Underestimation of 68Ga PET/CT SUV caused by activity overestimation using default calibrator settings. (March 2019)
- Record Type:
- Journal Article
- Title:
- Underestimation of 68Ga PET/CT SUV caused by activity overestimation using default calibrator settings. (March 2019)
- Main Title:
- Underestimation of 68Ga PET/CT SUV caused by activity overestimation using default calibrator settings
- Authors:
- Sanderson, Tom
Solomon, James
Nottage, Chris
Dickson, John - Abstract:
- Highlights: SUV quantification of a 68 Ga PET/CT phantom scan was underestimated. The primary cause was overestimation of 68 Ga activity in a radionuclide calibrator. 68 Ga activity was overestimated by 7.8–9.4% using default settings. Quantitative accuracy was improved by adjusting the 68 Ga calibrator factor. Abstract: Purpose: A PET/CT scan of a uniformly filled 68 Ga phantom resulted in an unexpectedly low SUVmean of 0.88. A potential contributing cause of underestimation of 68 Ga SUV is overestimation of 68 Ga activity in the Radionuclide Calibrator associated with the PET/CT scanner. To investigate this, a Radionuclide Calibrator cross-calibration exercise was performed. Methods: A source of 68 Ga was measured in 5 Capintec CRC-55tR calibrators using the pre-set 68 Ga calibrator factor of 416, and a Veenstra VDC-505 calibrator using a factor recommended by the manufacturer. The source was then measured in an externally located Fidelis Secondary Standard Radionuclide Calibrator. Manual adjustments were made to the Capintec calibrator factors to match the decay corrected Fidelis measurement, followed by a repeat PET/CT scan of a uniform 68 Ga phantom. Results: The cross-calibration results showed that the 5 Capintec calibrators systematically overestimated 68 Ga activity by 7.8–9.4% (mean 8.5%) compared to the Fidelis. The calibrator factors were adjusted to 456–464 to match the Fidelis measurement, and the repeat phantom scan resulted in a SUVmean of 0.97, within theHighlights: SUV quantification of a 68 Ga PET/CT phantom scan was underestimated. The primary cause was overestimation of 68 Ga activity in a radionuclide calibrator. 68 Ga activity was overestimated by 7.8–9.4% using default settings. Quantitative accuracy was improved by adjusting the 68 Ga calibrator factor. Abstract: Purpose: A PET/CT scan of a uniformly filled 68 Ga phantom resulted in an unexpectedly low SUVmean of 0.88. A potential contributing cause of underestimation of 68 Ga SUV is overestimation of 68 Ga activity in the Radionuclide Calibrator associated with the PET/CT scanner. To investigate this, a Radionuclide Calibrator cross-calibration exercise was performed. Methods: A source of 68 Ga was measured in 5 Capintec CRC-55tR calibrators using the pre-set 68 Ga calibrator factor of 416, and a Veenstra VDC-505 calibrator using a factor recommended by the manufacturer. The source was then measured in an externally located Fidelis Secondary Standard Radionuclide Calibrator. Manual adjustments were made to the Capintec calibrator factors to match the decay corrected Fidelis measurement, followed by a repeat PET/CT scan of a uniform 68 Ga phantom. Results: The cross-calibration results showed that the 5 Capintec calibrators systematically overestimated 68 Ga activity by 7.8–9.4% (mean 8.5%) compared to the Fidelis. The calibrator factors were adjusted to 456–464 to match the Fidelis measurement, and the repeat phantom scan resulted in a SUVmean of 0.97, within the local tolerance of 1.00 ± 5%. The result for the Veenstra calibrator was within the tolerance of ±5%. Conclusions: Underestimation of 68 Ga SUV was primarily caused by overestimation of 68 Ga activity using the pre-set calibrator factor setting on a Capintec CRC-55tR. Improvement in quantification accuracy was achieved by adjusting the 68 Ga calibrator factor based upon a cross-calibration exercise. We recommend SUV checks using a uniform phantom and regular calibrator cross-calibration exercises for all isotopes used for quantitative PET/CT imaging. … (more)
- Is Part Of:
- Physica medica. Volume 59(2019)
- Journal:
- Physica medica
- Issue:
- Volume 59(2019)
- Issue Display:
- Volume 59, Issue 2019 (2019)
- Year:
- 2019
- Volume:
- 59
- Issue:
- 2019
- Issue Sort Value:
- 2019-0059-2019-0000
- Page Start:
- 158
- Page End:
- 162
- Publication Date:
- 2019-03
- Subjects:
- PET -- SUV -- 68Ga -- Calibration
Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2019.03.006 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9719.xml