Feasibility of real-time MR thermal dose mapping for predicting radiofrequency ablation outcome in the myocardium in vivo. Issue 1 (December 2017)
- Record Type:
- Journal Article
- Title:
- Feasibility of real-time MR thermal dose mapping for predicting radiofrequency ablation outcome in the myocardium in vivo. Issue 1 (December 2017)
- Main Title:
- Feasibility of real-time MR thermal dose mapping for predicting radiofrequency ablation outcome in the myocardium in vivo
- Authors:
- Toupin, Solenn
Bour, Pierre
Lepetit-Coiffé, Matthieu
Ozenne, Valéry
Denis de Senneville, Baudouin
Schneider, Rainer
Vaussy, Alexis
Chaumeil, Arnaud
Cochet, Hubert
Sacher, Frédéric
Jaïs, Pierre
Quesson, Bruno - Abstract:
- Abstract Background Clinical treatment of cardiac arrhythmia by radiofrequency ablation (RFA) currently lacks quantitative and precise visualization of lesion formation in the myocardium during the procedure. This study aims at evaluating thermal dose (TD) imaging obtained from real-time magnetic resonance (MR) thermometry on the heart as a relevant indicator of the thermal lesion extent. Methods MR temperature mapping based on the Proton Resonance Frequency Shift (PRFS) method was performed at 1.5 T on the heart, with 4 to 5 slices acquired per heartbeat. Respiratory motion was compensated using navigator-based slice tracking. Residual in-plane motion and related magnetic susceptibility artifacts were corrected online. The standard deviation of temperature was measured on healthy volunteers (N = 5) in both ventricles. On animals, the MR-compatible catheter was positioned and visualized in the left ventricle (LV) using a bSSFP pulse sequence with active catheter tracking. Twelve MR-guided RFA were performed on three sheep in vivo at various locations in left ventricle (LV). The dimensions of the thermal lesions measured on thermal dose images, on 3D T1-weighted (T1-w) images acquired immediately after the ablation and at gross pathology were correlated. Results MR thermometry uncertainty was 1.5 °C on average over more than 96% of the pixels covering the left and right ventricles, on each volunteer. On animals, catheter repositioning in the LV with active slice tracking wasAbstract Background Clinical treatment of cardiac arrhythmia by radiofrequency ablation (RFA) currently lacks quantitative and precise visualization of lesion formation in the myocardium during the procedure. This study aims at evaluating thermal dose (TD) imaging obtained from real-time magnetic resonance (MR) thermometry on the heart as a relevant indicator of the thermal lesion extent. Methods MR temperature mapping based on the Proton Resonance Frequency Shift (PRFS) method was performed at 1.5 T on the heart, with 4 to 5 slices acquired per heartbeat. Respiratory motion was compensated using navigator-based slice tracking. Residual in-plane motion and related magnetic susceptibility artifacts were corrected online. The standard deviation of temperature was measured on healthy volunteers (N = 5) in both ventricles. On animals, the MR-compatible catheter was positioned and visualized in the left ventricle (LV) using a bSSFP pulse sequence with active catheter tracking. Twelve MR-guided RFA were performed on three sheep in vivo at various locations in left ventricle (LV). The dimensions of the thermal lesions measured on thermal dose images, on 3D T1-weighted (T1-w) images acquired immediately after the ablation and at gross pathology were correlated. Results MR thermometry uncertainty was 1.5 °C on average over more than 96% of the pixels covering the left and right ventricles, on each volunteer. On animals, catheter repositioning in the LV with active slice tracking was successfully performed and each ablation could be monitored in real-time by MR thermometry and thermal dosimetry. Thermal lesion dimensions on TD maps were found to be highly correlated with those observed on post-ablation T1-w images (R = 0.87) that also correlated (R = 0.89) with measurements at gross pathology. Conclusions Quantitative TD mapping from real-time rapid CMR thermometry during catheter-based RFA is feasible. It provides a direct assessment of the lesion extent in the myocardium with precision in the range of one millimeter. Real-time MR thermometry and thermal dosimetry may improve safety and efficacy of the RFA procedure by offering a reliable indicator of therapy outcome during the procedure. … (more)
- Is Part Of:
- Journal of cardiovascular magnetic resonance. Volume 19:Issue 1(2017)
- Journal:
- Journal of cardiovascular magnetic resonance
- Issue:
- Volume 19:Issue 1(2017)
- Issue Display:
- Volume 19, Issue 1 (2017)
- Year:
- 2017
- Volume:
- 19
- Issue:
- 1
- Issue Sort Value:
- 2017-0019-0001-0000
- Page Start:
- 1
- Page End:
- 12
- Publication Date:
- 2017-12
- Subjects:
- MR thermometry -- Radiofrequency ablation -- Catheter -- Arrhythmia -- Cardiac -- Electrophysiology
Cardiovascular system -- Magnetic resonance imaging -- Periodicals
616.1207548 - Journal URLs:
- http://jcmr-online.com/ ↗
http://www.informaworld.com/1532-429X ↗
http://www.tandfonline.com/ ↗
http://www.dekker.com/servlet/product/productid/JCMR ↗ - DOI:
- 10.1186/s12968-017-0323-0 ↗
- Languages:
- English
- ISSNs:
- 1097-6647
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4954.866600
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 10025.xml