New air speed calibration system at NMIJ for air speeds up to 90 m/s. (April 2019)
- Record Type:
- Journal Article
- Title:
- New air speed calibration system at NMIJ for air speeds up to 90 m/s. (April 2019)
- Main Title:
- New air speed calibration system at NMIJ for air speeds up to 90 m/s
- Authors:
- Iwai, Aya
Funaki, Tatsuya
Kurihara, Noboru
Terao, Yoshiya
Choi, Yong Moon - Abstract:
- Abstract: National Metrology Institute of Japan (NMIJ) has established a high air speed standard facility and has been providing a calibration service since 2015. The facility has an air speed range of 40 m/s to 90 m/s with a relative expanded uncertainty ( k = 2) of 0.63%. The purpose of this primary standard is mainly to contribute to the improvement of meteorological observation research and to the evaluation of flow field inside a turbo machine and around a high speed vehicle. The reference air speed is derived from the national primary gas flowrate standard of Japan. A conversion device from flowrate to air speed is installed in the test line of the closed-loop calibration facility. The reference air speed at the nozzle exit of the conversion device is obtained by comparing the integral of the air speed profiles and the reference volume flowrate. The total pressure tube used as a transfer standard is then calibrated against the reference air speed at the center of the nozzle exit. The Eiffel-type wind tunnel, which is a working standard for the daily calibration service, is calibrated using this total pressure tube. The present paper describes the calibration system, the traceability chain, and an uncertainty analysis using a validation method. Highlights: NMIJ has established a high air speed standard facility. The facilityhas an air speed range of 40 m/s to 90 m/s with a relative expanded uncertainty ( k = 2) of 0.63%. The reference air speed is derived from theAbstract: National Metrology Institute of Japan (NMIJ) has established a high air speed standard facility and has been providing a calibration service since 2015. The facility has an air speed range of 40 m/s to 90 m/s with a relative expanded uncertainty ( k = 2) of 0.63%. The purpose of this primary standard is mainly to contribute to the improvement of meteorological observation research and to the evaluation of flow field inside a turbo machine and around a high speed vehicle. The reference air speed is derived from the national primary gas flowrate standard of Japan. A conversion device from flowrate to air speed is installed in the test line of the closed-loop calibration facility. The reference air speed at the nozzle exit of the conversion device is obtained by comparing the integral of the air speed profiles and the reference volume flowrate. The total pressure tube used as a transfer standard is then calibrated against the reference air speed at the center of the nozzle exit. The Eiffel-type wind tunnel, which is a working standard for the daily calibration service, is calibrated using this total pressure tube. The present paper describes the calibration system, the traceability chain, and an uncertainty analysis using a validation method. Highlights: NMIJ has established a high air speed standard facility. The facilityhas an air speed range of 40 m/s to 90 m/s with a relative expanded uncertainty ( k = 2) of 0.63%. The reference air speed is derived from the national primary gas flowrate standard of Japan. The Pitot pressure tube iscalibrated using the Eiffel-type wind tunnelfor the daily calibration service. … (more)
- Is Part Of:
- Flow measurement and instrumentation. Volume 66(2019)
- Journal:
- Flow measurement and instrumentation
- Issue:
- Volume 66(2019)
- Issue Display:
- Volume 66, Issue 2019 (2019)
- Year:
- 2019
- Volume:
- 66
- Issue:
- 2019
- Issue Sort Value:
- 2019-0066-2019-0000
- Page Start:
- 132
- Page End:
- 140
- Publication Date:
- 2019-04
- Subjects:
- Air speed standard -- Wind tunnel -- Calibration -- Pitot static tube -- Hot-wire anemometer -- Volume flowrate
Fluid dynamic measurements -- Periodicals
Flow meters -- Periodicals
Fluides, Dynamique des -- Mesure -- Périodiques
Débitmètres -- Périodiques
681.2805 - Journal URLs:
- http://www.sciencedirect.com/science/journal/09555986 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.flowmeasinst.2019.02.004 ↗
- Languages:
- English
- ISSNs:
- 0955-5986
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3958.300000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 10031.xml