The sensitivity of gamma index analysis to detect multileaf collimator (MLC) positioning errors using Varian TrueBeam EPID and ArcCHECK for patient-specific prostate volumetric-modulated arc therapy (VMAT) quality assurance. Issue 1 (11th July 2017)
- Record Type:
- Journal Article
- Title:
- The sensitivity of gamma index analysis to detect multileaf collimator (MLC) positioning errors using Varian TrueBeam EPID and ArcCHECK for patient-specific prostate volumetric-modulated arc therapy (VMAT) quality assurance. Issue 1 (11th July 2017)
- Main Title:
- The sensitivity of gamma index analysis to detect multileaf collimator (MLC) positioning errors using Varian TrueBeam EPID and ArcCHECK for patient-specific prostate volumetric-modulated arc therapy (VMAT) quality assurance
- Authors:
- Maraghechi, Borna
Davis, Jack
Mitchell, Nicholas
Shah, Meeral
Fleck, Andre
Darko, Johnson
Osei, Ernest - Abstract:
- Abstract: Background: Due to the increased degree of modulation and complexity of volumetric-modulated arc therapy (VMAT) plans, it is necessary to have a pre-treatment patient-specific quality assurance (QA) programme. The gamma index is commonly used to quantitatively compare two dose distributions. In this study we investigated the sensitivity of single- and multi-gamma criteria techniques to detect multileaf collimator (MLC) positioning errors using the Varian TrueBeam Electronic Portal Imaging Device TM (EPID) dosimetry and the ArcCHECK TM device. Materials and methods: All active MLC positions of seven intact prostate patients VMAT plans were randomly changed with a mean value of 0.25, 0.5, 1 and 2 mm and a standard deviation of 0.1 mm on 25, 50, 75 and 100% of the control points. The change in gamma passing rates of six gamma criteria of 3%/3 mm, 3%/2 mm, 3%/1 mm, 2%/2 mm, 2%/1 mm and 1%/1 mm were analysed individually (single-gamma criterion) and as a group (multi-gamma criteria) as a function of the simulated errors. We used the improved and global gamma calculation algorithms with a low dose threshold of 10% in the EPID and ArcCHECK software, respectively. The changes in the planning target volume dose distributions and the organs at risk due to the MLC positioning errors were also studied. Results: When 25, 50, 75 and 100% of the control points were modified by the introduction of the simulated errors, the smallest detectable errors with the EPID were 2, 1, 0.5Abstract: Background: Due to the increased degree of modulation and complexity of volumetric-modulated arc therapy (VMAT) plans, it is necessary to have a pre-treatment patient-specific quality assurance (QA) programme. The gamma index is commonly used to quantitatively compare two dose distributions. In this study we investigated the sensitivity of single- and multi-gamma criteria techniques to detect multileaf collimator (MLC) positioning errors using the Varian TrueBeam Electronic Portal Imaging Device TM (EPID) dosimetry and the ArcCHECK TM device. Materials and methods: All active MLC positions of seven intact prostate patients VMAT plans were randomly changed with a mean value of 0.25, 0.5, 1 and 2 mm and a standard deviation of 0.1 mm on 25, 50, 75 and 100% of the control points. The change in gamma passing rates of six gamma criteria of 3%/3 mm, 3%/2 mm, 3%/1 mm, 2%/2 mm, 2%/1 mm and 1%/1 mm were analysed individually (single-gamma criterion) and as a group (multi-gamma criteria) as a function of the simulated errors. We used the improved and global gamma calculation algorithms with a low dose threshold of 10% in the EPID and ArcCHECK software, respectively. The changes in the planning target volume dose distributions and the organs at risk due to the MLC positioning errors were also studied. Results: When 25, 50, 75 and 100% of the control points were modified by the introduction of the simulated errors, the smallest detectable errors with the EPID were 2, 1, 0.5 and 0.5 mm, respectively, using the multi-gamma criteria technique. Similarly for the single-gamma criteria technique errors as small as 2, 1, 1 and 1 mm applied to 25, 50, 75 and 100% of the control points, respectively, were detectable using a 2%/2 mm criterion. However, the smallest detectable errors with the ArcCHECK when using the multi-gamma criteria technique were 2, 2 and 1 mm when MLC errors were applied on 50, 75 and 100% of the control points. When only 25% of the control points were affected the ArcCHECK were unable to detect any of the errors applied. No noticeable difference was observed in the sensitivity using the single- or the multi-gamma criteria techniques with the ArcCHECK. Conclusion: The Varian TrueBeam EPID dosimetry shows a higher sensitivity in detecting MLC positioning errors compared with the ArcCHECK regardless of using the single- or the multi-gamma criteria techniques. Higher sensitivity was observed using the multi-gamma criteria technique compared with the single-criterion technique when using the EPID. … (more)
- Is Part Of:
- Journal of radiotherapy in practice. Volume 17:Issue 1(2018)
- Journal:
- Journal of radiotherapy in practice
- Issue:
- Volume 17:Issue 1(2018)
- Issue Display:
- Volume 17, Issue 1 (2018)
- Year:
- 2018
- Volume:
- 17
- Issue:
- 1
- Issue Sort Value:
- 2018-0017-0001-0000
- Page Start:
- 66
- Page End:
- 77
- Publication Date:
- 2017-07-11
- Subjects:
- ArcCHECK, -- EPID dosimetry, -- gamma index, -- prostate cancer, -- VMAT quality assurance
Radiotherapy -- Periodicals
615.842005 - Journal URLs:
- http://journals.cambridge.org/action/displayJournal?jid=JRP ↗
- DOI:
- 10.1017/S1460396917000425 ↗
- Languages:
- English
- ISSNs:
- 1460-3969
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library HMNTS - ELD Digital store
- Ingest File:
- 9623.xml