Radiation dosimetry of 18F-FDG PET/CT: incorporating exam-specific parameters in dose estimates. Issue 1 (December 2016)
- Record Type:
- Journal Article
- Title:
- Radiation dosimetry of 18F-FDG PET/CT: incorporating exam-specific parameters in dose estimates. Issue 1 (December 2016)
- Main Title:
- Radiation dosimetry of 18F-FDG PET/CT: incorporating exam-specific parameters in dose estimates
- Authors:
- Quinn, Brian
Dauer, Zak
Pandit-Taskar, Neeta
Schoder, Heiko
Dauer, Lawrence - Abstract:
- Abstract Background Whole body fluorine-18 fluorodeoxyglucose positron emission tomography/computed tomography (PET/CT) is the standard of care in oncologic diagnosis and staging, and patient radiation dose must be well understood to balance exam benefits with the risk from radiation exposure. Although reference PET/CT patient doses are available, the potential for widely varying total dose prompts evaluation of clinic-specific patient dose. The aims of this study were to use exam-specific information to characterize the radiation dosimetry of PET/CT exams that used two different CT techniques for adult oncology patients and evaluate the practicality of employing an exam-specific approach to dose estimation. Methods Whole body PET/CT scans from two sets of consecutive adult patients were retrospectively reviewed. One set received a PET scan with a standard registration CT and the other a PET scan with a diagnostic quality CT. PET dose was calculated by modifying the standard reference phantoms in OLINDA/EXM 1.1 with patient-specific organ mass. CT dose was calculated using patient-specific data in ImPACT. International Commission on Radiological Protection publication 103 tissue weighting coefficients were used for effective dose. Results One hundred eighty three adult scans were evaluated (95 men, 88 women). The mean patient-specific effective dose from a mean injected 18F-FDG activity of 450 ± 32 MBq was 9.0 ± 1.6 mSv. For all standard PET/CT patients, mean effective mAsAbstract Background Whole body fluorine-18 fluorodeoxyglucose positron emission tomography/computed tomography (PET/CT) is the standard of care in oncologic diagnosis and staging, and patient radiation dose must be well understood to balance exam benefits with the risk from radiation exposure. Although reference PET/CT patient doses are available, the potential for widely varying total dose prompts evaluation of clinic-specific patient dose. The aims of this study were to use exam-specific information to characterize the radiation dosimetry of PET/CT exams that used two different CT techniques for adult oncology patients and evaluate the practicality of employing an exam-specific approach to dose estimation. Methods Whole body PET/CT scans from two sets of consecutive adult patients were retrospectively reviewed. One set received a PET scan with a standard registration CT and the other a PET scan with a diagnostic quality CT. PET dose was calculated by modifying the standard reference phantoms in OLINDA/EXM 1.1 with patient-specific organ mass. CT dose was calculated using patient-specific data in ImPACT. International Commission on Radiological Protection publication 103 tissue weighting coefficients were used for effective dose. Results One hundred eighty three adult scans were evaluated (95 men, 88 women). The mean patient-specific effective dose from a mean injected 18F-FDG activity of 450 ± 32 MBq was 9.0 ± 1.6 mSv. For all standard PET/CT patients, mean effective mAs was 39 ± 11 mAs, mean CT effective dose was 5.0 ± 1.0 mSv and mean total effective dose was 14 ± 1.3 mSv. For all diagnostic PET/CT patients, mean effective mAs was 120 ± 51 mAs, mean CT effective dose was 15.4 ± 5.0 mSv and mean total effective dose was 24.4 ± 4.3 mSv. The five organs receiving the highest organ equivalent doses in all exams were bladder, heart, brain, liver and lungs. Conclusions Patient-specific parameters optimize the patient dosimetry utilized in the medical justification of whole body PET/CT referrals and optimization of PET and CT acquisition parameters. Incorporating patient-specific data into dose estimates is a worthwhile effort for characterizing patient dose, and the specific dosimetric information assists in the justification of risk and optimization of PET/CT. … (more)
- Is Part Of:
- BMC medical imaging. Volume 16:Issue 1(2016)
- Journal:
- BMC medical imaging
- Issue:
- Volume 16:Issue 1(2016)
- Issue Display:
- Volume 16, Issue 1 (2016)
- Year:
- 2016
- Volume:
- 16
- Issue:
- 1
- Issue Sort Value:
- 2016-0016-0001-0000
- Page Start:
- 1
- Page End:
- 11
- Publication Date:
- 2016-12
- Subjects:
- PET/CT -- CT -- Radiation exposure -- Effective dose -- 18F-FDG
Diagnostic imaging -- Periodicals
616.075405 - Journal URLs:
- http://www.biomedcentral.com/bmcmedimaging/ ↗
http://www.pubmedcentral.nih.gov/tocrender.fcgi?journal=41 ↗
http://link.springer.com/ ↗ - DOI:
- 10.1186/s12880-016-0143-y ↗
- Languages:
- English
- ISSNs:
- 1471-2342
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 10034.xml