Assessing the feasibility of adaptive planning for prostate radiotherapy using Smartadapt deformable image registration. (February 2019)
- Record Type:
- Journal Article
- Title:
- Assessing the feasibility of adaptive planning for prostate radiotherapy using Smartadapt deformable image registration. (February 2019)
- Main Title:
- Assessing the feasibility of adaptive planning for prostate radiotherapy using Smartadapt deformable image registration
- Authors:
- Marin Anaya, Virginia
Fairfoul, Jamie - Abstract:
- Highlights: A method for geometric and dosimetric analysis of SmartAdapt is proposed. Geometric results for deformable registration are worse than for rigid registration. SmartAdapt tends to under-contour the prostate and rectum. The deformed prostate tends to overestimate the need for re-planning. The deformed rectum tends to underestimate the need for re-planning. Abstract: Background: The purpose of this study is to assess the feasibility of adaptive planning for prostate cancer using SmartAdapt deformable image registration (DIR), in an effort to reduce the effect of patient-specific variations. Methods: 18 prostate patients (74 Gy in 37 fractions) were selected. Each patient had a planning CT and cone-beam CT (CBCT) images acquired at different fractions. For each fraction, the planning CT was registered to the CBCT image dataset using SmartAdapt. A rigid registration was performed prior to the DIR. The manual contours on the CBCT images by the clinician were used as reference. For the geometric analysis, various evaluation metrics were used to compare the volumes for both rigid and DIR. For the dosimetric analysis, the initial treatment plan was recalculated on the smartadapted structures, using Eclipse. Several Dose-Volume Histogram (DVH) points were used for the evaluation. Results: The results of the geometric analysis indicate that DIR can result in worse evaluation metrics than rigid registration alone. In general, SmartAdapt tends to under-contour the volumes.Highlights: A method for geometric and dosimetric analysis of SmartAdapt is proposed. Geometric results for deformable registration are worse than for rigid registration. SmartAdapt tends to under-contour the prostate and rectum. The deformed prostate tends to overestimate the need for re-planning. The deformed rectum tends to underestimate the need for re-planning. Abstract: Background: The purpose of this study is to assess the feasibility of adaptive planning for prostate cancer using SmartAdapt deformable image registration (DIR), in an effort to reduce the effect of patient-specific variations. Methods: 18 prostate patients (74 Gy in 37 fractions) were selected. Each patient had a planning CT and cone-beam CT (CBCT) images acquired at different fractions. For each fraction, the planning CT was registered to the CBCT image dataset using SmartAdapt. A rigid registration was performed prior to the DIR. The manual contours on the CBCT images by the clinician were used as reference. For the geometric analysis, various evaluation metrics were used to compare the volumes for both rigid and DIR. For the dosimetric analysis, the initial treatment plan was recalculated on the smartadapted structures, using Eclipse. Several Dose-Volume Histogram (DVH) points were used for the evaluation. Results: The results of the geometric analysis indicate that DIR can result in worse evaluation metrics than rigid registration alone. In general, SmartAdapt tends to under-contour the volumes. The difference in DVH points between the deformed and the reference prostate was worse than the difference between the planning CT and the reference, overestimating the need for re-planning. However, there was a tendency for the deformed rectum to underestimate the need for re-planning. Conclusions: After cautious analysis of the results, it has been decided not to implement SmartAdapt clinically to trigger adaptive decisions … (more)
- Is Part Of:
- Medical engineering & physics. Volume 64(2019)
- Journal:
- Medical engineering & physics
- Issue:
- Volume 64(2019)
- Issue Display:
- Volume 64, Issue 2019 (2019)
- Year:
- 2019
- Volume:
- 64
- Issue:
- 2019
- Issue Sort Value:
- 2019-0064-2019-0000
- Page Start:
- 65
- Page End:
- 73
- Publication Date:
- 2019-02
- Subjects:
- Adaptive planning -- Deformable image registration -- Prostate -- Cone-beam CT
Biomedical engineering -- Periodicals
Biomedical Engineering -- Periodicals
Physics -- Periodicals
Génie biomédical -- Périodiques
Biomedical engineering
Electronic journals
Periodicals
610.28 - Journal URLs:
- http://www.medengphys.com ↗
http://www.sciencedirect.com/science/journal/13504533 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/13504533 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/13504533 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.medengphy.2019.01.001 ↗
- Languages:
- English
- ISSNs:
- 1350-4533
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5527.323000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9479.xml