Cone beam computed tomography-based monitoring and management of target and organ motion during external beam radiotherapy in cervical cancer. (January 2019)
- Record Type:
- Journal Article
- Title:
- Cone beam computed tomography-based monitoring and management of target and organ motion during external beam radiotherapy in cervical cancer. (January 2019)
- Main Title:
- Cone beam computed tomography-based monitoring and management of target and organ motion during external beam radiotherapy in cervical cancer
- Authors:
- Jensen, Nina Boje Kibsgaard
Assenholt, Marianne Sanggaard
Fokdal, Lars Ulrik
Vestergaard, Anne
Schouboe, Annette
Kjaersgaard, Eva Bruun
Boejen, Annette
Nyvang, Lars
Lindegaard, Jacob Christian
Tanderup, Kari - Abstract:
- Abstract: Background and purpose: Organ motion is a challenge during high-precision external beam radiotherapy in cervical cancer, and improved strategies for treatment adaptation and monitoring of target dose coverage are needed. This study evaluates a cone beam computed tomography (CBCT)-based approach. Materials and methods: In twenty-three patients, individualized internal target volumes (ITVs) were generated from pre-treatment MRI and CT scans with full and empty bladders. The target volumes encompassed high-risk clinical target volume (CTV-T HR) (gross tumor volume + remaining cervix) and low risk (LR) CTV-T (CTV-T HR + uterus + parametriae + upper vagina). Volumetric Modulated Arc Therapy (VMAT) was used to deliver a dose of 45 Gy in 25 fractions. CBCTs were used for setup and for radiation therapists (RTTs) to evaluate the target coverage (inside/outside the planning target volume). CBCTs were reviewed offline. Estimates of the dose delivered with minimum (point) doses across all fractions to CTV-T HR (aim 42.75 Gy) and CTV-T LR (aim 40 Gy) were assessed. In patients with insufficient dose coverage, re-plans were generated based on previous imaging. Results: Median (range) of the ITV-margins (mean of anterior-posterior margins) related to uterus and cervix was 1.2 (0.5–2.2 and 1.0–2.1) cm. RTTs were able to assess the target coverage in 90% of all CBCTs (505/563). With re-planning, one patient had considerable benefit (12.7 Gy increase of minimum dose) to CTV-TAbstract: Background and purpose: Organ motion is a challenge during high-precision external beam radiotherapy in cervical cancer, and improved strategies for treatment adaptation and monitoring of target dose coverage are needed. This study evaluates a cone beam computed tomography (CBCT)-based approach. Materials and methods: In twenty-three patients, individualized internal target volumes (ITVs) were generated from pre-treatment MRI and CT scans with full and empty bladders. The target volumes encompassed high-risk clinical target volume (CTV-T HR) (gross tumor volume + remaining cervix) and low risk (LR) CTV-T (CTV-T HR + uterus + parametriae + upper vagina). Volumetric Modulated Arc Therapy (VMAT) was used to deliver a dose of 45 Gy in 25 fractions. CBCTs were used for setup and for radiation therapists (RTTs) to evaluate the target coverage (inside/outside the planning target volume). CBCTs were reviewed offline. Estimates of the dose delivered with minimum (point) doses across all fractions to CTV-T HR (aim 42.75 Gy) and CTV-T LR (aim 40 Gy) were assessed. In patients with insufficient dose coverage, re-plans were generated based on previous imaging. Results: Median (range) of the ITV-margins (mean of anterior-posterior margins) related to uterus and cervix was 1.2 (0.5–2.2 and 1.0–2.1) cm. RTTs were able to assess the target coverage in 90% of all CBCTs (505/563). With re-planning, one patient had considerable benefit (12.7 Gy increase of minimum dose) to CTV-T LR_vagina, four patients had improved dose to the CTV-T LR_uterus (1.2–1.8 Gy), and 3 patients did not benefit from re-planning. Conclusions: Daily CBCT-based monitoring of target coverage by the RTTs has proven safe with limited workload. It allows for reduction in the treated volumes without compromising the target dose coverage. … (more)
- Is Part Of:
- Physics and imaging in radiation oncology. Volume 9(2019)
- Journal:
- Physics and imaging in radiation oncology
- Issue:
- Volume 9(2019)
- Issue Display:
- Volume 9, Issue 2019 (2019)
- Year:
- 2019
- Volume:
- 9
- Issue:
- 2019
- Issue Sort Value:
- 2019-0009-2019-0000
- Page Start:
- 14
- Page End:
- 20
- Publication Date:
- 2019-01
- Subjects:
- Image guidance -- External beam radiotherapy -- Adaptive radiotherapy -- Cervical cancer -- Cone-beam computed tomography -- Interfraction motion
Radiotherapy -- Periodicals
Radiation dosimetry -- Periodicals
Cancer -- Imaging -- Periodicals
Oncology -- Periodicals
615.842 - Journal URLs:
- http://www.sciencedirect.com/ ↗
https://www.journals.elsevier.com/physics-and-imaging-in-radiation-oncology/ ↗ - DOI:
- 10.1016/j.phro.2018.12.002 ↗
- Languages:
- English
- ISSNs:
- 2405-6316
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 10012.xml