A novel method for monitoring the constancy of beam path accuracy in CyberKnife1. Issue 5 (19th April 2019)
- Record Type:
- Journal Article
- Title:
- A novel method for monitoring the constancy of beam path accuracy in CyberKnife1. Issue 5 (19th April 2019)
- Main Title:
- A novel method for monitoring the constancy of beam path accuracy in CyberKnife1
- Authors:
- Yang, Bin
Wong, Wing Kei Rebecca
Lam, Wai Wang
Geng, Hui
Kong, Chi Wah
Cheung, Kin Yin
Yu, Siu Ki - Abstract:
- Abstract: The aim of current work was to present a novel evaluation procedure implemented for checking the constancy of beam path accuracy of a CyberKnife system based on ArcCHECK. A tailor‐made Styrofoam with four implanted fiducial markers was adopted to enable the fiducial tracking during beam deliveries. A simple two‐field plan and an isocentric plan were created for determining the density override of ArcCHECK in MultiPlan and the constancy of beam path accuracy respectively. Correlation curves for all diodes involved in the study were obtained by analyzing the dose distributions calculated by MultiPlan after introducing position shifts in anteroposterior, superoinferior, and left–right directions. The ability of detecting systematic position error was also evaluated by changing the position of alignment center intentionally. The one standard deviation (SD) result for reproducibility test showed the RMS of 0.054 mm and the maximum of 0.263 mm, which was comparable to the machine self‐test result. The mean of absolute value of position errors in the constancy test was measured to 0.091 mm with a SD of 0.035 mm, while the root‐mean‐square was 0.127 mm with a SD of 0.034 mm. All introduced systematic position errors range from 0.3 to 2 mm were detected successfully. Efficient method for evaluating the constancy of beam path accuracy of CyberKnife has been developed and proven to be sensitive enough for detecting a systematic drift of robotic manipulator. Once the workflowAbstract: The aim of current work was to present a novel evaluation procedure implemented for checking the constancy of beam path accuracy of a CyberKnife system based on ArcCHECK. A tailor‐made Styrofoam with four implanted fiducial markers was adopted to enable the fiducial tracking during beam deliveries. A simple two‐field plan and an isocentric plan were created for determining the density override of ArcCHECK in MultiPlan and the constancy of beam path accuracy respectively. Correlation curves for all diodes involved in the study were obtained by analyzing the dose distributions calculated by MultiPlan after introducing position shifts in anteroposterior, superoinferior, and left–right directions. The ability of detecting systematic position error was also evaluated by changing the position of alignment center intentionally. The one standard deviation (SD) result for reproducibility test showed the RMS of 0.054 mm and the maximum of 0.263 mm, which was comparable to the machine self‐test result. The mean of absolute value of position errors in the constancy test was measured to 0.091 mm with a SD of 0.035 mm, while the root‐mean‐square was 0.127 mm with a SD of 0.034 mm. All introduced systematic position errors range from 0.3 to 2 mm were detected successfully. Efficient method for evaluating the constancy of beam path accuracy of CyberKnife has been developed and proven to be sensitive enough for detecting a systematic drift of robotic manipulator. Once the workflow is streamlined, our proposed method will be an effective and easy quality assurance procedure for medical physicists. … (more)
- Is Part Of:
- Journal of applied clinical medical physics. Volume 20:Issue 5(2019)
- Journal:
- Journal of applied clinical medical physics
- Issue:
- Volume 20:Issue 5(2019)
- Issue Display:
- Volume 20, Issue 5 (2019)
- Year:
- 2019
- Volume:
- 20
- Issue:
- 5
- Issue Sort Value:
- 2019-0020-0005-0000
- Page Start:
- 109
- Page End:
- 119
- Publication Date:
- 2019-04-19
- Subjects:
- ArcCHECK -- beam path accuracy -- CyberKnife -- quality assurance
Medical physics -- Periodicals
Clinical medicine -- Periodicals
Health Physics
Clinical Medicine
Electronic journals
Periodicals
Periodicals
Fulltext
Internet Resources
610.153 - Journal URLs:
- http://aapm.onlinelibrary.wiley.com/hub/journal/10.1002/(ISSN)1526-9914/ ↗
http://bibpurl.oclc.org/web/7294 ↗
http://www.jacmp.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/acm2.12585 ↗
- Languages:
- English
- ISSNs:
- 1526-9914
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 10406.xml