Development and clinical evaluation of automatic fiducial detection for tumor tracking in cine megavoltage images during volumetric modulated arc therapy. Issue 3 (20th February 2013)
- Record Type:
- Journal Article
- Title:
- Development and clinical evaluation of automatic fiducial detection for tumor tracking in cine megavoltage images during volumetric modulated arc therapy. Issue 3 (20th February 2013)
- Main Title:
- Development and clinical evaluation of automatic fiducial detection for tumor tracking in cine megavoltage images during volumetric modulated arc therapy
- Authors:
- Azcona, Juan Diego
Li, Ruijiang
Mok, Edward
Hancock, Steven
Xing, Lei - Abstract:
- Abstract : Purpose: : Real‐time tracking of implanted fiducials in cine megavoltage (MV) imaging during volumetric modulated arc therapy (VMAT) delivery is complicated due to the inherent low contrast of MV images and potential blockage of dynamic leaves configurations. The purpose of this work is to develop a clinically practical autodetection algorithm for motion management during VMAT. Methods: : The expected field‐specific segments and the planned fiducial position from the Eclipse (Varian Medical Systems, Palo Alto, CA) treatment planning system were projected onto the MV images. The fiducials were enhanced by applying a Laplacian of Gaussian filter in the spatial domain for each image, with a blob‐shaped object as the impulse response. The search of implanted fiducials was then performed on a region of interest centered on the projection of the fiducial when it was within an open field including the case when it was close to the field edge or partially occluded by the leaves. A universal template formula was proposed for template matching and normalized cross correlation was employed for its simplicity and computational efficiency. The search region for every image was adaptively updated through a prediction model that employed the 3D position of the fiducial estimated from the localized positions in previous images. This prediction model allowed the actual fiducial position to be tracked dynamically and was used to initialize the search region. The artifacts caused byAbstract : Purpose: : Real‐time tracking of implanted fiducials in cine megavoltage (MV) imaging during volumetric modulated arc therapy (VMAT) delivery is complicated due to the inherent low contrast of MV images and potential blockage of dynamic leaves configurations. The purpose of this work is to develop a clinically practical autodetection algorithm for motion management during VMAT. Methods: : The expected field‐specific segments and the planned fiducial position from the Eclipse (Varian Medical Systems, Palo Alto, CA) treatment planning system were projected onto the MV images. The fiducials were enhanced by applying a Laplacian of Gaussian filter in the spatial domain for each image, with a blob‐shaped object as the impulse response. The search of implanted fiducials was then performed on a region of interest centered on the projection of the fiducial when it was within an open field including the case when it was close to the field edge or partially occluded by the leaves. A universal template formula was proposed for template matching and normalized cross correlation was employed for its simplicity and computational efficiency. The search region for every image was adaptively updated through a prediction model that employed the 3D position of the fiducial estimated from the localized positions in previous images. This prediction model allowed the actual fiducial position to be tracked dynamically and was used to initialize the search region. The artifacts caused by electronic interference during the acquisition were effectively removed. A score map was computed by combining both morphological information and image intensity. The pixel location with the highest score was selected as the detected fiducial position. The sets of cine MV images taken during treatment were analyzed with in‐house developed software written inMATLAB (The Mathworks, Inc., Natick, MA). Five prostate patients were analyzed to assess the algorithm performance by measuring their positioning accuracy during treatment. Results: : The algorithm was able to accurately localize the fiducial position on MV images with success rates of more than 90% per case. The percentage of images in which each fiducial was localized in the studied cases varied between 23% and 65%, with at least one fiducial having been localized between 40% and 95% of the images. This depended mainly on the modulation of the plan and fiducial blockage. The prostate movement in the presented cases varied between 0.8 and 3.5 mm (mean values). The maximum displacement detected among all patients was of 5.7 mm. Conclusions: : An algorithm for automatic detection of fiducial markers in cine MV images has been developed and tested with five clinical cases. Despite the challenges posed by complex beam aperture shapes, fiducial localization close to the field edge, partial occlusion of fiducials, fast leaf and gantry movement, and inherently low MV image quality, good localization results were achieved in patient images. This work provides a technique for enabling real‐time accurate fiducial detection and tumor tracking during VMAT treatments without the use of extra imaging dose. … (more)
- Is Part Of:
- Medical physics. Volume 40:Issue 3(2013)
- Journal:
- Medical physics
- Issue:
- Volume 40:Issue 3(2013)
- Issue Display:
- Volume 40, Issue 3 (2013)
- Year:
- 2013
- Volume:
- 40
- Issue:
- 3
- Issue Sort Value:
- 2013-0040-0003-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2013-02-20
- Subjects:
- Therapeutic applications, including brachytherapy -- Artifacts and distortion -- Edge enhancement -- Computed tomography -- Image enhancement -- Biomedical imaging
computerised tomography -- filtering theory -- image enhancement -- image matching -- image motion analysis -- medical image processing -- object tracking -- radiation therapy -- tumours
image‐guided radiation therapy -- real‐time tumor tracking -- MV imaging -- VMAT -- intrafraction motion
Computerised tomographs -- Radiation therapy -- Digital computing or data processing equipment or methods, specially adapted for specific applications -- Image data processing or generation, in general -- Image enhancement or restoration, e.g. from bit‐mapped to bit‐mapped creating a similar image -- Analysis of motion
Medical imaging -- Cancer -- Image guided radiation therapy -- Intensity modulated radiation therapy -- Medical image quality -- Multileaf collimators -- Image analysis -- Medical X‐ray imaging -- Computed tomography -- Dosimetry
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4791646 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File: