Minimizing eddy currents induced in the ground plane of a large phased-array ultrasound applicator for echo-planar imaging-based MR thermometry. Issue 1 (December 2016)
- Record Type:
- Journal Article
- Title:
- Minimizing eddy currents induced in the ground plane of a large phased-array ultrasound applicator for echo-planar imaging-based MR thermometry. Issue 1 (December 2016)
- Main Title:
- Minimizing eddy currents induced in the ground plane of a large phased-array ultrasound applicator for echo-planar imaging-based MR thermometry
- Authors:
- Lechner-Greite, Silke
Hehn, Nicolas
Werner, Beat
Zadicario, Eyal
Tarasek, Matthew
Yeo, Desmond - Abstract:
- Abstract Background The study aims to investigate different ground plane segmentation designs of an ultrasound transducer to reduce gradient field induced eddy currents and the associated geometric distortion and temperature map errors in echo-planar imaging (EPI)-based MR thermometry in transcranial magnetic resonance (MR)-guided focused ultrasound (tcMRgFUS). Methods Six different ground plane segmentations were considered and the efficacy of each in suppressing eddy currents was investigated in silico and in operando. For the latter case, the segmented ground planes were implemented in a transducer mockup model for validation. Robust spoiled gradient (SPGR) echo sequences and multi-shot EPI sequences were acquired. For each sequence and pattern, geometric distortions were quantified in the magnitude images and expressed in millimeters. Phase images were used for extracting the temperature maps on the basis of the temperature-dependent proton resonance frequency shift phenomenon. The means, standard deviations, and signal-to-noise ratios (SNRs) were extracted and contrasted with the geometric distortions of all patterns. Results The geometric distortion analysis and temperature map evaluations showed that more than one pattern could be considered the best-performing transducer. In the sagittal plane, the star (d) (3.46 ± 2.33 mm) and star-ring patterns (f) (2.72 ± 2.8 mm) showed smaller geometric distortions than the currently available seven-segment sheet (c)Abstract Background The study aims to investigate different ground plane segmentation designs of an ultrasound transducer to reduce gradient field induced eddy currents and the associated geometric distortion and temperature map errors in echo-planar imaging (EPI)-based MR thermometry in transcranial magnetic resonance (MR)-guided focused ultrasound (tcMRgFUS). Methods Six different ground plane segmentations were considered and the efficacy of each in suppressing eddy currents was investigated in silico and in operando. For the latter case, the segmented ground planes were implemented in a transducer mockup model for validation. Robust spoiled gradient (SPGR) echo sequences and multi-shot EPI sequences were acquired. For each sequence and pattern, geometric distortions were quantified in the magnitude images and expressed in millimeters. Phase images were used for extracting the temperature maps on the basis of the temperature-dependent proton resonance frequency shift phenomenon. The means, standard deviations, and signal-to-noise ratios (SNRs) were extracted and contrasted with the geometric distortions of all patterns. Results The geometric distortion analysis and temperature map evaluations showed that more than one pattern could be considered the best-performing transducer. In the sagittal plane, the star (d) (3.46 ± 2.33 mm) and star-ring patterns (f) (2.72 ± 2.8 mm) showed smaller geometric distortions than the currently available seven-segment sheet (c) (5.54 ± 4.21 mm) and were both comparable to the reference scenario (a) (2.77 ± 2.24 mm). Contrasting these results with the temperature maps revealed that (d) performs as well as (a) in SPGR and EPI. Conclusions We demonstrated that segmenting the transducer ground plane into a star pattern reduces eddy currents to a level wherein multi-plane EPI for accurate MR thermometry in tcMRgFUS is feasible. … (more)
- Is Part Of:
- Journal of therapeutic ultrasound. Volume 4:Issue 1(2016)
- Journal:
- Journal of therapeutic ultrasound
- Issue:
- Volume 4:Issue 1(2016)
- Issue Display:
- Volume 4, Issue 1 (2016)
- Year:
- 2016
- Volume:
- 4
- Issue:
- 1
- Issue Sort Value:
- 2016-0004-0001-0000
- Page Start:
- 1
- Page End:
- 14
- Publication Date:
- 2016-12
- Subjects:
- MR thermometry -- Focused ultrasound -- Eddy currents -- Proton resonance frequency shift -- Echo-planar imaging -- Phased-array transducer
Ultrasonic waves -- Therapeutic use -- Periodicals
Ultrasonics in medicine -- Periodicals
Periodicals
615.832305 - Journal URLs:
- http://www.jtultrasound.com/ ↗
http://www.jtultrasound.com/content ↗
http://link.springer.com/ ↗ - DOI:
- 10.1186/s40349-016-0047-x ↗
- Languages:
- English
- ISSNs:
- 2050-5736
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 10043.xml