Biotin interference in TSH, FT4, and FT3 assays based on the LOCI technology: Identifying interference by dilution. Issue 2 (17th September 2018)
- Record Type:
- Journal Article
- Title:
- Biotin interference in TSH, FT4, and FT3 assays based on the LOCI technology: Identifying interference by dilution. Issue 2 (17th September 2018)
- Main Title:
- Biotin interference in TSH, FT4, and FT3 assays based on the LOCI technology: Identifying interference by dilution
- Authors:
- Dasgupta, Amitava
Bourgeois, Laverne - Abstract:
- Abstract: Background: Although biotin interferences in TSH, FT3, FT4, and other biotinylated antibody‐based assays manufactured by Roche Diagnostics have been well studied, there are relatively few reports on biotin interference in biotin‐based assays manufactured by other companies. We investigated biotin interferences in TSH, FT4, and FT3 assays based on the LOCI (luminescent oxygen channeling assay) technology using the Dimension Vista 1500 analyzer (Siemens). Methods: We prepared four serum pools using leftover specimens. Three serum pools were prepared initially for the original study but the 4 th pool was prepared three months later. The aliquots of serum pool one and two were supplemented with various amounts of biotin (50 ‐1200 ng/mL) followed by determination of TSH, FT4, and FT3 concentrations. The aliquots of third pool were also supplemented with biotin to investigate whether 1:3 dilution could identify biotin interference. Aliquots of serum pool four were supplemented with biotin in order to study reproducibility of our original data. Results: We observed significantly elevated FT3 levels at biotin concentration of 100 ng/mL. In contrast, FT4 levels were falsely elevated but TSH levels were falsely decreased at a biotin level of 500 ng/mL. We also observed nonlinearity in dilution experiment. Conclusions: We conclude that FT3 assay is most susceptible to biotin interference (threshold: 100 ng/mL) while the FT4 and TSH assays are less affected (threshold:Abstract: Background: Although biotin interferences in TSH, FT3, FT4, and other biotinylated antibody‐based assays manufactured by Roche Diagnostics have been well studied, there are relatively few reports on biotin interference in biotin‐based assays manufactured by other companies. We investigated biotin interferences in TSH, FT4, and FT3 assays based on the LOCI (luminescent oxygen channeling assay) technology using the Dimension Vista 1500 analyzer (Siemens). Methods: We prepared four serum pools using leftover specimens. Three serum pools were prepared initially for the original study but the 4 th pool was prepared three months later. The aliquots of serum pool one and two were supplemented with various amounts of biotin (50 ‐1200 ng/mL) followed by determination of TSH, FT4, and FT3 concentrations. The aliquots of third pool were also supplemented with biotin to investigate whether 1:3 dilution could identify biotin interference. Aliquots of serum pool four were supplemented with biotin in order to study reproducibility of our original data. Results: We observed significantly elevated FT3 levels at biotin concentration of 100 ng/mL. In contrast, FT4 levels were falsely elevated but TSH levels were falsely decreased at a biotin level of 500 ng/mL. We also observed nonlinearity in dilution experiment. Conclusions: We conclude that FT3 assay is most susceptible to biotin interference (threshold: 100 ng/mL) while the FT4 and TSH assays are less affected (threshold: 500 ng/mL). In addition, we also observed nonlinearity upon 1:3 dilution, which may indicate biotin interference (or interference from other compounds). … (more)
- Is Part Of:
- Journal of clinical laboratory analysis. Volume 33:Issue 2(2019)
- Journal:
- Journal of clinical laboratory analysis
- Issue:
- Volume 33:Issue 2(2019)
- Issue Display:
- Volume 33, Issue 2 (2019)
- Year:
- 2019
- Volume:
- 33
- Issue:
- 2
- Issue Sort Value:
- 2019-0033-0002-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2018-09-17
- Subjects:
- biotin -- FT3 -- FT4 -- LOCI technology -- TSH
Diagnosis, Laboratory -- Periodicals
Medical laboratory technology -- Periodicals
616 - Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
- DOI:
- 10.1002/jcla.22667 ↗
- Languages:
- English
- ISSNs:
- 0887-8013
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4958.520000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9549.xml