Distributive quality assurance and delivery of stereotactic ablative radiotherapy treatments amongst beam matched linear accelerators: A feasibility study. Issue 4 (18th March 2019)
- Record Type:
- Journal Article
- Title:
- Distributive quality assurance and delivery of stereotactic ablative radiotherapy treatments amongst beam matched linear accelerators: A feasibility study. Issue 4 (18th March 2019)
- Main Title:
- Distributive quality assurance and delivery of stereotactic ablative radiotherapy treatments amongst beam matched linear accelerators: A feasibility study
- Authors:
- Rijken, James
Schachenmayr, Hilmar
Crowe, Scott
Kairn, Tanya
Trapp, Jamie - Abstract:
- Abstract: Purpose: Beam matching occurs on all linacs to some degree and when two are more are matched to each other, patients are able to be transferred between machines. Quality assurance of plans can also be performed "distributively" on any of the matched linacs. The degree to which machines are matched and how this translates to like delivery of plans has been the focus of a number of studies. This concept has not yet been explored for stereotactic techniques which require a higher degree of accuracy. This study proposes beam matching criteria which allows for the distributive delivery and quality assurance of stereotactic body radiotherapy (SBRT) plans. Method: Two clinically relevant and complex volumetric modulated arc therapy (VMAT) SBRT spine and lung plans were chosen as benchmarking cases. These were delivered on nine previously beam matched linacs with quality assurance performed through ArcCheck and film exposure in the sagittal plane. Measured doses were compared to their treatment planning system predictions through gamma analysis at a range of criteria. Results: Despite differences in beam match parameters and variations in small fields, all nine linacs produced accurate deliveries with a tight deviation in the population sample. Pass rates were well above suggested tolerances at the recommended gamma criterion. Film was able to detect dose errors to a greater degree than ArcCheck. Conclusion: Distributive quality assurance and delivery of stereotacticAbstract: Purpose: Beam matching occurs on all linacs to some degree and when two are more are matched to each other, patients are able to be transferred between machines. Quality assurance of plans can also be performed "distributively" on any of the matched linacs. The degree to which machines are matched and how this translates to like delivery of plans has been the focus of a number of studies. This concept has not yet been explored for stereotactic techniques which require a higher degree of accuracy. This study proposes beam matching criteria which allows for the distributive delivery and quality assurance of stereotactic body radiotherapy (SBRT) plans. Method: Two clinically relevant and complex volumetric modulated arc therapy (VMAT) SBRT spine and lung plans were chosen as benchmarking cases. These were delivered on nine previously beam matched linacs with quality assurance performed through ArcCheck and film exposure in the sagittal plane. Measured doses were compared to their treatment planning system predictions through gamma analysis at a range of criteria. Results: Despite differences in beam match parameters and variations in small fields, all nine linacs produced accurate deliveries with a tight deviation in the population sample. Pass rates were well above suggested tolerances at the recommended gamma criterion. Film was able to detect dose errors to a greater degree than ArcCheck. Conclusion: Distributive quality assurance and delivery of stereotactic ablative radiotherapy treatments amongst beam matched linacs is certainly feasible provided the linacs are matched to a strict protocol like that suggested in this study and regular quality assurance is performed on the matched fleet. Distributive quality assurance and delivery of SBRT provides the possibility of efficiency gains for physicists as well as treatment staff. … (more)
- Is Part Of:
- Journal of applied clinical medical physics. Volume 20:Issue 4(2019)
- Journal:
- Journal of applied clinical medical physics
- Issue:
- Volume 20:Issue 4(2019)
- Issue Display:
- Volume 20, Issue 4 (2019)
- Year:
- 2019
- Volume:
- 20
- Issue:
- 4
- Issue Sort Value:
- 2019-0020-0004-0000
- Page Start:
- 99
- Page End:
- 105
- Publication Date:
- 2019-03-18
- Subjects:
- beam match -- match -- SBRT -- spine -- stereo -- VMAT
Medical physics -- Periodicals
Clinical medicine -- Periodicals
Health Physics
Clinical Medicine
Electronic journals
Periodicals
Periodicals
Fulltext
Internet Resources
610.153 - Journal URLs:
- http://aapm.onlinelibrary.wiley.com/hub/journal/10.1002/(ISSN)1526-9914/ ↗
http://bibpurl.oclc.org/web/7294 ↗
http://www.jacmp.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/acm2.12567 ↗
- Languages:
- English
- ISSNs:
- 1526-9914
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9733.xml