Dose response of selected solid state detectors in applied homogeneous transverse and longitudinal magnetic fields. Issue 9 (20th August 2014)
- Record Type:
- Journal Article
- Title:
- Dose response of selected solid state detectors in applied homogeneous transverse and longitudinal magnetic fields. Issue 9 (20th August 2014)
- Main Title:
- Dose response of selected solid state detectors in applied homogeneous transverse and longitudinal magnetic fields
- Authors:
- Reynolds, M.
Fallone, B. G.
Rathee, S. - Abstract:
- Abstract : Purpose: : MR‐Linac devices under development worldwide will require standard calibration, commissioning, and quality assurance. Solid state radiation detectors are often used for dose profiles and percent depth dose measurements. The dose response of selected solid state detectors is therefore evaluated in varying transverse and longitudinal magnetic fields for this purpose. Methods: : The Monte Carlo code PENELOPE was used to model irradiation of a PTW 60003 diamond detector and IBA PFD diode detector in the presence of a magnetic field. The field itself was varied in strength, and oriented both transversely and longitudinally with respect to the incident photon beam. The long axis of the detectors was oriented either parallel or perpendicular to the photon beam. The dose to the active volume of each detector in air was scored, and its ratio to dose with zero magnetic field strength was determined as the "dose response" in magnetic field. Measurements at low fields for both detectors in transverse magnetic fields were taken to evaluate the accuracy of the simulations. Additional simulations were performed in a water phantom to obtain few representative points for beam profile and percent depth dose measurements. Results: : Simulations show significant dose response as a function of magnetic field in transverse field geometries. This response can be near 20% at 1.5 T, and it is highly dependent on the detectors' relative orientation to the magnetic field, theAbstract : Purpose: : MR‐Linac devices under development worldwide will require standard calibration, commissioning, and quality assurance. Solid state radiation detectors are often used for dose profiles and percent depth dose measurements. The dose response of selected solid state detectors is therefore evaluated in varying transverse and longitudinal magnetic fields for this purpose. Methods: : The Monte Carlo code PENELOPE was used to model irradiation of a PTW 60003 diamond detector and IBA PFD diode detector in the presence of a magnetic field. The field itself was varied in strength, and oriented both transversely and longitudinally with respect to the incident photon beam. The long axis of the detectors was oriented either parallel or perpendicular to the photon beam. The dose to the active volume of each detector in air was scored, and its ratio to dose with zero magnetic field strength was determined as the "dose response" in magnetic field. Measurements at low fields for both detectors in transverse magnetic fields were taken to evaluate the accuracy of the simulations. Additional simulations were performed in a water phantom to obtain few representative points for beam profile and percent depth dose measurements. Results: : Simulations show significant dose response as a function of magnetic field in transverse field geometries. This response can be near 20% at 1.5 T, and it is highly dependent on the detectors' relative orientation to the magnetic field, the energy of the photon beam, and detector composition. Measurements at low transverse magnetic fields verify the simulations for both detectors in their relative orientations to radiation beam. Longitudinal magnetic fields, in contrast, show little dose response, rising slowly with magnetic field, and reaching 0.5%–1% at 1.5 T regardless of detector orientation. Water tank and in air simulation results were the same within simulation uncertainty where lateral electronic equilibrium is present and expectedly differed at the beam edge in transverse field orientations only. Due to the difference in design, the two detectors behaved differently. Conclusions: : When transverse magnetic fields are present, great care must be taken when using diamond or diode detectors. Dose response varies with relative detector orientation, magnetic field strength, and between detectors. This response can be considerable (∼20% for both detectors). Both detectors in longitudinal fields exhibit little to no dose response as a function of magnetic field. Water tank simulations seem to suggest that the diode detector is better suited to general beam commissioning, and each detector must be investigated separately. … (more)
- Is Part Of:
- Medical physics. Volume 41:Issue 9(2014)
- Journal:
- Medical physics
- Issue:
- Volume 41:Issue 9(2014)
- Issue Display:
- Volume 41, Issue 9 (2014)
- Year:
- 2014
- Volume:
- 41
- Issue:
- 9
- Issue Sort Value:
- 2014-0041-0009-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2014-08-20
- Subjects:
- Dosimetry/exposure assessment -- Monte Carlo simulations -- Solid‐state detectors
dosimetry -- Monte Carlo methods -- particle detectors -- phantoms
Monte Carlo -- Monte Carlo methods -- linac‐MR -- longitudinal and transverse linac‐MR -- linear accelerators -- biomedical MRI -- magnetic field -- dosimetry -- solid state detectors -- diamond detector -- diode detector
Measurement of nuclear or x‐radiation -- Tubes for determining the presence, intensity, density or energy of radiation or particles -- Scintigraphy
Magnetic field sensors -- Magnetic fields -- Diamond -- Particle beam detectors -- Magnetic field measurements -- Dosimetry -- Water vapor -- Linear accelerators -- Tungsten -- Photons
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4893276 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9333.xml