Characterizing the clinical implementation of a novel activation-repolarization metric to identify targets for catheter ablation of ventricular tachycardias using computational models. (May 2019)
- Record Type:
- Journal Article
- Title:
- Characterizing the clinical implementation of a novel activation-repolarization metric to identify targets for catheter ablation of ventricular tachycardias using computational models. (May 2019)
- Main Title:
- Characterizing the clinical implementation of a novel activation-repolarization metric to identify targets for catheter ablation of ventricular tachycardias using computational models
- Authors:
- Campos, Fernando O.
Orini, Michele
Taggart, Peter
Hanson, Ben
Lambiase, Pier D.
Porter, Bradley
Rinaldi, Christopher Aldo
Gill, Jaswinder
Bishop, Martin J. - Abstract:
- Abstract: Identification of targets for catheter ablation of ventricular tachycardias (VTs) remains a significant challenge. VTs are often driven by re-entrant circuits resulting from a complex interaction between the front (activation) and tail (repolarization) of the electrical wavefront. Most mapping techniques do not take into account the tissue repolarization which may hinder the detection of ablation targets. The re-entry vulnerability index (RVI), a recently proposed mapping procedure, incorporates both activation and repolarization times to uncover VT circuits. The method showed potential in a series of experiments, but it still requires further development to enable its incorporation into a clinical protocol. Here, in-silico experiments were conducted to thoroughly assess RVI maps constructed under clinically-relevant mapping conditions. Within idealized as well as anatomically realistic infarct models, we show that parameters of the algorithm such as the search radius can significantly alter the specificity and sensitivity of the RVI maps. When constructed on sparse grids obtained following various placements of clinical recording catheters, RVI maps can identify vulnerable regions as long as two electrodes were placed on both sides of the line of block. Moreover, maps computed during pacing without inducing VT can reveal areas of abnormal repolarization and slow conduction but not directly vulnerability. In conclusion, the RVI algorithm can detect re-entrantAbstract: Identification of targets for catheter ablation of ventricular tachycardias (VTs) remains a significant challenge. VTs are often driven by re-entrant circuits resulting from a complex interaction between the front (activation) and tail (repolarization) of the electrical wavefront. Most mapping techniques do not take into account the tissue repolarization which may hinder the detection of ablation targets. The re-entry vulnerability index (RVI), a recently proposed mapping procedure, incorporates both activation and repolarization times to uncover VT circuits. The method showed potential in a series of experiments, but it still requires further development to enable its incorporation into a clinical protocol. Here, in-silico experiments were conducted to thoroughly assess RVI maps constructed under clinically-relevant mapping conditions. Within idealized as well as anatomically realistic infarct models, we show that parameters of the algorithm such as the search radius can significantly alter the specificity and sensitivity of the RVI maps. When constructed on sparse grids obtained following various placements of clinical recording catheters, RVI maps can identify vulnerable regions as long as two electrodes were placed on both sides of the line of block. Moreover, maps computed during pacing without inducing VT can reveal areas of abnormal repolarization and slow conduction but not directly vulnerability. In conclusion, the RVI algorithm can detect re-entrant circuits during VT from low resolution mapping grids resembling the clinical setting. Furthermore, RVI maps may provide information about the underlying tissue electrophysiology to guide catheter ablation without the need of inducing potentially harmful VT during the clinical procedure. Finally, the ability of the RVI maps to identify vulnerable regions with specificity in high resolution computer models could potentially improve the prediction of optimal ablation targets of simulation-based strategies. Highlights: Safe and accurate detection of targets for catheter ablation remains a challenge. We conducted a thorough assessment of the Re-entry Vulnerability Index (RVI). Parameters of the algorithm can alter the specificity and sensitivity of RVI maps. When constructed on sparse grids RVI maps could still detect arrhythmogenic sites. In absence of arrhythmia, RVI maps revealed abnormal sites, but not vulnerability. … (more)
- Is Part Of:
- Computers in biology and medicine. Volume 108(2019)
- Journal:
- Computers in biology and medicine
- Issue:
- Volume 108(2019)
- Issue Display:
- Volume 108, Issue 2019 (2019)
- Year:
- 2019
- Volume:
- 108
- Issue:
- 2019
- Issue Sort Value:
- 2019-0108-2019-0000
- Page Start:
- 263
- Page End:
- 275
- Publication Date:
- 2019-05
- Subjects:
- Arrhythmia -- Ventricular tachycardia -- Ablation -- Computer simulation
Medicine -- Data processing -- Periodicals
Biology -- Data processing -- Periodicals
610.285 - Journal URLs:
- http://www.sciencedirect.com/science/journal/00104825/ ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.compbiomed.2019.03.018 ↗
- Languages:
- English
- ISSNs:
- 0010-4825
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3394.880000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 10332.xml