A randomized prospective controlled trial comparing the laryngeal tube suction disposable and the supreme laryngeal mask airway: the influence of head and neck position on oropharyngeal seal pressure. Issue 1 (December 2015)
- Record Type:
- Journal Article
- Title:
- A randomized prospective controlled trial comparing the laryngeal tube suction disposable and the supreme laryngeal mask airway: the influence of head and neck position on oropharyngeal seal pressure. Issue 1 (December 2015)
- Main Title:
- A randomized prospective controlled trial comparing the laryngeal tube suction disposable and the supreme laryngeal mask airway: the influence of head and neck position on oropharyngeal seal pressure
- Authors:
- Somri, Mostafa
Vaida, Sonia
Fornari, Gustavo
Mendoza, Gabriela
Charco-Mora, Pedro
Hawash, Naser
Matter, Ibrahim
Swaid, Forat
Gaitini, Luis - Abstract:
- Abstract Background The Laryngeal Tube Suction Disposable (LTS-D) and the Supreme Laryngeal Mask Airway (SLMA) are second generation supraglottic airway devices (SADs) with an added channel to allow gastric drainage. We studied the efficacy of these devices when using pressure controlled mechanical ventilation during general anesthesia for short and medium duration surgical procedures and compared the oropharyngeal seal pressure in different head and-neck positions. Methods Eighty patients in each group had either LTS-D or SLMA for airway management. The patients were recruited in two different institutions. Primary outcome variables were the oropharyngeal seal pressures in neutral, flexion, extension, right and left head-neck position. Secondary outcome variables were time to achieve an effective airway, ease of insertion, number of attempts, maneuvers necessary during insertion, ventilatory parameters, success of gastric tube insertion and incidence of complications. Results The oropharyngeal seal pressure achieved with the LTS-D was higher than the SLMA in, (extension (p =0.0150) and right position (p =0.0268 at 60 cm H2 O intracuff pressures and nearly significant in neutral position (p = 0.0571). The oropharyngeal seal pressure was significantly higher with the LTS-D during neck extension as compared to SLMA (p = 0.015). Similar oropharyngeal seal pressures were detected in all other positions with each device. The secondary outcomes were comparable between bothAbstract Background The Laryngeal Tube Suction Disposable (LTS-D) and the Supreme Laryngeal Mask Airway (SLMA) are second generation supraglottic airway devices (SADs) with an added channel to allow gastric drainage. We studied the efficacy of these devices when using pressure controlled mechanical ventilation during general anesthesia for short and medium duration surgical procedures and compared the oropharyngeal seal pressure in different head and-neck positions. Methods Eighty patients in each group had either LTS-D or SLMA for airway management. The patients were recruited in two different institutions. Primary outcome variables were the oropharyngeal seal pressures in neutral, flexion, extension, right and left head-neck position. Secondary outcome variables were time to achieve an effective airway, ease of insertion, number of attempts, maneuvers necessary during insertion, ventilatory parameters, success of gastric tube insertion and incidence of complications. Results The oropharyngeal seal pressure achieved with the LTS-D was higher than the SLMA in, (extension (p =0.0150) and right position (p =0.0268 at 60 cm H2 O intracuff pressures and nearly significant in neutral position (p = 0.0571). The oropharyngeal seal pressure was significantly higher with the LTS-D during neck extension as compared to SLMA (p = 0.015). Similar oropharyngeal seal pressures were detected in all other positions with each device. The secondary outcomes were comparable between both groups. Patients ventilated with LTS-D had higher incidence of sore throat (p = 0.527). No major complications occurred. Conclusions Better oropharyngeal seal pressure was achieved with the LTS-D in head-neck right and extension positions , although it did not appear to have significance in alteration of management using pressure control mechanical ventilation in neutral position. The fiberoptic view was better with the SLMA. The post-operative sore throat incidence was higher in the LTS-D. Trial registration ClinicalTrials.gov ID:NCT02856672, Unique Protocol ID:BnaiZionMC-16-LG-001, Registered: August 2016. … (more)
- Is Part Of:
- BMC anesthesiology. Volume 16:Issue 1(2016)
- Journal:
- BMC anesthesiology
- Issue:
- Volume 16:Issue 1(2016)
- Issue Display:
- Volume 16, Issue 1 (2016)
- Year:
- 2016
- Volume:
- 16
- Issue:
- 1
- Issue Sort Value:
- 2016-0016-0001-0000
- Page Start:
- 1
- Page End:
- 8
- Publication Date:
- 2015-12
- Subjects:
- Laryngeal tube -- Supreme laryngeal mask airway -- Oropharyngeal seal pressure
Anesthetics -- Periodicals
Anesthesia -- Periodicals
Anesthesiology -- Periodicals
617.9605 - Journal URLs:
- http://www.biomedcentral.com/bmcanesthesiol/ ↗
http://www.pubmedcentral.nih.gov/tocrender.fcgi?journal=11 ↗
http://link.springer.com/ ↗ - DOI:
- 10.1186/s12871-016-0237-7 ↗
- Languages:
- English
- ISSNs:
- 1471-2253
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - Digital store
British Library HMNTS - ELD Digital store - Ingest File:
- 9973.xml