Prospect of robotic assistance for fully automated brachytherapy seed placement into skull base: Experimental validation in phantom and cadaver. (February 2019)
- Record Type:
- Journal Article
- Title:
- Prospect of robotic assistance for fully automated brachytherapy seed placement into skull base: Experimental validation in phantom and cadaver. (February 2019)
- Main Title:
- Prospect of robotic assistance for fully automated brachytherapy seed placement into skull base: Experimental validation in phantom and cadaver
- Authors:
- Zhu, Jian-Hua
Wang, Jing
Wang, Yong-Gui
Li, Meng
Liu, Xiao-Jing
Guo, Chuan-Bin - Abstract:
- Abstract: Background and purpose: To investigate the feasibility and accuracy of robot-assisted brachytherapy for skull base tumours. Material and methods: A custom robot system was tested on both phantom and cadaveric specimen. Cone beam CT (CBCT) images were transferred to the graphical user interface (GUI) for planning trajectories and the data were sent to the robot control unit. Following registration, the puncture needle was inserted into the target by the robot under navigation guidance, and seeds were implanted. Placement error was instantly displayed on the GUI; the result was verified after postoperative image scanning. Results: A total of 150 seeds (100 for phantom experiments, 50 for cadaveric studies) were deposited by the robot system. In phantom experiments the mean placement error was 0.57 ± 0.21 mm (measured by the navigation system) vs. 1.41 ± 0.38 mm (measured by image fusion) ( p < 0.001); in cadaveric studies the corresponding figures were 0.60 ± 0.30 mm vs. 2.48 ± 0.32 mm ( p < 0.001). There was no significant difference for comparison of accuracy test in phantom experiments ( p = 0.173) as well as in cadaveric studies ( p = 0.354). Accuracy was better in the phantom experiment than in cadaveric studies ( p < 0.001). Conclusions: The performance of robot-assisted skull base brachytherapy is feasible and accurate. Dosimetric coverage will need to be demonstrated in further studies.
- Is Part Of:
- Radiotherapy and oncology. Volume 131(2019)
- Journal:
- Radiotherapy and oncology
- Issue:
- Volume 131(2019)
- Issue Display:
- Volume 131, Issue 2019 (2019)
- Year:
- 2019
- Volume:
- 131
- Issue:
- 2019
- Issue Sort Value:
- 2019-0131-2019-0000
- Page Start:
- 160
- Page End:
- 165
- Publication Date:
- 2019-02
- Subjects:
- Robot -- Brachytherapy -- Skull base -- Navigation -- Seed localization accuracy
Oncology -- Periodicals
Radiotherapy -- Periodicals
Tumors -- Periodicals
Medical Oncology -- Periodicals
Neoplasms -- radiotherapy -- Periodicals
Radiotherapy -- Periodicals
Radiothérapie -- Périodiques
Cancérologie -- Périodiques
Tumeurs -- Périodiques
Electronic journals
616.9940642 - Journal URLs:
- http://www.sciencedirect.com/science/journal/01678140 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/01678140 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/01678140 ↗
http://www.estro.org/ ↗
http://www.elsevier.com/journals ↗
http://www.journals.elsevier.com/radiotherapy-and-oncology/ ↗ - DOI:
- 10.1016/j.radonc.2017.11.031 ↗
- Languages:
- English
- ISSNs:
- 0167-8140
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 7240.790000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9646.xml