Development of custom RF coils for use in a small animal platform for magnetic resonance-guided focused ultrasound hyperthermia compatible with a clinical MRI scanner. (1st January 2019)
- Record Type:
- Journal Article
- Title:
- Development of custom RF coils for use in a small animal platform for magnetic resonance-guided focused ultrasound hyperthermia compatible with a clinical MRI scanner. (1st January 2019)
- Main Title:
- Development of custom RF coils for use in a small animal platform for magnetic resonance-guided focused ultrasound hyperthermia compatible with a clinical MRI scanner
- Authors:
- Abraham, Christopher Brian
Loree-Spacek, Jak
Andrew Drainville, Robert
Pichardo, Samuel
Curiel, Laura - Abstract:
- Abstract: Three different magnetic resonance imaging (MRI) coils were developed and assessed for use with an experimental platform designed to generate hyperthermia in mice using magnetic resonance-guided focused ultrasound (MRgFUS). An ergonomic animal treatment bed was integrated with MRI coils. Three different coil designs optimized for small targets were tested, and performance in targeting and conducting accurate temperature imaging was evaluated. Two transmit/receive surface coils of different diameters (4 and 7 cm) and a transmit-only/receive-only (TORO) coil were used. A software platform was developed to provide real-time targeting and temperature maps and to deliver controlled ultrasound exposure. MR thermometry was conducted on different targets, including fresh chicken breasts and mouse cadavers. Multiple experiments were performed in which tissues were targeted with high reproducibility. The TORO coil was the most resilient to temperature drift, resulting in an increase in the calculated temperature of 0.29 ± 0.12 °C, compared to 1.27 ± 0.13 °C and 0.47 ± 0.04 °C for the medium and small coils, respectively. Controlled closed-loop hyperthermia exposure was successfully performed with all three coils. Considering all assessments, the TORO coil exhibited the best overall performance for thermometry acquisition when accounting for stability, precision, temperature spread and resilience to temperature drift. B1 maps of the three coils confirmed that the TORO coilAbstract: Three different magnetic resonance imaging (MRI) coils were developed and assessed for use with an experimental platform designed to generate hyperthermia in mice using magnetic resonance-guided focused ultrasound (MRgFUS). An ergonomic animal treatment bed was integrated with MRI coils. Three different coil designs optimized for small targets were tested, and performance in targeting and conducting accurate temperature imaging was evaluated. Two transmit/receive surface coils of different diameters (4 and 7 cm) and a transmit-only/receive-only (TORO) coil were used. A software platform was developed to provide real-time targeting and temperature maps and to deliver controlled ultrasound exposure. MR thermometry was conducted on different targets, including fresh chicken breasts and mouse cadavers. Multiple experiments were performed in which tissues were targeted with high reproducibility. The TORO coil was the most resilient to temperature drift, resulting in an increase in the calculated temperature of 0.29 ± 0.12 °C, compared to 1.27 ± 0.13 °C and 0.47 ± 0.04 °C for the medium and small coils, respectively. Controlled closed-loop hyperthermia exposure was successfully performed with all three coils. Considering all assessments, the TORO coil exhibited the best overall performance for thermometry acquisition when accounting for stability, precision, temperature spread and resilience to temperature drift. B1 maps of the three coils confirmed that the TORO coil exhibited the most homogeneous B1 field, which explained the improved thermometry performance. The use of coils specifically designed for small targets within the proposed experimental platform allowed accurate thermometry during hyperthermia. … (more)
- Is Part Of:
- International journal of hyperthermia. Volume 35:Number 1(2018)
- Journal:
- International journal of hyperthermia
- Issue:
- Volume 35:Number 1(2018)
- Issue Display:
- Volume 35, Issue 1 (2018)
- Year:
- 2018
- Volume:
- 35
- Issue:
- 1
- Issue Sort Value:
- 2018-0035-0001-0000
- Page Start:
- 348
- Page End:
- 360
- Publication Date:
- 2019-01-01
- Subjects:
- MRgFUS -- MR thermometry -- ultrasound hyperthermia -- MRI small animal coils -- small animal-focused ultrasound
Thermotherapy -- Periodicals
615.832 - Journal URLs:
- http://informahealthcare.com/loi/hth ↗
http://www.tandf.co.uk/journals/titles/02656736.asp ↗
http://informahealthcare.com ↗ - DOI:
- 10.1080/02656736.2018.1503344 ↗
- Languages:
- English
- ISSNs:
- 0265-6736
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4542.297000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 9427.xml