Coagulation and ablation patterns of high-intensity focused ultrasound on a tissue-mimicking phantom and cadaveric skin. Issue 9 (December 2015)
- Record Type:
- Journal Article
- Title:
- Coagulation and ablation patterns of high-intensity focused ultrasound on a tissue-mimicking phantom and cadaveric skin. Issue 9 (December 2015)
- Main Title:
- Coagulation and ablation patterns of high-intensity focused ultrasound on a tissue-mimicking phantom and cadaveric skin
- Authors:
- Kim, Hee-Jin
Kim, Han
Zheng, Zhenlong
Park, Hyoun
Yoon, Jeung
Oh, Wook
Lee, Cheol
Cho, Sung - Abstract:
- Abstract High-intensity focused ultrasound (HIFU) can be applied noninvasively to create focused zones of tissue coagulation on various skin layers. We performed a comparative study of HIFU, evaluating patterns of focused tissue coagulation and ablation upon application thereof. A tissue-mimicking (TM) phantom was prepared with bovine serum albumin and polyacrylamide hydrogel to evaluate the geometric patterns of HIFU-induced thermal injury zones (TIZs) for five different HIFU devices. Additionally, for each device, we investigated histologic patterns of HIFU-induced coagulation and ablation in serial sections of cadaveric skin of the face and neck. All HIFU devices generated remarkable TIZs in the TM phantom, with different geometric values of coagulation for each device. Most of the TIZs seemed to be separated into two or more tiny parts. In cadaveric skin, characteristic patterns of HIFU-induced ablation and coagulation were noted along the mid to lower dermis at the focal penetration depth of 3 mm and along subcutaneous fat to the superficial musculoaponeurotic system or the platysma muscle of the neck at 4.5 mm. Additionally, remarkable pre-focal areas of tissue coagulation were observed in the upper and mid dermis at the focal penetration depth of 3 mm and mid to lower dermis at 4.5 mm. For five HIFU devices, we outlined various patterns of HIFU-induced TIZ formation along pre-focal, focal, and post-focal areas of TM phantom and cadaveric skin of the face and neck.
- Is Part Of:
- Lasers in medical science. Volume 30:Issue 9(2015)
- Journal:
- Lasers in medical science
- Issue:
- Volume 30:Issue 9(2015)
- Issue Display:
- Volume 30, Issue 9 (2015)
- Year:
- 2015
- Volume:
- 30
- Issue:
- 9
- Issue Sort Value:
- 2015-0030-0009-0000
- Page Start:
- 2251
- Page End:
- 2258
- Publication Date:
- 2015-12
- Subjects:
- High-intensity focused ultrasound -- Coagulation -- Ablation -- Tissue phantom -- Cadaver -- Comparison study
Lasers in medicine -- Periodicals
610.28 - Journal URLs:
- http://firstsearch.oclc.org ↗
http://link.springer.com/journal/10103 ↗
http://www.springer.com/gb/ ↗
http://www.springer.com/gb/ ↗
http://link.springer.de/link/service/journals/10103/index.htm ↗ - DOI:
- 10.1007/s10103-015-1804-8 ↗
- Languages:
- English
- ISSNs:
- 0268-8921
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5156.680800
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9899.xml