Electrochemical immunosensor based on an antibody-hierarchical mesoporous SiO2 for the detection of Staphylococcus aureus. Issue 28 (24th May 2019)
- Record Type:
- Journal Article
- Title:
- Electrochemical immunosensor based on an antibody-hierarchical mesoporous SiO2 for the detection of Staphylococcus aureus. Issue 28 (24th May 2019)
- Main Title:
- Electrochemical immunosensor based on an antibody-hierarchical mesoporous SiO2 for the detection of Staphylococcus aureus
- Authors:
- Wang, Hongsu
Xiu, Yi
Chen, Yan
Sun, Liping
Yang, Libin
Chen, Honghao
Niu, Xiaodi - Abstract:
- Abstract : Electrochemical immunosensor based on an Ab-HMS materials using butterfly wings as biotemplate is successfully prepared. It is not only simple to manufacture, but also has a fast detection time and can detect S. aureus with low sensitivity. Abstract : The outbreak of food-borne pathogens has become a serious concern; therefore, the detection of pathogenic bacteria in food is required. Untreated, sensitive, and reliable sensors should be developed for the detection of Staphylococcus aureus ( S. aureus ). In this study, a sensitive antibody-based electrochemical immunosensor was developed using antibody (Ab)-hierarchical mesoporous silica (HMS) bio-conjugates for label-free detection of low concentrations of S. aureus . First, a bio-template method based on butterfly wings was used to prepare the HMS. Then, the carrier material was amino-functionalized to cross-link the antibody with glutaraldehyde. The Ab-HMS bio-conjugates were then immobilized on a glassy carbon electrode (GCE), and the presence of S. aureus was detected by analyzing the changes in the peak currents after the antigen–antibody complex formation. Differential pulse voltammetry (DPV) was performed with bacterial concentrations ranging from 10 to 2 × 10 3 colony forming units (CFU) mL −1 . Selective tests were performed using Escherichia coli ( E. coli ), Listeria monocytogenes ( L. monocytohenes ), and Salmonella, and the selective assays showed specific detection of S. aureus using the sensor. InAbstract : Electrochemical immunosensor based on an Ab-HMS materials using butterfly wings as biotemplate is successfully prepared. It is not only simple to manufacture, but also has a fast detection time and can detect S. aureus with low sensitivity. Abstract : The outbreak of food-borne pathogens has become a serious concern; therefore, the detection of pathogenic bacteria in food is required. Untreated, sensitive, and reliable sensors should be developed for the detection of Staphylococcus aureus ( S. aureus ). In this study, a sensitive antibody-based electrochemical immunosensor was developed using antibody (Ab)-hierarchical mesoporous silica (HMS) bio-conjugates for label-free detection of low concentrations of S. aureus . First, a bio-template method based on butterfly wings was used to prepare the HMS. Then, the carrier material was amino-functionalized to cross-link the antibody with glutaraldehyde. The Ab-HMS bio-conjugates were then immobilized on a glassy carbon electrode (GCE), and the presence of S. aureus was detected by analyzing the changes in the peak currents after the antigen–antibody complex formation. Differential pulse voltammetry (DPV) was performed with bacterial concentrations ranging from 10 to 2 × 10 3 colony forming units (CFU) mL −1 . Selective tests were performed using Escherichia coli ( E. coli ), Listeria monocytogenes ( L. monocytohenes ), and Salmonella, and the selective assays showed specific detection of S. aureus using the sensor. In addition, the immunosensor showed a good linear relationship between the peak current increase and logarithmic S. aureus concentration ( R 2 = 0.9759) with a fast detection time (20 min) and detection limit of 11 CFU mL −1 . When the electrochemical impedance spectroscopy (EIS) was performed under the same conditions, the results showed a good linear relationship between the impedance change value and the bacterial concentration ( R 2 = 0.9720), the limit of detection (LOD) was 12 CFU mL −1 . The performance of the sensor was compared with that of the colony counting method in the spiked milk sample test. The results showed no significant difference in the test results. Hence, this electrochemical immunosensor can be used to quickly detect S. aureus in actual food samples with a high sensitivity, specificity and stability. … (more)
- Is Part Of:
- RSC advances. Volume 9:Issue 28(2019)
- Journal:
- RSC advances
- Issue:
- Volume 9:Issue 28(2019)
- Issue Display:
- Volume 9, Issue 28 (2019)
- Year:
- 2019
- Volume:
- 9
- Issue:
- 28
- Issue Sort Value:
- 2019-0009-0028-0000
- Page Start:
- 16278
- Page End:
- 16287
- Publication Date:
- 2019-05-24
- Subjects:
- Chemistry -- Periodicals
540.5 - Journal URLs:
- http://pubs.rsc.org/en/Journals/JournalIssues/RA ↗
http://www.rsc.org/ ↗ - DOI:
- 10.1039/c9ra00907h ↗
- Languages:
- English
- ISSNs:
- 2046-2069
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 8036.750300
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 10669.xml