
Persat, F.
Hendrickx, M.
Piarroux, R.
Wallon, M. - Abstract:
- Abstract: Objectives: The purpose of this study was to compare the efficiency of mould identification of two matrix-assisted laser desorption ionization time of flight mass spectrometry (MALDI-TOF MS) systems – Vitek MS (VMS) and Microflex LT (MLT) – and the MSI application. Methods: Moulds were collected retrospectively and prospectively to display epidemiological diversity of a microbiology laboratory. All of them were identified via sequencing. Strains were then identified using the VMS v3.0, the MLT, and the MSI software applied on MLT spectra. Rates of correct identifications to the species, to the complex, and to the genus level were compared with the molecular reference standard. Results: A total of 102 isolates were collected. The rate of correct identification to the species level with the MLT was 42.2% (43/102) with a threshold of 1.7 (vs. 16.7% (17/102) with a threshold of 2.0, p < 0.05). The VMS performed better than the MLT with a threshold of 1.7 for species (49.0% (50/102), p 0.33) and complex level identifications (71.6% (73/102) vs. 54.9% (56/102), p < 0.05). However the highest performances were observed when the MLT spectra were analysed via the Mass Spectrometry Identification (MSI) software reaching 90.2% (92/102) of correct identification to the species, 92.2% (94/102) to the species complex and 94.1% (96/102) to the genus level. Conclusions: The VMS performed better than the MLT for mould identification. However, it remains of utmost importance toAbstract: Objectives: The purpose of this study was to compare the efficiency of mould identification of two matrix-assisted laser desorption ionization time of flight mass spectrometry (MALDI-TOF MS) systems – Vitek MS (VMS) and Microflex LT (MLT) – and the MSI application. Methods: Moulds were collected retrospectively and prospectively to display epidemiological diversity of a microbiology laboratory. All of them were identified via sequencing. Strains were then identified using the VMS v3.0, the MLT, and the MSI software applied on MLT spectra. Rates of correct identifications to the species, to the complex, and to the genus level were compared with the molecular reference standard. Results: A total of 102 isolates were collected. The rate of correct identification to the species level with the MLT was 42.2% (43/102) with a threshold of 1.7 (vs. 16.7% (17/102) with a threshold of 2.0, p < 0.05). The VMS performed better than the MLT with a threshold of 1.7 for species (49.0% (50/102), p 0.33) and complex level identifications (71.6% (73/102) vs. 54.9% (56/102), p < 0.05). However the highest performances were observed when the MLT spectra were analysed via the Mass Spectrometry Identification (MSI) software reaching 90.2% (92/102) of correct identification to the species, 92.2% (94/102) to the species complex and 94.1% (96/102) to the genus level. Conclusions: The VMS performed better than the MLT for mould identification. However, it remains of utmost importance to expand commercial databases, as performances of the MLT highly improved when using the MSI software and its extended database, reaching far above the VMS system. Thus the VMS could benefit from the use of this online tool. … (more)
- Is Part Of:
- Clinical microbiology and infection. Volume 25:Number 7(2019)
- Journal:
- Clinical microbiology and infection
- Issue:
- Volume 25:Number 7(2019)
- Issue Display:
- Volume 25, Issue 7 (2019)
- Year:
- 2019
- Volume:
- 25
- Issue:
- 7
- Issue Sort Value:
- 2019-0025-0007-0000
- Page Start:
- 892
- Page End:
- 897
- Publication Date:
- 2019-07
- Subjects:
- Filamentous fungi -- Identification -- MALDI-TOF -- Mass spectrometry -- Moulds
Medical microbiology -- Periodicals
Diagnostic microbiology -- Periodicals
Communicable diseases -- Periodicals
Infection -- Periodicals
616.01 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1469-0691 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1016/j.cmi.2018.10.013 ↗
- Languages:
- English
- ISSNs:
- 1198-743X
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3286.305520
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 10711.xml