A study of matching fluid loss in a biomedical microwave tomography system. Issue 2 (25th January 2013)
- Record Type:
- Journal Article
- Title:
- A study of matching fluid loss in a biomedical microwave tomography system. Issue 2 (25th January 2013)
- Main Title:
- A study of matching fluid loss in a biomedical microwave tomography system
- Authors:
- Gilmore, Colin
Zakaria, Amer
LoVetri, Joe
Pistorius, Stephen - Abstract:
- Abstract : Purpose: : Effective imaging of human tissue with microwave tomography systems requires a matching fluid to reduce the wave reflections at the tissue boundary. Further, in order to match the idealized mathematical model used for imaging with the complicated physical measurement environment, loss must be added to the matching fluid. Both too little and too much loss result in low‐quality images, but due to the nonlinear nature of the imaging problem, the exact nature of loss‐to‐image quality cannot be predicted a priori . Possible optimal loss levels include a single, highly sensitive value, or a broad range of acceptable losses. Herein, the authors outline a process of determining an appropriate level of loss inside the matching fluid and attempt to determine the bounds for which the images are the highest quality. Methods: : Our biomedical microwave tomography system is designed for 2D limb imaging, operating from 0.8 to 1.2 GHz. Our matching fluid consists of deionized water with various levels of loss introduced by the addition of table salt. Using two homogeneous tissue‐mimicking phantoms, and eight different matching fluids of varying salt concentrations, the authors introduce quantitative image quality metrics based on L ‐norms, mean values, and standard deviations to test the tomography system and assess image quality. Images are generated with a balanced multiplicative regularized contrast source inversion algorithm. The authors further generate images ofAbstract : Purpose: : Effective imaging of human tissue with microwave tomography systems requires a matching fluid to reduce the wave reflections at the tissue boundary. Further, in order to match the idealized mathematical model used for imaging with the complicated physical measurement environment, loss must be added to the matching fluid. Both too little and too much loss result in low‐quality images, but due to the nonlinear nature of the imaging problem, the exact nature of loss‐to‐image quality cannot be predicted a priori . Possible optimal loss levels include a single, highly sensitive value, or a broad range of acceptable losses. Herein, the authors outline a process of determining an appropriate level of loss inside the matching fluid and attempt to determine the bounds for which the images are the highest quality. Methods: : Our biomedical microwave tomography system is designed for 2D limb imaging, operating from 0.8 to 1.2 GHz. Our matching fluid consists of deionized water with various levels of loss introduced by the addition of table salt. Using two homogeneous tissue‐mimicking phantoms, and eight different matching fluids of varying salt concentrations, the authors introduce quantitative image quality metrics based on L ‐norms, mean values, and standard deviations to test the tomography system and assess image quality. Images are generated with a balanced multiplicative regularized contrast source inversion algorithm. The authors further generate images of a human forearm which may be analyzed qualitatively. Results: : The image metrics for the phantoms support the claim that the worst images occur at the extremes of high and low salt concentrations. Importantly, the image metrics show that there exists a broad range of salt concentrations that result in high‐quality images, not a single optimal value. In particular, 2.5–4.5 g of table salt per liter of deionized water provide the best reconstruction quality for simple phantoms. The authors argue that qualitatively, the human forearm data provide the best images at approximately the same salt concentrations. Conclusions: : There exists a relatively large‐range of matching fluid losses (i.e., salt concentrations) that provide similar image quality. In particular, it is not necessary to spend time highly optimizing the level of loss in the matching fluid. … (more)
- Is Part Of:
- Medical physics. Volume 40:Issue 2(2013)
- Journal:
- Medical physics
- Issue:
- Volume 40:Issue 2(2013)
- Issue Display:
- Volume 40, Issue 2 (2013)
- Year:
- 2013
- Volume:
- 40
- Issue:
- 2
- Issue Sort Value:
- 2013-0040-0002-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2013-01-25
- Subjects:
- Non‐ionizing radiation equipment and techniques -- Reconstruction
biological tissues -- image reconstruction -- inverse problems -- medical image processing -- microwave imaging -- phantoms -- tomography
microwave imaging -- electromagnetic tomography
Digital computing or data processing equipment or methods, specially adapted for specific applications -- Image data processing or generation, in general
Medical imaging -- Permittivity -- Medical image quality -- Image reconstruction -- Antennas -- Tissues -- Medical image reconstruction -- Medical image contrast -- Microwaves -- Data sets
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4788640 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9924.xml