Measuring energy expenditure in the intensive care unit: a comparison of indirect calorimetry by E-sCOVX and Quark RMR with Deltatrac II in mechanically ventilated critically ill patients. Issue 1 (December 2016)
- Record Type:
- Journal Article
- Title:
- Measuring energy expenditure in the intensive care unit: a comparison of indirect calorimetry by E-sCOVX and Quark RMR with Deltatrac II in mechanically ventilated critically ill patients. Issue 1 (December 2016)
- Main Title:
- Measuring energy expenditure in the intensive care unit: a comparison of indirect calorimetry by E-sCOVX and Quark RMR with Deltatrac II in mechanically ventilated critically ill patients
- Authors:
- Rehal, Martin
Fiskaare, Erik
Tjäder, Inga
Norberg, Åke
Rooyackers, Olav
Wernerman, Jan - Abstract:
- Abstract Background Indirect calorimetry allows the determination of energy expenditure in critically ill patients by measuring oxygen consumption (VO2 ) and carbon dioxide production (VCO2 ). Recent studies have demonstrated variable performance of "breath-by-breath" instruments compared to mixing chamber technology. The aim of this study was to validate two modern devices (E-sCOVX and Quark RMR) against a reference method (Deltatrac II). Method Measurements of VO2 /VCO2 with the test and reference devices were performed simultaneously over a 20-min period in mechanically ventilated adult intensive care unit patients. Accuracy and precision of instruments were analyzed using Bland-Altman plots. Results Forty-eight measurements in 22 patients were included for analysis. Both E-sCOVX and Quark RMR overestimated VO2 and VCO2 compared to Deltatrac II, corresponding to a 10 % higher mean resting energy expenditure. Limits of agreement of resting energy expenditure within ±2 standard deviations were ±461 kcal/24 h (±21 % expressed as percentage error) for ΔE-sCOVX–Deltatrac II and ±465 kcal/24 h (±22 %) for ΔQuark RMR–Deltatrac II. Conclusion Both test devices overestimate VO2 and VCO2 compared to Deltatrac II. The observed limits of agreement are comparable to those commonly accepted in evaluations of circulatory monitoring, and significantly less than results from predictive equations. We hypothesize that the discrepancy between methods is due to patient/ventilator-relatedAbstract Background Indirect calorimetry allows the determination of energy expenditure in critically ill patients by measuring oxygen consumption (VO2 ) and carbon dioxide production (VCO2 ). Recent studies have demonstrated variable performance of "breath-by-breath" instruments compared to mixing chamber technology. The aim of this study was to validate two modern devices (E-sCOVX and Quark RMR) against a reference method (Deltatrac II). Method Measurements of VO2 /VCO2 with the test and reference devices were performed simultaneously over a 20-min period in mechanically ventilated adult intensive care unit patients. Accuracy and precision of instruments were analyzed using Bland-Altman plots. Results Forty-eight measurements in 22 patients were included for analysis. Both E-sCOVX and Quark RMR overestimated VO2 and VCO2 compared to Deltatrac II, corresponding to a 10 % higher mean resting energy expenditure. Limits of agreement of resting energy expenditure within ±2 standard deviations were ±461 kcal/24 h (±21 % expressed as percentage error) for ΔE-sCOVX–Deltatrac II and ±465 kcal/24 h (±22 %) for ΔQuark RMR–Deltatrac II. Conclusion Both test devices overestimate VO2 and VCO2 compared to Deltatrac II. The observed limits of agreement are comparable to those commonly accepted in evaluations of circulatory monitoring, and significantly less than results from predictive equations. We hypothesize that the discrepancy between methods is due to patient/ventilator-related factors that affect the synchronization of gas and spirometry waveforms. Trial registration Australian New Zealand Clinical Trials Registry, Trial IDACTRN12615000205538 . Date registered 3 March 2015. … (more)
- Is Part Of:
- Critical care. Volume 20:Issue 1(2016)
- Journal:
- Critical care
- Issue:
- Volume 20:Issue 1(2016)
- Issue Display:
- Volume 20, Issue 1 (2016)
- Year:
- 2016
- Volume:
- 20
- Issue:
- 1
- Issue Sort Value:
- 2016-0020-0001-0000
- Page Start:
- 1
- Page End:
- 9
- Publication Date:
- 2016-12
- Subjects:
- Indirect calorimetry -- Mechanical ventilation -- Intensive care -- Validation
Critical care medicine -- Periodicals
616.02805 - Journal URLs:
- http://ccforum.com/currentissue/browse.asp ↗
http://www.biomedcentral.com/1364-8535/ ↗
http://www.pubmedcentral.nih.gov/tocrender.fcgi?action=archive&journal=9 ↗
http://link.springer.com/ ↗ - DOI:
- 10.1186/s13054-016-1232-6 ↗
- Languages:
- English
- ISSNs:
- 1364-8535
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9919.xml