Scanner influence on the mechanical response of QCT-based finite element analysis of long bones. (27th March 2019)
- Record Type:
- Journal Article
- Title:
- Scanner influence on the mechanical response of QCT-based finite element analysis of long bones. (27th March 2019)
- Main Title:
- Scanner influence on the mechanical response of QCT-based finite element analysis of long bones
- Authors:
- Katz, Yekutiel
Dahan, Gal
Sosna, Jacob
Shelef, Ilan
Cherniavsky, Evgenia
Yosibash, Zohar - Abstract:
- Abstract: Patient-specific QCT-based finite element (QCTFE) analyses enable highly accurate quantification of bone strength. We evaluated CT scanner influence on QCTFE models of long bones. A femur, humerus, and proximal femur without the head were scanned with K2 HPO4 phantoms by seven CT scanners (four models) using typical clinical protocols. QCTFE models were constructed. The geometrical dimensions, as well as the QCT-values expressed in Hounsfield unit (HU) distribution was compared. Principal strains at representative regions of interest (ROIs), and maximum principal strains (associated with fracture risk) were compared. Intraclass correlation coefficients (ICCs) were calculated to evaluate strain prediction reliability for different scanners. Repeatability was examined by scanning the femur twice and comparing resulting QCTFE models. Maximum difference in geometry was 2.3%. HU histograms before phantom calibration showed wide variation between QCT scans; however, bone density histogram variability was reduced after calibration and algorithmic manipulation. Relative standard deviation (RSD) in principal strains at ROIs was <10.7%. ICC estimates between scanners were >0.9. Fracture-associated strain had 6.7%, 8.1%, and 13.3% maximum RSD for the femur, humerus, and proximal femur, respectively. The difference in maximum strain location was <2 mm. The average difference with repeat scans was 2.7%. Quantification of strain differences showed mean RSD bounded by ∼6% inAbstract: Patient-specific QCT-based finite element (QCTFE) analyses enable highly accurate quantification of bone strength. We evaluated CT scanner influence on QCTFE models of long bones. A femur, humerus, and proximal femur without the head were scanned with K2 HPO4 phantoms by seven CT scanners (four models) using typical clinical protocols. QCTFE models were constructed. The geometrical dimensions, as well as the QCT-values expressed in Hounsfield unit (HU) distribution was compared. Principal strains at representative regions of interest (ROIs), and maximum principal strains (associated with fracture risk) were compared. Intraclass correlation coefficients (ICCs) were calculated to evaluate strain prediction reliability for different scanners. Repeatability was examined by scanning the femur twice and comparing resulting QCTFE models. Maximum difference in geometry was 2.3%. HU histograms before phantom calibration showed wide variation between QCT scans; however, bone density histogram variability was reduced after calibration and algorithmic manipulation. Relative standard deviation (RSD) in principal strains at ROIs was <10.7%. ICC estimates between scanners were >0.9. Fracture-associated strain had 6.7%, 8.1%, and 13.3% maximum RSD for the femur, humerus, and proximal femur, respectively. The difference in maximum strain location was <2 mm. The average difference with repeat scans was 2.7%. Quantification of strain differences showed mean RSD bounded by ∼6% in ROIs. Fracture-associated strains in "regular" bones showed a mean RSD bounded by ∼8%. Strains were obtained within a ±10% difference relative to the mean; thus, in a longitudinal study only changes larger than 20% in the principal strains may be significant. ICCs indicated high reliability of QCTFE models derived from different scanners. … (more)
- Is Part Of:
- Journal of biomechanics. Volume 86(2019)
- Journal:
- Journal of biomechanics
- Issue:
- Volume 86(2019)
- Issue Display:
- Volume 86, Issue 2019 (2019)
- Year:
- 2019
- Volume:
- 86
- Issue:
- 2019
- Issue Sort Value:
- 2019-0086-2019-0000
- Page Start:
- 149
- Page End:
- 159
- Publication Date:
- 2019-03-27
- Subjects:
- QCT -- Femur -- Finite element analysis -- Humerus -- Patient-specific -- Personalized medicine
Animal mechanics -- Periodicals
Biomechanics -- Periodicals
Biomechanics -- Periodicals
Mécanique animale -- Périodiques
Biomécanique -- Périodiques
Electronic journals
571.4305 - Journal URLs:
- http://www.sciencedirect.com/science/journal/00219290 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/00219290 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/00219290 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.jbiomech.2019.01.049 ↗
- Languages:
- English
- ISSNs:
- 0021-9290
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4953.600000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9642.xml