153 Dose monitoring in synchrotron radiation breast computed tomography with radiochromic films. (December 2018)
- Record Type:
- Journal Article
- Title:
- 153 Dose monitoring in synchrotron radiation breast computed tomography with radiochromic films. (December 2018)
- Main Title:
- 153 Dose monitoring in synchrotron radiation breast computed tomography with radiochromic films
- Authors:
- Masi, M.
Di Lillo, F.
Mettivier, G.
Sarno, A.
Arfelli, F.
Brombal, L.
Donato, S.
Fedon, C.
Rigon, L.
Russo, P. - Abstract:
- Abstract : Purpose: To measure the dose distribution in a phantom during a synchrotron radiation (SR) computed tomography (CT) scan using radiochromic films, for procedure assessment and for offline dose monitoring in SR breast CT. Methods: We used XR-QA2 radiochromic films (calibrated in terms of free air kerma) and a 14-cm PMMA cylindrical phantom. Measurements were performed at the SYRMEP beamline of ELETTRA facility, within the SYRMA-3D project for phase-contrast breast CT[1] . The phantom was positioned at isocenter, pending from the patient bed. Measurements were performed at 38 keV, beam size of 3 mm × 150 mm (H × W), at an air kerma corresponding to a mean glandular dose of 5 mGy in a 14-cm diameter 50% glandular breast. The phantom was irradiated over 360° first with a film at the mid-plane, and then with film strips wrapped on its surface. Additional tests were made over 180° with an axial scan pitch of 1, 0.8 or 0.5. Results: Results show a dose map at mid-plane the phantom, indicating the cup-shaped horizontal profile due to primary and scatter dose. Analogous maps were recorded with the films at the periphery of the phantom, representing the surface dose maps. Conclusions: We performed measurements of the dose distribution in a phantom during a SR CT scan with different irradiation geometries. This technique produces evidence for selecting the suitable scan procedure. Moreover, as an offline monitor system, a piece of calibrated radiochromic film wrapped on aAbstract : Purpose: To measure the dose distribution in a phantom during a synchrotron radiation (SR) computed tomography (CT) scan using radiochromic films, for procedure assessment and for offline dose monitoring in SR breast CT. Methods: We used XR-QA2 radiochromic films (calibrated in terms of free air kerma) and a 14-cm PMMA cylindrical phantom. Measurements were performed at the SYRMEP beamline of ELETTRA facility, within the SYRMA-3D project for phase-contrast breast CT[1] . The phantom was positioned at isocenter, pending from the patient bed. Measurements were performed at 38 keV, beam size of 3 mm × 150 mm (H × W), at an air kerma corresponding to a mean glandular dose of 5 mGy in a 14-cm diameter 50% glandular breast. The phantom was irradiated over 360° first with a film at the mid-plane, and then with film strips wrapped on its surface. Additional tests were made over 180° with an axial scan pitch of 1, 0.8 or 0.5. Results: Results show a dose map at mid-plane the phantom, indicating the cup-shaped horizontal profile due to primary and scatter dose. Analogous maps were recorded with the films at the periphery of the phantom, representing the surface dose maps. Conclusions: We performed measurements of the dose distribution in a phantom during a SR CT scan with different irradiation geometries. This technique produces evidence for selecting the suitable scan procedure. Moreover, as an offline monitor system, a piece of calibrated radiochromic film wrapped on a suitable holder around the breast could record the history of the organ irradiation. … (more)
- Is Part Of:
- Physica medica. Volume 56(2018)Supplement 2
- Journal:
- Physica medica
- Issue:
- Volume 56(2018)Supplement 2
- Issue Display:
- Volume 56, Issue 2 (2018)
- Year:
- 2018
- Volume:
- 56
- Issue:
- 2
- Issue Sort Value:
- 2018-0056-0002-0000
- Page Start:
- 158
- Page End:
- Publication Date:
- 2018-12
- Subjects:
- Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2018.04.164 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9460.xml