Myers, Matthew R. - Abstract:
- Abstract: Objective: Blenderized diets are gaining increasing popularity among enteral tube users. Connectors in gastrostomy tubes (G-tubes) are undergoing standardization to reduce misconnections. These standardized G-tubes are referred to as ENFit G-tubes. This study was performed to quantify the in vitro performance of existing (legacy) G-tubes and compare them with ENFit G-tubes for blenderized diets. Method: Patient blenderized diet recipes and practices were obtained through patient advocacy groups. Different blenders and blending times were studied. Five legacy G-tube brands and three corresponding ENFit brands, sized between 14 Fr and 24 Fr, were studied under gravity and push modes of feeding. Results: Considering both thin and thick blenderized gravity mode diets, an average increase in feeding time from 20 minutes to 32 ± 18 minutes in transitioning from legacy to ENFit was observed with standard G-tubes, compared to 22 ± 3.5 minutes for low profiles. For push-mode diets, a 60-second push with standard ENFit G-tubes was easier compared to standard legacy G-tubes (61% ± 21% as much force), but faster 5-second pushes required considerably more effort for ENFit standard G-tubes (167% ± 96%). Low-profile ENFit G-tubes required slightly less effort compared to low-profile legacies for both 60-second and 5-second pushes (72% ± 22% and 90% ± 19%, respectively). Clogging was common in both legacy and ENFit devices, particularly under gravity mode. Conclusions: For a pushAbstract: Objective: Blenderized diets are gaining increasing popularity among enteral tube users. Connectors in gastrostomy tubes (G-tubes) are undergoing standardization to reduce misconnections. These standardized G-tubes are referred to as ENFit G-tubes. This study was performed to quantify the in vitro performance of existing (legacy) G-tubes and compare them with ENFit G-tubes for blenderized diets. Method: Patient blenderized diet recipes and practices were obtained through patient advocacy groups. Different blenders and blending times were studied. Five legacy G-tube brands and three corresponding ENFit brands, sized between 14 Fr and 24 Fr, were studied under gravity and push modes of feeding. Results: Considering both thin and thick blenderized gravity mode diets, an average increase in feeding time from 20 minutes to 32 ± 18 minutes in transitioning from legacy to ENFit was observed with standard G-tubes, compared to 22 ± 3.5 minutes for low profiles. For push-mode diets, a 60-second push with standard ENFit G-tubes was easier compared to standard legacy G-tubes (61% ± 21% as much force), but faster 5-second pushes required considerably more effort for ENFit standard G-tubes (167% ± 96%). Low-profile ENFit G-tubes required slightly less effort compared to low-profile legacies for both 60-second and 5-second pushes (72% ± 22% and 90% ± 19%, respectively). Clogging was common in both legacy and ENFit devices, particularly under gravity mode. Conclusions: For a push mode of feeding, patients will largely be unimpacted after the transition to ENFit. For a gravity mode of feeding, some ENFit users may need higher-powered blenders and should expect increased feeding times. … (more)
- Is Part Of:
- Journal of the American College of Nutrition. Volume 38:Number 4(2019)
- Journal:
- Journal of the American College of Nutrition
- Issue:
- Volume 38:Number 4(2019)
- Issue Display:
- Volume 38, Issue 4 (2019)
- Year:
- 2019
- Volume:
- 38
- Issue:
- 4
- Issue Sort Value:
- 2019-0038-0004-0000
- Page Start:
- 311
- Page End:
- 317
- Publication Date:
- 2019-05-19
- Subjects:
- ENFit -- gastrostomy tubes -- feeding tubes -- misconnections -- blenderized diets
Nutrition -- Periodicals
Nutrition disorders -- Periodicals
613.2 - Journal URLs:
- http://www.tandfonline.com/action/aboutThisJournal?journalCode=uacn20 ↗
http://www.tandfonline.com/ ↗ - DOI:
- 10.1080/07315724.2018.1509247 ↗
- Languages:
- English
- ISSNs:
- 0731-5724
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4685.780000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 10327.xml