Scaling Relationship of In Vivo Muscle Contraction Strength of Rabbits Exposed to High-Frequency Nanosecond Pulse Bursts. (16th July 2018)
- Record Type:
- Journal Article
- Title:
- Scaling Relationship of In Vivo Muscle Contraction Strength of Rabbits Exposed to High-Frequency Nanosecond Pulse Bursts. (16th July 2018)
- Main Title:
- Scaling Relationship of In Vivo Muscle Contraction Strength of Rabbits Exposed to High-Frequency Nanosecond Pulse Bursts
- Authors:
- Mi, Yan
Xu, Jin
Tang, Xuefeng
Bian, Changhao
Liu, Hongliang
Yang, Qiyu
Tang, Junying - Abstract:
- We studied the influence of various parameters of high-frequency nanosecond pulse bursts on the strength of rabbit muscle contractions. Ten unipolar high-frequency pulse bursts with various field intensities E (1 kV/cm, 4 kV/cm, and 8 kV/cm), intraburst frequencies f (10 kHz, 100 kHz, and 1 MHz), and intraburst pulse numbers N (1, 10, and 100) were applied using a pair of plate electrodes to the surface skin of the rabbits' biceps femoris, and the acceleration signal of muscle contraction near the electrode was measured using a 3-axis acceleration sensor. A time- and frequency-domain analysis of the acceleration signals showed that the peak value of the signal increases with the increasing strength of the pulse burst and that the frequency spectra of the signals measured under various pulse bursts have characteristic frequencies (at approximately 2 Hz, 32 Hz, 45 Hz, and 55 Hz). Furthermore, we processed the data through multivariate nonlinear regression analysis and variance analysis and determined that the peak value of the signal scales with the logarithm to the base 10 of EN x, where x is a value that scales with the logarithm to the base 10 of intraburst frequency (f). These results indicate that for high-frequency nanosecond pulse treatment of solid tumors in or near muscles, when the field strength is relatively high, the intraburst frequency and the intraburst pulse number require appropriate selection to limit the strength of muscle contraction as much as possible.
- Is Part Of:
- Technology in cancer research & treatment. Volume 17(2018)
- Journal:
- Technology in cancer research & treatment
- Issue:
- Volume 17(2018)
- Issue Display:
- Volume 17, Issue 2018 (2018)
- Year:
- 2018
- Volume:
- 17
- Issue:
- 2018
- Issue Sort Value:
- 2018-0017-2018-0000
- Page Start:
- Page End:
- Publication Date:
- 2018-07-16
- Subjects:
- high-frequency nanosecond pulse bursts -- rabbit muscle contractions -- dose effect -- acceleration signal -- spectrum distribution -- function relation
Oncology -- Periodicals
Cancer -- Diagnosis -- Periodicals
Cancer -- Treatment -- Technological innovations -- Periodicals
616.994 - Journal URLs:
- http://tct.sagepub.com/ ↗
http://www.tcrt.org ↗
http://www.sagepub.com ↗ - DOI:
- 10.1177/1533033818788078 ↗
- Languages:
- English
- ISSNs:
- 1533-0346
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9605.xml