Multicenter Evaluation of the Thermo Scientific Prelude for Measurement of Immunosuppressant Drugs Using Sample Preparation Liquid Chromatography–Tandem Mass Spectrometry. Issue 2 (April 2015)
- Record Type:
- Journal Article
- Title:
- Multicenter Evaluation of the Thermo Scientific Prelude for Measurement of Immunosuppressant Drugs Using Sample Preparation Liquid Chromatography–Tandem Mass Spectrometry. Issue 2 (April 2015)
- Main Title:
- Multicenter Evaluation of the Thermo Scientific Prelude for Measurement of Immunosuppressant Drugs Using Sample Preparation Liquid Chromatography–Tandem Mass Spectrometry
- Authors:
- Peake, Roy W. A.
Hartigan, Christina R.
Esposito, Christopher L.
Kellogg, Mark D.
Gabler, Jessica
Wang, Sihe
Breaud, Autumn
Di Bussolo, Joseph
Moskowitz, Joshua
Clarke, William - Abstract:
- Abstract : Abstract: Immunosuppressant drugs (ISDs) are commonly prescribed to solid organ transplant patients. Their narrow therapeutic index and potential for toxicity necessitates careful monitoring of blood concentrations. Liquid chromatography–tandem mass spectrometry (LC-MS/MS) methods are increasingly used for ISD measurement. However, there remain many challenges with this methodology, particularly regarding interassay variability. The Thermo Scientific Prelude is an online extraction/liquid chromatography platform that uses turbulent flow technology coupled with MS/MS. A multicenter evaluation of the Prelude for the measurement of cyclosporine A, tacrolimus, and sirolimus is described. ISDs were measured at each site using standardized protocols. Sample preparation liquid chromatography–MS/MS was performed using the Prelude coupled to a TSQ Vantage. Chromatography was achieved with a Cyclone-P TurboFlow/Accucore C8 column combination using a multisolvent loading and eluting pump system. Mass spectrometry acquisitions were performed in selective reaction monitoring mode and data processed using TraceFinder (version 3.1). Multisite mean imprecision for cyclosporine A ranged from 8.8% (54 mcg/L) to 9.8% (450 mcg/L); for tacrolimus, 4.7% (15.5 mcg/L) to 12.6% (2.5 mcg/L); for sirolimus, 7.4% (19.9 mcg/L) to 16.5% (2.6 mcg/L). Approximately 110 specimens were used for method comparison. For cyclosporine A, mean bias against the multisite mean ranged from −18% to 1%; forAbstract : Abstract: Immunosuppressant drugs (ISDs) are commonly prescribed to solid organ transplant patients. Their narrow therapeutic index and potential for toxicity necessitates careful monitoring of blood concentrations. Liquid chromatography–tandem mass spectrometry (LC-MS/MS) methods are increasingly used for ISD measurement. However, there remain many challenges with this methodology, particularly regarding interassay variability. The Thermo Scientific Prelude is an online extraction/liquid chromatography platform that uses turbulent flow technology coupled with MS/MS. A multicenter evaluation of the Prelude for the measurement of cyclosporine A, tacrolimus, and sirolimus is described. ISDs were measured at each site using standardized protocols. Sample preparation liquid chromatography–MS/MS was performed using the Prelude coupled to a TSQ Vantage. Chromatography was achieved with a Cyclone-P TurboFlow/Accucore C8 column combination using a multisolvent loading and eluting pump system. Mass spectrometry acquisitions were performed in selective reaction monitoring mode and data processed using TraceFinder (version 3.1). Multisite mean imprecision for cyclosporine A ranged from 8.8% (54 mcg/L) to 9.8% (450 mcg/L); for tacrolimus, 4.7% (15.5 mcg/L) to 12.6% (2.5 mcg/L); for sirolimus, 7.4% (19.9 mcg/L) to 16.5% (2.6 mcg/L). Approximately 110 specimens were used for method comparison. For cyclosporine A, mean bias against the multisite mean ranged from −18% to 1%; for tacrolimus, values ranged from −7% to 4%; for sirolimus, values ranged from −4% to 2%. Comparisons of multisite mean Prelude results with routine ISD method results was also performed for cyclosporine A (slope = 0.7878, intercept = 24.16, r = 0.98), tacrolimus (slope = 0.9391, intercept = 0.1017, r = 98), and sirolimus (slope = 0.9618, intercept = 0.1483, r = 0.97). The Prelude ISD method offers acceptable and comparable multisite performance. This study has also highlighted the importance of adopting standardized protocols and LC-MS/MS methods for better comparability between ISD assays. … (more)
- Is Part Of:
- Therapeutic drug monitoring. Volume 37:Issue 2(2015:Apr.)
- Journal:
- Therapeutic drug monitoring
- Issue:
- Volume 37:Issue 2(2015:Apr.)
- Issue Display:
- Volume 37, Issue 2 (2015)
- Year:
- 2015
- Volume:
- 37
- Issue:
- 2
- Issue Sort Value:
- 2015-0037-0002-0000
- Page Start:
- Page End:
- Publication Date:
- 2015-04
- Subjects:
- immunosuppressants -- LC-MS/MS -- Prelude -- standardization -- comparison
Pharmacokinetics -- Periodicals
Patient monitoring -- Periodicals
Drugs -- Analysis -- Periodicals
Body fluids -- Analysis -- Periodicals
Drug Therapy -- Periodicals
Monitoring, Physiologic -- Periodicals
Pharmacology -- Periodicals
615.7 - Journal URLs:
- http://journals.lww.com/drug-monitoring/pages/default.aspx ↗
http://ovidsp.ovid.com/ovidweb.cgi?T=JS&NEWS=n&CSC=Y&PAGE=toc&D=yrovft&AN=00007691-000000000-00000 ↗
http://www.drug-monitoring.com/ ↗
http://journals.lww.com ↗
http://www.lww.com/Product/0163-4356 ↗ - DOI:
- 10.1097/FTD.0000000000000120 ↗
- Languages:
- English
- ISSNs:
- 0163-4356
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 8814.643000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9894.xml