54. Spatio-temporal generalisation of Model Observer for Low Contrast Detectability assessment of dynamic angiographic images: comparison with an innovative Statistical method and 2AFC experiments. (December 2018)
- Record Type:
- Journal Article
- Title:
- 54. Spatio-temporal generalisation of Model Observer for Low Contrast Detectability assessment of dynamic angiographic images: comparison with an innovative Statistical method and 2AFC experiments. (December 2018)
- Main Title:
- 54. Spatio-temporal generalisation of Model Observer for Low Contrast Detectability assessment of dynamic angiographic images: comparison with an innovative Statistical method and 2AFC experiments
- Authors:
- Villa, R.
Paruccini, N.
Spadavecchia, C.
Baglivi, A.
Signoriello, M.
Crespi, A. - Abstract:
- Abstract : Purpose: Evaluate the feasibility of different less time consuming approaches for protocol optimisation in interventional radiology. Methods: A Allura FD20 (Philips, The Netherlands) has been employed in this study. Low Dose Fluoroscopy was selected due to its spatial resolution, characterised by a lower enhancement compared to other modalities. Simulated images have been obtained adding a blurred circular insert to a series of acquired homogeneous images. A total of 3 insert dimensions (1.25, 2, 3 mm of diameter), 25 contrast levels and 4 different visualised frame rate (5, 10, 15 and 30 fps) have been selected to characterise low contrast detectability. A series of 2AFC experiments have been performed by two observers. Two different Model Observers has been considered: Non-PreWhitening Eye filter (NPWE) with spatio-temporal contrast sensitivity function and Channelized Hotelling Observer (CHO) with spatio-temporal Gabor filters. Furthermore, Low Contrast Detectability has been evaluated with a statistical method using uniform images. This analysis has been performed on an averaging of frames depending from the visualised fps. In particular, the averaging process was performed on groups of frames that correspond to the noise integration time of human eye-brain system of 200 ms, e.g. 3 frames with 15 fps. Results: Analysis with both Model Observers and 2AFC experiments suffered from the general low quality of the protocol considered (Fig. 1). Nevertheless, a goodAbstract : Purpose: Evaluate the feasibility of different less time consuming approaches for protocol optimisation in interventional radiology. Methods: A Allura FD20 (Philips, The Netherlands) has been employed in this study. Low Dose Fluoroscopy was selected due to its spatial resolution, characterised by a lower enhancement compared to other modalities. Simulated images have been obtained adding a blurred circular insert to a series of acquired homogeneous images. A total of 3 insert dimensions (1.25, 2, 3 mm of diameter), 25 contrast levels and 4 different visualised frame rate (5, 10, 15 and 30 fps) have been selected to characterise low contrast detectability. A series of 2AFC experiments have been performed by two observers. Two different Model Observers has been considered: Non-PreWhitening Eye filter (NPWE) with spatio-temporal contrast sensitivity function and Channelized Hotelling Observer (CHO) with spatio-temporal Gabor filters. Furthermore, Low Contrast Detectability has been evaluated with a statistical method using uniform images. This analysis has been performed on an averaging of frames depending from the visualised fps. In particular, the averaging process was performed on groups of frames that correspond to the noise integration time of human eye-brain system of 200 ms, e.g. 3 frames with 15 fps. Results: Analysis with both Model Observers and 2AFC experiments suffered from the general low quality of the protocol considered (Fig. 1). Nevertheless, a good agreement (<8%) can be found considering the effects of internal eye noise. Results with the Statistical Method show different degree of accordance related to insert dimension; thus, a single correction factor can be used to obtain an agreement within 8% for every protocols considered (Fig. 1). Conclusions: The agreement with human observer experiments underlines the feasibility of the proposed generalisations. Thus, they could be introduced for a deeper modalities characterisation, in terms of image quality, or for clinical protocol optimization involving cine images. … (more)
- Is Part Of:
- Physica medica. Volume 56(2018)Supplement 2
- Journal:
- Physica medica
- Issue:
- Volume 56(2018)Supplement 2
- Issue Display:
- Volume 56, Issue 2 (2018)
- Year:
- 2018
- Volume:
- 56
- Issue:
- 2
- Issue Sort Value:
- 2018-0056-0002-0000
- Page Start:
- 97
- Page End:
- Publication Date:
- 2018-12
- Subjects:
- Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2018.04.064 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9461.xml