Calculation of kQ factors for Farmer-type ionization chambers following the recent recommendations on new key dosimetry data. (January 2019)
- Record Type:
- Journal Article
- Title:
- Calculation of kQ factors for Farmer-type ionization chambers following the recent recommendations on new key dosimetry data. (January 2019)
- Main Title:
- Calculation of kQ factors for Farmer-type ionization chambers following the recent recommendations on new key dosimetry data
- Authors:
- Pimpinella, Maria
Silvi, Luca
Pinto, Massimo - Abstract:
- Highlights: kQ factors are calculated by Monte Carlo using up-to-date basic data. The influence of Monte Carlo simulation modality on kQ values is investigated. Impact of ICRU 90 recommendations on kQ calculation is evaluated. Up-to-date MC kQ values are compared with values provided by dosimetry protocols. Abstract: Purpose: To calculate by Monte Carlo simulations kQ factors for Farmer-type ionization chambers in megavoltage photon beams using the new key dosimetry data recommended by the International Commission on Radiation Units and Measurements (ICRU) Report 90. Methods: Monte Carlo calculations were performed with the EGSnrc code system using both the ICRU 90 and the ICRU 37 data. Farmer-type ionization chambers with graphite and plastic walls and with graphite wall and a plastic waterproofing sleeve were considered (Nuclear Enterprise NE 2571, IBA FC65-G and FC65-P). kQ factors were calculated for photon beams in the range 6–25 MV using phase-space files as input radiation sources. The photon beam qualities in terms of TPR20, 10 and %dd (10)x were established by simulating the depth-dose curves in water. Absorbed doses to the air cavity and to water were calculated using the egs_chamber user code with a target statistical uncertainty below 0.1%. Results: The update of key dosimetry data according to the ICRU report 90 had an impact of −0.2% in the absorbed dose to water and up to 0.5% in the absorbed dose to the air cavity. Nevertheless, changes partially offset eachHighlights: kQ factors are calculated by Monte Carlo using up-to-date basic data. The influence of Monte Carlo simulation modality on kQ values is investigated. Impact of ICRU 90 recommendations on kQ calculation is evaluated. Up-to-date MC kQ values are compared with values provided by dosimetry protocols. Abstract: Purpose: To calculate by Monte Carlo simulations kQ factors for Farmer-type ionization chambers in megavoltage photon beams using the new key dosimetry data recommended by the International Commission on Radiation Units and Measurements (ICRU) Report 90. Methods: Monte Carlo calculations were performed with the EGSnrc code system using both the ICRU 90 and the ICRU 37 data. Farmer-type ionization chambers with graphite and plastic walls and with graphite wall and a plastic waterproofing sleeve were considered (Nuclear Enterprise NE 2571, IBA FC65-G and FC65-P). kQ factors were calculated for photon beams in the range 6–25 MV using phase-space files as input radiation sources. The photon beam qualities in terms of TPR20, 10 and %dd (10)x were established by simulating the depth-dose curves in water. Absorbed doses to the air cavity and to water were calculated using the egs_chamber user code with a target statistical uncertainty below 0.1%. Results: The update of key dosimetry data according to the ICRU report 90 had an impact of −0.2% in the absorbed dose to water and up to 0.5% in the absorbed dose to the air cavity. Nevertheless, changes partially offset each other when entering in kQ as ratio, and the final impact on the kQ values was below 0.3%. Conclusions: The calculated values of kQ tend to be lower than the current values in the IAEA TRS-398 protocol with differences up to about 0.5%. A slightly better agreement (within 0.3%) is observed with the Monte-Carlo calculated values provided by the addendum to the AAPM's TG-51 protocol. … (more)
- Is Part Of:
- Physica medica. Volume 57(2019)
- Journal:
- Physica medica
- Issue:
- Volume 57(2019)
- Issue Display:
- Volume 57, Issue 2019 (2019)
- Year:
- 2019
- Volume:
- 57
- Issue:
- 2019
- Issue Sort Value:
- 2019-0057-2019-0000
- Page Start:
- 221
- Page End:
- 230
- Publication Date:
- 2019-01
- Subjects:
- Beam quality correction factor -- kQ -- ICRU 90 -- Monte-Carlo calculated kQ -- Radiotherapy dosimetry
Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2018.12.006 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9503.xml