SAPIEN valve for percutaneous transcatheter pulmonary valve replacement without "pre‐stenting": A multi‐institutional experience. Issue 2 (23rd October 2018)
- Record Type:
- Journal Article
- Title:
- SAPIEN valve for percutaneous transcatheter pulmonary valve replacement without "pre‐stenting": A multi‐institutional experience. Issue 2 (23rd October 2018)
- Main Title:
- SAPIEN valve for percutaneous transcatheter pulmonary valve replacement without "pre‐stenting": A multi‐institutional experience
- Authors:
- Morgan, Gareth J.
Sadeghi, Soraya
Salem, Moris M.
Wilson, Neil
Kay, Joseph
Rothman, Abraham
Galindo, Alvaro
Martin, Mary Hunt
Gray, Robert
Ross, Michael
Aboulhosn, Jamil A.
Levi, Daniel Steven - Abstract:
- Abstract: Objectives: To describe a multicenter experience of percutaneous transcatheter pulmonary valve replacement (TPVR) using the Edwards SAPIEN S3 valve without the use of a prior stent ("pre‐stenting"). Background: The SAPIEN S3 and XT valves have durable cobalt–chromium stent frames which may allow for TPVR in large diameter dysfunctional right ventricular outflow tracts (RVOTs) without pre‐stenting the landing zone. Methods: A retrospective review was performed of all patients with Congenital Heart Disease and dysfunctional RVOT who underwent TPVR using the SAPIEN valve without the use of a pre‐stent. Imaging data, procedural elements, and clinical follow‐up data were collected to evaluate short and intermediate‐term results. Results: Fifty‐seven patients underwent percutaneous placement of the SAPIEN valve in the pulmonary position without the use of pre‐stenting. The anatomical substrate varied: native RVOTs ( n = 41), conduits ( n = 10), and bioprosthetic valves ( n = 6). There were no cases in which the valve could not be implanted and no cases of valve embolization or misplacement. On follow‐up (range 1 month to 2.2 years, median 5.3 months), no patients had significant obstruction or regurgitation around the valve. There were no frame fractures. There were no procedural deaths. Major complications included severe aortic compression ( n = 1) requiring surgical explantation and tricuspid valve injury requiring surgical intervention ( n = 2). Conclusions: ThisAbstract: Objectives: To describe a multicenter experience of percutaneous transcatheter pulmonary valve replacement (TPVR) using the Edwards SAPIEN S3 valve without the use of a prior stent ("pre‐stenting"). Background: The SAPIEN S3 and XT valves have durable cobalt–chromium stent frames which may allow for TPVR in large diameter dysfunctional right ventricular outflow tracts (RVOTs) without pre‐stenting the landing zone. Methods: A retrospective review was performed of all patients with Congenital Heart Disease and dysfunctional RVOT who underwent TPVR using the SAPIEN valve without the use of a pre‐stent. Imaging data, procedural elements, and clinical follow‐up data were collected to evaluate short and intermediate‐term results. Results: Fifty‐seven patients underwent percutaneous placement of the SAPIEN valve in the pulmonary position without the use of pre‐stenting. The anatomical substrate varied: native RVOTs ( n = 41), conduits ( n = 10), and bioprosthetic valves ( n = 6). There were no cases in which the valve could not be implanted and no cases of valve embolization or misplacement. On follow‐up (range 1 month to 2.2 years, median 5.3 months), no patients had significant obstruction or regurgitation around the valve. There were no frame fractures. There were no procedural deaths. Major complications included severe aortic compression ( n = 1) requiring surgical explantation and tricuspid valve injury requiring surgical intervention ( n = 2). Conclusions: This limited multi‐institutional experience demonstrates that the SAPIEN valve can be used for TPVR without the use of a pre‐stent without medium‐term risk of frame fracture, paravalvar leak, or embolization. Longer term follow‐up is required to fully assess this method. … (more)
- Is Part Of:
- Catheterization and cardiovascular interventions. Volume 93:Issue 2(2019)
- Journal:
- Catheterization and cardiovascular interventions
- Issue:
- Volume 93:Issue 2(2019)
- Issue Display:
- Volume 93, Issue 2 (2019)
- Year:
- 2019
- Volume:
- 93
- Issue:
- 2
- Issue Sort Value:
- 2019-0093-0002-0000
- Page Start:
- 324
- Page End:
- 329
- Publication Date:
- 2018-10-23
- Subjects:
- percutaneous pulmonary valve replacement -- right ventricular outflow tract stenting -- stent fracture
Heart -- Diseases -- Diagnosis -- Periodicals
Cardiac catheterization -- Periodicals
616.1207572 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1522-726X ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/ccd.27932 ↗
- Languages:
- English
- ISSNs:
- 1522-1946
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3092.992000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 9478.xml