Fractography of clinically fractured, implant-supported dental computer-aided design and computer-aided manufacturing crowns. (17th November 2017)
- Record Type:
- Journal Article
- Title:
- Fractography of clinically fractured, implant-supported dental computer-aided design and computer-aided manufacturing crowns. (17th November 2017)
- Main Title:
- Fractography of clinically fractured, implant-supported dental computer-aided design and computer-aided manufacturing crowns
- Authors:
- Lohbauer, Ulrich
Belli, Renan
Cune, Marco S
Schepke, Ulf - Abstract:
- Today, a substantial part of the dental crown production uses computer-aided design and computer-aided manufacturing (CAD/CAM) technology. A recent step in restorative dentistry is the replacement of natural tooth structure with pre-polymerized and machined resin-based methacrylic polymers. Recently, a new CAD/CAM composite was launched for the crown indication in the load-bearing area, but the clinical reality forced the manufacturer to withdraw this specific indication. In parallel, a randomized clinical trial of CAD/CAM composite crowns luted on zirconia implant abutments revealed a high incidence of failure within the first year of service. Fractured crowns of this clinical trial were retrieved and submitted to a fractographic examination. The aim of the case series presented in this article was to identify failure reasons for a new type of CAD/CAM composite crown material (Lava Ultimate; 3M Oral Care, St. Paul, Minnesota, USA) via fractographic examinations and analytical assessment of luting surfaces and water absorption behavior. As a result, the debonding of the composite crowns from the zirconia implant abutments was identified as the central reason for failure. The adhesive interface was found the weakest link. A lack of silica at the zirconia surface certainly has compromised the bonding potential of the adhesive system from the beginning. Additionally, the hydrolytic stress released from swelling of the resin-based crown (water absorption) and transfer to theToday, a substantial part of the dental crown production uses computer-aided design and computer-aided manufacturing (CAD/CAM) technology. A recent step in restorative dentistry is the replacement of natural tooth structure with pre-polymerized and machined resin-based methacrylic polymers. Recently, a new CAD/CAM composite was launched for the crown indication in the load-bearing area, but the clinical reality forced the manufacturer to withdraw this specific indication. In parallel, a randomized clinical trial of CAD/CAM composite crowns luted on zirconia implant abutments revealed a high incidence of failure within the first year of service. Fractured crowns of this clinical trial were retrieved and submitted to a fractographic examination. The aim of the case series presented in this article was to identify failure reasons for a new type of CAD/CAM composite crown material (Lava Ultimate; 3M Oral Care, St. Paul, Minnesota, USA) via fractographic examinations and analytical assessment of luting surfaces and water absorption behavior. As a result, the debonding of the composite crowns from the zirconia implant abutments was identified as the central reason for failure. The adhesive interface was found the weakest link. A lack of silica at the zirconia surface certainly has compromised the bonding potential of the adhesive system from the beginning. Additionally, the hydrolytic stress released from swelling of the resin-based crown (water absorption) and transfer to the luting interface further added to the interfacial stress and most probably contributed to a great extend to the debonding failure. … (more)
- Is Part Of:
- SAGE open medical case reports. Volume 5(2017)
- Journal:
- SAGE open medical case reports
- Issue:
- Volume 5(2017)
- Issue Display:
- Volume 5, Issue 2017 (2017)
- Year:
- 2017
- Volume:
- 5
- Issue:
- 2017
- Issue Sort Value:
- 2017-0005-2017-0000
- Page Start:
- Page End:
- Publication Date:
- 2017-11-17
- Subjects:
- Dentistry -- fractography -- computer-aided design and computer-aided manufacturing composite -- adhesive failure -- crown restoration -- lava -- ultimate -- hydrolytic degradation
Clinical medicine -- Case studies -- Periodicals
616.09 - Journal URLs:
- http://sco.sagepub.com/ ↗
http://www.uk.sagepub.com/home.nav ↗ - DOI:
- 10.1177/2050313X17741015 ↗
- Languages:
- English
- ISSNs:
- 2050-313X
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9600.xml