Does the condylar lift-off method or the separation method better detect loss of contact between tibial and femoral implants based on analysis of single-plane radiographs following total knee arthroplasty?. (27th March 2019)
- Record Type:
- Journal Article
- Title:
- Does the condylar lift-off method or the separation method better detect loss of contact between tibial and femoral implants based on analysis of single-plane radiographs following total knee arthroplasty?. (27th March 2019)
- Main Title:
- Does the condylar lift-off method or the separation method better detect loss of contact between tibial and femoral implants based on analysis of single-plane radiographs following total knee arthroplasty?
- Authors:
- Hull, Maury L.
Ross, Derrick
Nicolet-Petersen, Stephanie
Howell, Stephen M. - Abstract:
- Abstract: Background: Loss of contact between the femoral and tibial implants following total knee arthroplasty (TKA) has been related to accelerated polyethylene wear and other complications. Two methods have been used to detect loss of contact in single-plane fluoroscopy, the condylar lift-off method and the separation method. The objectives were to assess the ability of each method to detect loss of contact. Methods: TKA was performed on ten cadaveric knee specimens. Tibial force was measured in each compartment as specimens were flexed from 0° to 90° while internal-external and varus-valgus moments were applied. Single-plane radiographs taken simultaneously with tibial force were analyzed for loss of contact using the two methods. Receiver operating characteristic (ROC) and optimum threshold distances were determined. Results: For the lift-off method and the separation method, the areas under the ROC curves were 0.89 vs 0.60 for the lateral compartment only and 0.81 vs 0.70 for the medial compartment only, respectively. For the lift-off method, the optimum threshold distances were 0.7 mm in the lateral compartment only and 0.1 mm in the medial compartment only but the false positive rate for the medial compartment only almost doubled. For both compartments jointly, the areas under the ROC curves decreased to 0.70 and 0.59 for the lift-off and separation methods, respectively. Conclusion: When detecting loss of contact using single-plane fluoroscopy, the lift-off methodAbstract: Background: Loss of contact between the femoral and tibial implants following total knee arthroplasty (TKA) has been related to accelerated polyethylene wear and other complications. Two methods have been used to detect loss of contact in single-plane fluoroscopy, the condylar lift-off method and the separation method. The objectives were to assess the ability of each method to detect loss of contact. Methods: TKA was performed on ten cadaveric knee specimens. Tibial force was measured in each compartment as specimens were flexed from 0° to 90° while internal-external and varus-valgus moments were applied. Single-plane radiographs taken simultaneously with tibial force were analyzed for loss of contact using the two methods. Receiver operating characteristic (ROC) and optimum threshold distances were determined. Results: For the lift-off method and the separation method, the areas under the ROC curves were 0.89 vs 0.60 for the lateral compartment only and 0.81 vs 0.70 for the medial compartment only, respectively. For the lift-off method, the optimum threshold distances were 0.7 mm in the lateral compartment only and 0.1 mm in the medial compartment only but the false positive rate for the medial compartment only almost doubled. For both compartments jointly, the areas under the ROC curves decreased to 0.70 and 0.59 for the lift-off and separation methods, respectively. Conclusion: When detecting loss of contact using single-plane fluoroscopy, the lift-off method is useful for the lateral compartment only but not for the medial compartment only and not for both compartments jointly. The separation method is not useful. … (more)
- Is Part Of:
- Journal of biomechanics. Volume 86(2019)
- Journal:
- Journal of biomechanics
- Issue:
- Volume 86(2019)
- Issue Display:
- Volume 86, Issue 2019 (2019)
- Year:
- 2019
- Volume:
- 86
- Issue:
- 2019
- Issue Sort Value:
- 2019-0086-2019-0000
- Page Start:
- 40
- Page End:
- 47
- Publication Date:
- 2019-03-27
- Subjects:
- Fluoroscopy -- Bi-planar -- Knee replacement -- Aseptic loosening -- Tibial insert wear
Animal mechanics -- Periodicals
Biomechanics -- Periodicals
Biomechanics -- Periodicals
Mécanique animale -- Périodiques
Biomécanique -- Périodiques
Electronic journals
571.4305 - Journal URLs:
- http://www.sciencedirect.com/science/journal/00219290 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/00219290 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/00219290 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.jbiomech.2019.01.038 ↗
- Languages:
- English
- ISSNs:
- 0021-9290
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4953.600000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9642.xml