- Technical Note: A method for improving the calibration reproducibility of an ionization chamber detector array
- Authors:
- Goodall, Simon
Morgan, Steve - Abstract:
- Abstract : Purpose: : This paper describes an extension to a wide field calibration method implemented on a commercial detector array in order to improve the reproducibility of the calibration procedure. Methods: : Following the standard array calibration procedure, two additional 10 × 10 cm exposures were acquired for each array axis with the detector array shifted by ±10 cm in the transverse or axial axes, or by ±10 2 cm in the positive or negative diagonal axes. These exposures were compared with a final baseline 10 × 10 cm exposure captured with the detector repositioned at the isocenter. The measurements were used to calculate a linear off‐axis correction gradient which was then applied to the stored calibration factors. Results: : The mean coefficient of variation between five repeat calibrations was reduced from 4.17% to 0.48% and the maximum percentage error in individual calibration factors was reduced from 6.46% to 0.77%. Conclusions: : The reproducibility of the calibration factors of an ionization chamber array was increased by capturing a baseline exposure and two further off‐axis readings per calibration axis.
- Is Part Of:
- Medical physics. Volume 41:Issue 9(2014)
- Journal:
- Medical physics
- Issue:
- Volume 41:Issue 9(2014)
- Issue Display:
- Volume 41, Issue 9 (2014)
- Year:
- 2014
- Volume:
- 41
- Issue:
- 9
- Issue Sort Value:
- 2014-0041-0009-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2014-08-13
- Subjects:
- Dose‐volume analysis -- Standards and calibration -- Gas‐filled counters: ionization chambers, proportional, and avalanche counters
calibration -- dosimeters -- dosimetry -- ionisation chambers -- radiation therapy
calibration reproducibility -- array calibration -- radiation dosimetry
Radiation therapy -- Testing or calibrating of apparatus or arrangements provided for in groups G01D1/00 to G01D15/00 -- Measurement of nuclear or x‐radiation -- Dosimeters -- Calibrating of instruments or apparatus -- Tubes for determining the presence, intensity, density or energy of radiation or particles -- Scintigraphy -- with scintillation detectors -- Ionisation chambers
Calibration -- Inductively coupled plasma -- Linear accelerators -- Detector arrays -- Ionization chambers -- Dosimetry -- Quality assurance
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4892607 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9333.xml