Optimal positioning of self‐expanding valves before deployment decreases paravalvular regurgitation following transcatheter aortic valve replacement. Issue 1 (14th September 2018)
- Record Type:
- Journal Article
- Title:
- Optimal positioning of self‐expanding valves before deployment decreases paravalvular regurgitation following transcatheter aortic valve replacement. Issue 1 (14th September 2018)
- Main Title:
- Optimal positioning of self‐expanding valves before deployment decreases paravalvular regurgitation following transcatheter aortic valve replacement
- Authors:
- Dadu, Razvan T.
Xu, Jiaqiong
Rehman, Hasan
von Ballmoos, Moritz Wyler
Barker, Colin M.
Reyes, Manuel
Ramchandani, Mahesh
Reul, Ross M.
Reardon, Michael J.
Kleiman, Neal S. - Abstract:
- Abstract: Objectives: To evaluate the association between measurements performed during Medtronic CoreValve (MCV) deployment and paravalvular leak (PVL). Background: The MCV can be recaptured and repositioned, allowing the TAVR operator to implant at a more favorable position. The association between angiographic measurements of MCV position while the valve is recapturable and PVL post deployment has not been investigated. Methods: 493 patients undergoing TAVR with MCV (January 2011–July 2017) were included. PVL was defined as intraprocedural aortic regurgitation that was judged clinically to require balloon postdilation. Depth of the valve at the left coronary cusp (LCC) and noncoronary cusp (NCC) were measured when the valve was 80% deployed. An optimal cutoff value for the ratio LCC/NCC for PVL was identified in 40 patients. Using this cutoff value, the association between LCC/NCC and PVL was then validated in 453 patients. Results: The median LCC/NCC was 1.51 (interquartile range 1.06–1.89).The optimal cutoff value for LCC/NCC was 1.48 (93% sensitivity, 77% specificity, AUC0.85). In the validation group 112 (24.7%) patients had PVL. For LCC/NCC ≥ 1.48, the incidence of PVL was lower compared to LCC/NCC < 1.48 (9.58% vs. 41.78%, P < 0.0001). LCC/NCC of 1.48 had a sensitivity of 79.5% and specificity of 63.6% for PVL (AUC0.72). In a multivariate model, LCC/NCC < 1.48 independently predicted PVL (OR = 6.67, 95% CI 3.96–11.23, P < 0.0001). Conclusion: Positioning the MCVAbstract: Objectives: To evaluate the association between measurements performed during Medtronic CoreValve (MCV) deployment and paravalvular leak (PVL). Background: The MCV can be recaptured and repositioned, allowing the TAVR operator to implant at a more favorable position. The association between angiographic measurements of MCV position while the valve is recapturable and PVL post deployment has not been investigated. Methods: 493 patients undergoing TAVR with MCV (January 2011–July 2017) were included. PVL was defined as intraprocedural aortic regurgitation that was judged clinically to require balloon postdilation. Depth of the valve at the left coronary cusp (LCC) and noncoronary cusp (NCC) were measured when the valve was 80% deployed. An optimal cutoff value for the ratio LCC/NCC for PVL was identified in 40 patients. Using this cutoff value, the association between LCC/NCC and PVL was then validated in 453 patients. Results: The median LCC/NCC was 1.51 (interquartile range 1.06–1.89).The optimal cutoff value for LCC/NCC was 1.48 (93% sensitivity, 77% specificity, AUC0.85). In the validation group 112 (24.7%) patients had PVL. For LCC/NCC ≥ 1.48, the incidence of PVL was lower compared to LCC/NCC < 1.48 (9.58% vs. 41.78%, P < 0.0001). LCC/NCC of 1.48 had a sensitivity of 79.5% and specificity of 63.6% for PVL (AUC0.72). In a multivariate model, LCC/NCC < 1.48 independently predicted PVL (OR = 6.67, 95% CI 3.96–11.23, P < 0.0001). Conclusion: Positioning the MCV such that the LCC/NCC is ≥1.48 may result in less PVL. … (more)
- Is Part Of:
- Catheterization and cardiovascular interventions. Volume 93:Issue 1(2019)
- Journal:
- Catheterization and cardiovascular interventions
- Issue:
- Volume 93:Issue 1(2019)
- Issue Display:
- Volume 93, Issue 1 (2019)
- Year:
- 2019
- Volume:
- 93
- Issue:
- 1
- Issue Sort Value:
- 2019-0093-0001-0000
- Page Start:
- 149
- Page End:
- 155
- Publication Date:
- 2018-09-14
- Subjects:
- aortic stenosis -- core valve -- medtronic -- paravalvular aortic regurgitation -- transcatheter aortic valve replacement
Heart -- Diseases -- Diagnosis -- Periodicals
Cardiac catheterization -- Periodicals
616.1207572 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1522-726X ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/ccd.27829 ↗
- Languages:
- English
- ISSNs:
- 1522-1946
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3092.992000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 9370.xml