Impact of low-dose SPECT imaging on normal databases and myocardial perfusion scores. (March 2019)
- Record Type:
- Journal Article
- Title:
- Impact of low-dose SPECT imaging on normal databases and myocardial perfusion scores. (March 2019)
- Main Title:
- Impact of low-dose SPECT imaging on normal databases and myocardial perfusion scores
- Authors:
- Scabbio, Camilla
Malaspina, Simona
Capozza, Antonella
Selvaggi, Claudio
Matheoud, Roberta
Del Sole, Angelo
Lecchi, Michela - Abstract:
- Highlights: NDBs variation depends significantly on study count-statistics. Significantly lower SR% and SS% were found for 25%-count statistics. A 50% decrease in patient dose is the limit for Astonish™ with default parameters. Abstract: Purpose: We have recently demonstrated that iterative reconstruction algorithms with resolution recovery require the adoption of specific normal databases (NDBs) for perfusion SPECT quantification. This work was aimed at investigating the impact of patient low-dose imaging on NDBs and percent summed rest (SR%) and stress (SS%) scores. Methods: Assuming that count statistics of shorter acquisition time may simulate that of lower patient dose, three simultaneous scans were acquired (BrightView, Philips) with different acquisition-time/projection: 30, 15 and 8 s (from 100% to 25% of the reference). Fifty-two normal patients with low likelihood of coronary artery disease were enrolled and three homemade NDBs were then generated and compared (Astonish™ algorithm with default parameters): 100%-HM-NDBs, 50%-HM-NDBs and 25%-HM-NDBs. SR% and SS% were subsequently calculated for another group of 38 patients (normal/abnormal = 5/33). SR% and SS% values of 100%-HM-NDBs were compared with those obtained with the NDBs available on the workstation. Moreover, the impact of the study count statistics on perfusion scores was evaluated using the count-specific NDBs. Results: Significantly higher standard-deviation values were found for 25%-HM-NDBs compared toHighlights: NDBs variation depends significantly on study count-statistics. Significantly lower SR% and SS% were found for 25%-count statistics. A 50% decrease in patient dose is the limit for Astonish™ with default parameters. Abstract: Purpose: We have recently demonstrated that iterative reconstruction algorithms with resolution recovery require the adoption of specific normal databases (NDBs) for perfusion SPECT quantification. This work was aimed at investigating the impact of patient low-dose imaging on NDBs and percent summed rest (SR%) and stress (SS%) scores. Methods: Assuming that count statistics of shorter acquisition time may simulate that of lower patient dose, three simultaneous scans were acquired (BrightView, Philips) with different acquisition-time/projection: 30, 15 and 8 s (from 100% to 25% of the reference). Fifty-two normal patients with low likelihood of coronary artery disease were enrolled and three homemade NDBs were then generated and compared (Astonish™ algorithm with default parameters): 100%-HM-NDBs, 50%-HM-NDBs and 25%-HM-NDBs. SR% and SS% were subsequently calculated for another group of 38 patients (normal/abnormal = 5/33). SR% and SS% values of 100%-HM-NDBs were compared with those obtained with the NDBs available on the workstation. Moreover, the impact of the study count statistics on perfusion scores was evaluated using the count-specific NDBs. Results: Significantly higher standard-deviation values were found for 25%-HM-NDBs compared to the other HM-NDBs (p < 0.02). Significantly higher SS% were also found for the 100%-HM-NDBs compared to the workstation NDBs (95%CI: 0.15–2.11%). Moreover, a post-hoc test showed significantly lower SR% and SS% for 25%-count statistics compared to 100%-HM-NDBs (p < 0.03). Conclusions: NDBs and perfusion scores depend significantly on study count-statistics. A 50% reduction in patient dose is ultimately the limit for Astonish™ (with the default parameters) in order to prevent a significant variation in myocardial perfusion quantification. … (more)
- Is Part Of:
- Physica medica. Volume 59(2019)
- Journal:
- Physica medica
- Issue:
- Volume 59(2019)
- Issue Display:
- Volume 59, Issue 2019 (2019)
- Year:
- 2019
- Volume:
- 59
- Issue:
- 2019
- Issue Sort Value:
- 2019-0059-2019-0000
- Page Start:
- 163
- Page End:
- 169
- Publication Date:
- 2019-03
- Subjects:
- MPI myocardial perfusion imaging -- IRR iterative reconstruction algorithms with resolution recovery -- LV left ventricle -- GP general purpose -- BMI body mass index -- SR% percent summed rest score -- SS% percent summed stress score
Myocardial perfusion SPECT imaging -- Iterative reconstruction -- Normal databases -- Patient dose
Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2019.03.009 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9738.xml