Effects of heart valve prostheses on phase contrast flow measurements in Cardiovascular Magnetic Resonance – a phantom study. Issue 1 (December 2017)
- Record Type:
- Journal Article
- Title:
- Effects of heart valve prostheses on phase contrast flow measurements in Cardiovascular Magnetic Resonance – a phantom study. Issue 1 (December 2017)
- Main Title:
- Effects of heart valve prostheses on phase contrast flow measurements in Cardiovascular Magnetic Resonance – a phantom study
- Authors:
- Richau, Johanna
Dieringer, Matthias
Traber, Julius
von Knobelsdorff-Brenkenhoff, Florian
Greiser, Andreas
Schwenke, Carsten
Schulz-Menger, Jeanette - Abstract:
- Abstract Background Cardiovascular Magnetic Resonance is often used to evaluate patients after heart valve replacement. This study systematically analyses the influence of heart valve prostheses on phase contrast measurements in a phantom trial. Methods Two biological and one mechanical aortic valve prostheses were integrated in a flow phantom. B0 maps and phase contrast measurements were acquired at a 1.5 T MR scanner using conventional gradient-echo sequences in predefined distances to the prostheses. Results were compared to measurements with a synthetic metal-free aortic valve. Results The flow results at the level of the prosthesis differed significantly from the reference flow acquired before the level of the prosthesis. The maximum flow miscalculation was 154 ml/s for one of the biological prostheses and 140 ml/s for the mechanical prosthesis. Measurements with the synthetic aortic valve did not show significant deviations. Flow values measured approximately 20 mm distal to the level of the prosthesis agreed with the reference flow for all tested all prostheses. Conclusions The tested heart valve prostheses lead to a significant deviation of the measured flow rates compared to a reference. A distance of 20 mm was effective in our setting to avoid this influence.
- Is Part Of:
- Journal of cardiovascular magnetic resonance. Volume 19:Issue 1(2017)
- Journal:
- Journal of cardiovascular magnetic resonance
- Issue:
- Volume 19:Issue 1(2017)
- Issue Display:
- Volume 19, Issue 1 (2017)
- Year:
- 2017
- Volume:
- 19
- Issue:
- 1
- Issue Sort Value:
- 2017-0019-0001-0000
- Page Start:
- 1
- Page End:
- 11
- Publication Date:
- 2017-12
- Subjects:
- Cardiovascular magnetic resonance -- Phase contrast -- Flow phantom -- Valve prosthesis -- Artefacts
Cardiovascular system -- Magnetic resonance imaging -- Periodicals
616.1207548 - Journal URLs:
- http://jcmr-online.com/ ↗
http://www.informaworld.com/1532-429X ↗
http://www.tandfonline.com/ ↗
http://www.dekker.com/servlet/product/productid/JCMR ↗ - DOI:
- 10.1186/s12968-016-0319-1 ↗
- Languages:
- English
- ISSNs:
- 1097-6647
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4954.866600
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 10024.xml