Detector dose response in megavoltage small photon beams. I. Theoretical concepts. Issue 10 (25th September 2015)
- Record Type:
- Journal Article
- Title:
- Detector dose response in megavoltage small photon beams. I. Theoretical concepts. Issue 10 (25th September 2015)
- Main Title:
- Detector dose response in megavoltage small photon beams. I. Theoretical concepts
- Authors:
- Bouchard, Hugo
Seuntjens, Jan
Duane, Simon
Kamio, Yuji
Palmans, Hugo - Abstract:
- Abstract : Purpose: To explain the reasons for significant quality correction factors in megavoltage small photon fields and clarify the underlying concepts relevant to dosimetry under such conditions. Methods: The validity of cavity theory and the requirement of charged particle equilibrium (CPE) are addressed from a theoretical point of view in the context of nonstandard beams. Perturbation effects are described into four main subeffects, explaining their nature and pointing out their relative importance in small photon fields. Results: It is demonstrated that the failure to meet classical cavity theory requirements, such as CPE, is not the reason for significant quality correction factors. On the contrary, it is shown that the lack of CPE alone cannot explain these corrections and that what matters most, apart from volume averaging effects, is the relationship between the lack of CPE in the small field itself and the density of the detector cavity. The density perturbation effect is explained based on Fano's theorem, describing the compensating effect of two main contributions to cavity absorbed dose. Using the same approach, perturbation effects arising from the difference in atomic properties of the cavity medium and the presence of extracameral components are explained. Volume averaging effects are also discussed in detail. Conclusions: Quality correction factors of small megavoltage photon fields are mainly due to differences in electron density between water and theAbstract : Purpose: To explain the reasons for significant quality correction factors in megavoltage small photon fields and clarify the underlying concepts relevant to dosimetry under such conditions. Methods: The validity of cavity theory and the requirement of charged particle equilibrium (CPE) are addressed from a theoretical point of view in the context of nonstandard beams. Perturbation effects are described into four main subeffects, explaining their nature and pointing out their relative importance in small photon fields. Results: It is demonstrated that the failure to meet classical cavity theory requirements, such as CPE, is not the reason for significant quality correction factors. On the contrary, it is shown that the lack of CPE alone cannot explain these corrections and that what matters most, apart from volume averaging effects, is the relationship between the lack of CPE in the small field itself and the density of the detector cavity. The density perturbation effect is explained based on Fano's theorem, describing the compensating effect of two main contributions to cavity absorbed dose. Using the same approach, perturbation effects arising from the difference in atomic properties of the cavity medium and the presence of extracameral components are explained. Volume averaging effects are also discussed in detail. Conclusions: Quality correction factors of small megavoltage photon fields are mainly due to differences in electron density between water and the detector medium and to volume averaging over the detector cavity. Other effects, such as the presence of extracameral components and differences in atomic properties of the detection medium with respect to water, can also play an accentuated role in small photon fields compared to standard beams. … (more)
- Is Part Of:
- Medical physics. Volume 42:Issue 10(2015)
- Journal:
- Medical physics
- Issue:
- Volume 42:Issue 10(2015)
- Issue Display:
- Volume 42, Issue 10 (2015)
- Year:
- 2015
- Volume:
- 42
- Issue:
- 10
- Issue Sort Value:
- 2015-0042-0010-0000
- Page Start:
- 6033
- Page End:
- 6047
- Publication Date:
- 2015-09-25
- Subjects:
- dosimetry -- particle detectors -- radiation therapy
Dosimetry/exposure assessment
Radiation therapy -- Scintigraphy
beam calibration -- cavity theory -- Monte Carlo simulations -- nonstandard beams -- perturbation effects -- photon beams -- quality correction factors -- radiation detectors -- radiation dosimetry -- small fields
Photons -- Dosimetry -- Particle beam detectors -- General atomic properties -- Collisional energy loss -- Perturbation theory -- Monte Carlo methods -- Radiation detectors -- Classical field theory -- Cavitation
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4930053 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9939.xml