Systematic Review and Meta-Analysis of Method Comparison Studies of Masimo Pulse Co-Oximeters (Radical-7™ or Pronto-7™) and HemoCue® Absorption Spectrometers (B-Hemoglobin or 201+) with Laboratory Haemoglobin Estimation. Issue 3 (May 2015)
- Record Type:
- Journal Article
- Title:
- Systematic Review and Meta-Analysis of Method Comparison Studies of Masimo Pulse Co-Oximeters (Radical-7™ or Pronto-7™) and HemoCue® Absorption Spectrometers (B-Hemoglobin or 201+) with Laboratory Haemoglobin Estimation. Issue 3 (May 2015)
- Main Title:
- Systematic Review and Meta-Analysis of Method Comparison Studies of Masimo Pulse Co-Oximeters (Radical-7™ or Pronto-7™) and HemoCue® Absorption Spectrometers (B-Hemoglobin or 201+) with Laboratory Haemoglobin Estimation
- Authors:
- Hiscock, R.
Kumar, D.
Simmons, S. W. - Abstract:
- We assessed agreement in haemoglobin measurement between Masimo pulse co-oximeters (Rad-7 ™ and Pronto-7 ™ ) and HemoCue ® photometers (201+ or B-Hemoglobin) with laboratory-based determination and identified 39 relevant studies (2915 patients in Masimo group and 3084 patients in HemoCue group). In the Masimo group, the overall mean difference was -0.03 g/dl (95% prediction interval -0.30 to 0.23) and 95% limits of agreement -3.0 to 2.9 g/dl compared to 0.08 g/dl (95% prediction interval -0.04 to 0.20) and 95% limits of agreement -1.3 to 1.4 g/dl in the HemoCue group. Only B-Hemoglobin exhibited bias (0.53, 95% prediction interval 0.27 to 0.78). The overall standard deviation of difference was larger (1.42 g/dl versus 0.64 g/dl) for Masimo pulse co-oximeters compared to HemoCue photometers. Masimo devices and HemoCue 201+ both provide an unbiased, pooled estimate of laboratory haemoglobin. However, Masimo devices have lower precision and wider 95% limits of agreement than HemoCue devices. Clinicians should carefully consider these limits of agreement before basing transfusion or other clinical decisions on these point-of-care measurements alone.
- Is Part Of:
- Anaesthesia and intensive care. Volume 43:Issue 3(2015)
- Journal:
- Anaesthesia and intensive care
- Issue:
- Volume 43:Issue 3(2015)
- Issue Display:
- Volume 43, Issue 3 (2015)
- Year:
- 2015
- Volume:
- 43
- Issue:
- 3
- Issue Sort Value:
- 2015-0043-0003-0000
- Page Start:
- 341
- Page End:
- 350
- Publication Date:
- 2015-05
- Subjects:
- haemoglobin -- oximetry
Anesthesiology -- Periodicals
Intensive Care Units -- Periodicals
617.96 - Journal URLs:
- https://journals.sagepub.com/home/aic ↗
- DOI:
- 10.1177/0310057X1504300310 ↗
- Languages:
- English
- ISSNs:
- 0310-057X
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library HMNTS - ELD Digital store
- Ingest File:
- 9671.xml