Do the representative beam data for TrueBeam™ linear accelerators represent average data?. Issue 2 (13th January 2019)
- Record Type:
- Journal Article
- Title:
- Do the representative beam data for TrueBeam™ linear accelerators represent average data?. Issue 2 (13th January 2019)
- Main Title:
- Do the representative beam data for TrueBeam™ linear accelerators represent average data?
- Authors:
- Tanaka, Yoshihiro
Mizuno, Hirokazu
Akino, Yuichi
Isono, Masaru
Masai, Norimasa
Yamamoto, Toshijiro - Abstract:
- Abstract: If the vendor's representative beam data (RBD) for TrueBeam linear accelerators are to be valid for use in clinical practice, the variations in the beam data used for beam modeling must be small. Although a few studies have reported the variation of the beam data of the TrueBeam machines, the numbers of machines analyzed in those studies were small. In this study, we investigated the variation in the beam data for 21 TrueBeam machines collected from 17 institutions with their agreement. In the exponential regions, the percent depth dose (PDD) values showed very small variation, <1% for all the photon energies analyzed. Similarly, the off‐center ratio (OCR) values also showed small variation for all energies. In the field regions, the standard deviations of the values of dose difference (DD) between the data for each machine and the study average were <1% for field sizes ≥100 × 100 mm 2 . The maximum distance‐to‐agreement from the average data was <0.5 mm in the penumbra regions. The output factor (OPF) values also showed very small variation (<1%) for all energies and field sizes. Both the PDD and OCR of the average study data showed good agreement with the vendor's RBD for field sizes ≥100 × 100 mm 2 . The OPF of the average study data also showed good agreement with the vendor's RBD for all field sizes. However, although all the institutions used ionization chambers with similar cavity volumes, the 30 × 30 mm 2 field size showed large DD variations (≥2%) in OCRAbstract: If the vendor's representative beam data (RBD) for TrueBeam linear accelerators are to be valid for use in clinical practice, the variations in the beam data used for beam modeling must be small. Although a few studies have reported the variation of the beam data of the TrueBeam machines, the numbers of machines analyzed in those studies were small. In this study, we investigated the variation in the beam data for 21 TrueBeam machines collected from 17 institutions with their agreement. In the exponential regions, the percent depth dose (PDD) values showed very small variation, <1% for all the photon energies analyzed. Similarly, the off‐center ratio (OCR) values also showed small variation for all energies. In the field regions, the standard deviations of the values of dose difference (DD) between the data for each machine and the study average were <1% for field sizes ≥100 × 100 mm 2 . The maximum distance‐to‐agreement from the average data was <0.5 mm in the penumbra regions. The output factor (OPF) values also showed very small variation (<1%) for all energies and field sizes. Both the PDD and OCR of the average study data showed good agreement with the vendor's RBD for field sizes ≥100 × 100 mm 2 . The OPF of the average study data also showed good agreement with the vendor's RBD for all field sizes. However, although all the institutions used ionization chambers with similar cavity volumes, the 30 × 30 mm 2 field size showed large DD variations (≥2%) in OCR in the field regions. We conclude that the intermachine variability of TrueBeam linear accelerators was very small except for small field dosimetry, supporting the validity of the use of the RBD for clinical applications. The use of the vendor's RBD might greatly facilitate the quick installation of a new linear accelerator. … (more)
- Is Part Of:
- Journal of applied clinical medical physics. Volume 20:Issue 2(2019)
- Journal:
- Journal of applied clinical medical physics
- Issue:
- Volume 20:Issue 2(2019)
- Issue Display:
- Volume 20, Issue 2 (2019)
- Year:
- 2019
- Volume:
- 20
- Issue:
- 2
- Issue Sort Value:
- 2019-0020-0002-0000
- Page Start:
- 51
- Page End:
- 62
- Publication Date:
- 2019-01-13
- Subjects:
- commissioning -- linear accelerator -- representative beam data
Medical physics -- Periodicals
Clinical medicine -- Periodicals
Health Physics
Clinical Medicine
Electronic journals
Periodicals
Periodicals
Fulltext
Internet Resources
610.153 - Journal URLs:
- http://aapm.onlinelibrary.wiley.com/hub/journal/10.1002/(ISSN)1526-9914/ ↗
http://bibpurl.oclc.org/web/7294 ↗
http://www.jacmp.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/acm2.12518 ↗
- Languages:
- English
- ISSNs:
- 1526-9914
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9522.xml