Mechanical performance of conical implant-abutment connections under different cyclic loading conditions. (February 2019)
- Record Type:
- Journal Article
- Title:
- Mechanical performance of conical implant-abutment connections under different cyclic loading conditions. (February 2019)
- Main Title:
- Mechanical performance of conical implant-abutment connections under different cyclic loading conditions
- Authors:
- Yao, Kuang-Ta
Kao, Hung-Chan
Cheng, Cheng-Kung
Fang, Hsu-Wei
Huang, Chang-Hung
Hsu, Ming-Lun - Abstract:
- Abstract: Objectives: Conical implant-abutment connections are popular for its anti-bending performance; on the other hand, the torsional and axial forces also play important roles in occlusion. However, so far there were scarce studies on their effects on connection stability. Therefore, this study seeks to investigate the mechanical performance of conical connections under different cyclic loading conditions. Methods: 15 conical implant-abutment assembles (Cowell Medi, Busan, South Korea) were divided into 3 groups according to different cyclic loadings. In group BTA, the loading condition of the posterior occlusion was simulated (20–200 N, 30° off-axis and 4 mm eccentric to implant axis), generating a bending moment, a torsional moment, and an axial loading. In group BT, a bending moment and a torsional moment of the posterior occlusion were applied (10–100 N, 90° off-axis and 4 mm eccentric to implant axis). In group B, only a bending moment was applied (10–100 N, 90° off-axis and through implant axis). The fatigue testing machine ran at 10 Hz until failure, or to the upper limit of 10 6 cycles. The fatigue cycles and failure modes were recorded. Besides, the value of the torque loss of the abutment screw, the difference between initial torque and post-load reverse torque, was calculated. The data were statistically analyzed. Morphologies of the abutment conical surface were examined by scanning electron microscopy. Results: In group B and BTA, all samples passed theAbstract: Objectives: Conical implant-abutment connections are popular for its anti-bending performance; on the other hand, the torsional and axial forces also play important roles in occlusion. However, so far there were scarce studies on their effects on connection stability. Therefore, this study seeks to investigate the mechanical performance of conical connections under different cyclic loading conditions. Methods: 15 conical implant-abutment assembles (Cowell Medi, Busan, South Korea) were divided into 3 groups according to different cyclic loadings. In group BTA, the loading condition of the posterior occlusion was simulated (20–200 N, 30° off-axis and 4 mm eccentric to implant axis), generating a bending moment, a torsional moment, and an axial loading. In group BT, a bending moment and a torsional moment of the posterior occlusion were applied (10–100 N, 90° off-axis and 4 mm eccentric to implant axis). In group B, only a bending moment was applied (10–100 N, 90° off-axis and through implant axis). The fatigue testing machine ran at 10 Hz until failure, or to the upper limit of 10 6 cycles. The fatigue cycles and failure modes were recorded. Besides, the value of the torque loss of the abutment screw, the difference between initial torque and post-load reverse torque, was calculated. The data were statistically analyzed. Morphologies of the abutment conical surface were examined by scanning electron microscopy. Results: In group B and BTA, all samples passed the test (10 6 cycles). While, in group BT, all abutments generated rotation within 140 cycles, showing significant differences compared to the other two groups ( p < 0.001). However, from SEM observations, both group B and BT showed marked fretting wear, indicating obvious micromotion in the connection. Whereas group BTA showed indentation of tight contact, attributed to the axial loading. In terms of the torque loss of the abutment screw, the torque loss in group BT was much more than the other two groups with statistically significant differences (p < 0.05). Conclusion: Owing to the effect of the bending moment, marked fretting wear was generated in the conical connections and further led to loss of the anti-torsional ability. However, adding an axial loading could improve their anti-torsional ability significantly. Graphical abstract: Highlights: Effects of different loading conditions on conical connections are verified. Bending moments generate fretting wear and undermine conical connection stability. Adding axial loadings counteracts bending moments and stabilizes the connection. … (more)
- Is Part Of:
- Journal of the mechanical behavior of biomedical materials. Volume 90(2019)
- Journal:
- Journal of the mechanical behavior of biomedical materials
- Issue:
- Volume 90(2019)
- Issue Display:
- Volume 90, Issue 2019 (2019)
- Year:
- 2019
- Volume:
- 90
- Issue:
- 2019
- Issue Sort Value:
- 2019-0090-2019-0000
- Page Start:
- 426
- Page End:
- 432
- Publication Date:
- 2019-02
- Subjects:
- Implant-abutment connection -- Conical -- Cyclic loading -- Fretting wear -- SEM
Biomedical materials -- Periodicals
Biomedical materials -- Mechanical properties -- Periodicals
Biomedical materials
Biomedical materials -- Mechanical properties
Periodicals
Electronic journals
610.28 - Journal URLs:
- http://www.sciencedirect.com/science/journal/17516161 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.jmbbm.2018.10.039 ↗
- Languages:
- English
- ISSNs:
- 1751-6161
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5015.809000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9395.xml