Integrated PET/MR imaging: Automatic attenuation correction of flexible RF coils. Issue 8 (9th July 2013)
- Record Type:
- Journal Article
- Title:
- Integrated PET/MR imaging: Automatic attenuation correction of flexible RF coils. Issue 8 (9th July 2013)
- Main Title:
- Integrated PET/MR imaging: Automatic attenuation correction of flexible RF coils
- Authors:
- Kartmann, René
Paulus, Daniel H.
Braun, Harald
Aklan, Bassim
Ziegler, Susanne
Navalpakkam, Bharath K.
Lentschig, Markus
Quick, Harald H. - Abstract:
- Abstract : Purpose: : Flexible radiofrequency (RF) surface coils used in simultaneous PET/MR imaging are currently disregarded in PET attenuation correction (AC) since their position and individual geometry are unknown in whole‐body patient scans. The attenuation of PET emission data due to the presence of RF surface coils has been investigated by several research groups but so far no automatic approach for the incorporation of RF surface coils into PET AC has been described. In this work, an algorithm is presented and evaluated which automatically determines the position of multiple RF surface coils and corrects for their attenuation of the PET emission data. Methods: : The presented algorithm nonrigidly registers pre‐acquired CT‐based three‐dimensional attenuation templates of RF surface coils into attenuation maps used for PET AC. Transformation parameters are obtained by nonrigid B‐spline landmark registration of marker positions in the CT‐based attenuation templates of the RF surface coils to marker positions in the current MR images of the patient. The use of different marker patterns enables the registration algorithm to distinguish multiple partly overlapping RF surface coils. To evaluate the registration algorithm, two different PET emission scans of a NEMA standard body phantom with six active lesions and of a large rectangular body phantom were performed on an integrated whole‐body PET/MR scanner. The phantoms were scanned with and without one (NEMA phantom scan)Abstract : Purpose: : Flexible radiofrequency (RF) surface coils used in simultaneous PET/MR imaging are currently disregarded in PET attenuation correction (AC) since their position and individual geometry are unknown in whole‐body patient scans. The attenuation of PET emission data due to the presence of RF surface coils has been investigated by several research groups but so far no automatic approach for the incorporation of RF surface coils into PET AC has been described. In this work, an algorithm is presented and evaluated which automatically determines the position of multiple RF surface coils and corrects for their attenuation of the PET emission data. Methods: : The presented algorithm nonrigidly registers pre‐acquired CT‐based three‐dimensional attenuation templates of RF surface coils into attenuation maps used for PET AC. Transformation parameters are obtained by nonrigid B‐spline landmark registration of marker positions in the CT‐based attenuation templates of the RF surface coils to marker positions in the current MR images of the patient. The use of different marker patterns enables the registration algorithm to distinguish multiple partly overlapping RF surface coils. To evaluate the registration algorithm, two different PET emission scans of a NEMA standard body phantom with six active lesions and of a large rectangular body phantom were performed on an integrated whole‐body PET/MR scanner. The phantoms were scanned with and without one (NEMA phantom scan) or three (large body phantom scan) flexible six‐channel RF surface coils placed on top. Additionally, the accuracy and performance of the algorithm were evaluated on volunteer scans ( n = 5) and on a patient scan using a typical clinical setup of three RF surface coils. Results: : Overall loss of true counts due to the presence of the RF surface coils was 5.1% for the NEMA phantom, 3.6% for the large body phantom, and 2.1% for the patient scan. Considerable local underestimation of measured activity concentration up to 15.4% in the top part of the phantoms and 15.5% for a lesion near the body surface of the patient was measured close to the high attenuating hardware components of the RF coils. The attenuation maps generated by the registration algorithm reduced the quantification errors due to the RF surface coils to values ranging from −3.9% to 4.3%. Concerning the volunteer examinations, the attenuation templates of the three RF surface coils were registered to their correct positions with an overall accuracy of about 3 mm. Conclusions: : The presence of flexible RF surface coils leads to considerable local errors in the simultaneously measured PET activity concentration up to 15.5% especially in regions close to the coils. The presented automatic algorithm accurately and reliably reduces the PET quantification errors caused by multiple partly overlapping flexible RF surface coils to values of 4.3% or better. … (more)
- Is Part Of:
- Medical physics. Volume 40:Issue 8(2013)
- Journal:
- Medical physics
- Issue:
- Volume 40:Issue 8(2013)
- Issue Display:
- Volume 40, Issue 8 (2013)
- Year:
- 2013
- Volume:
- 40
- Issue:
- 8
- Issue Sort Value:
- 2013-0040-0008-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2013-07-09
- Subjects:
- Positron emission tomography (PET) -- Magnetic resonance imaging
biomedical MRI -- coils -- image registration -- medical image processing -- phantoms -- positron emission tomography
simultaneous whole‐body PET/MR -- attenuation correction (AC) -- radiofrequency (RF) surface coil -- B‐spline registration -- hybrid imaging
Involving electronic [emr] or nuclear [nmr] magnetic resonance, e.g. magnetic resonance imaging -- Digital computing or data processing equipment or methods, specially adapted for specific applications -- Image data processing or generation, in general -- Coils -- Coils; Windings; Conductive connections
Positron emission tomography -- Medical imaging -- Medical magnetic resonance imaging -- Computed tomography -- Medical image reconstruction -- Tissues -- Liver -- Image reconstruction -- Foams -- Surface patterning
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4812685 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9932.xml