Image quality dependency on system configuration and tube voltage in chest tomosynthesis—A visual grading study using an anthropomorphic chest phantom. Issue 3 (18th February 2015)
- Record Type:
- Journal Article
- Title:
- Image quality dependency on system configuration and tube voltage in chest tomosynthesis—A visual grading study using an anthropomorphic chest phantom. Issue 3 (18th February 2015)
- Main Title:
- Image quality dependency on system configuration and tube voltage in chest tomosynthesis—A visual grading study using an anthropomorphic chest phantom
- Authors:
- Söderman, Christina
Asplund, Sara
Allansdotter Johnsson, Åse
Vikgren, Jenny
Rossi Norrlund, Rauni
Molnar, David
Svalkvist, Angelica
Gunnar Månsson, Lars
Båth, Magnus - Abstract:
- Abstract : Purpose: To investigate the potential benefit of increasing the dose per projection image in chest tomosynthesis, performed at the current standard dose level, by reducing the angular range covered or the projection image density and to evaluate the influence of the tube voltage on the image quality. Methods: An anthropomorphic chest phantom was imaged using nine different projection image configurations and ten different tube voltages with the GE VolumeRAD tomosynthesis system. The resulting image sets were representative of being acquired at the same total effective dose. This was achieved partly by applying a simulated dose reduction to the projection images due to restrictions concerning the tube load settings on the VolumeRAD system. Four observers were included in a visual grading study where the reconstructed tomosynthesis section images were rated according to a set of image quality criteria. Image quality was evaluated relative to the default configuration and default tube voltage on the VolumeRAD system. Results: Overall, the image quality decreased with decreasing projection image density. Regarding angular range covered by the projection images, the image quality increased with decreasing angular range for two of the criteria, whereas for a criterion related to the depth resolution in the section images the reduced angular ranges resulted in inferior image quality as compared to the default configuration. The image quality showed little dependence onAbstract : Purpose: To investigate the potential benefit of increasing the dose per projection image in chest tomosynthesis, performed at the current standard dose level, by reducing the angular range covered or the projection image density and to evaluate the influence of the tube voltage on the image quality. Methods: An anthropomorphic chest phantom was imaged using nine different projection image configurations and ten different tube voltages with the GE VolumeRAD tomosynthesis system. The resulting image sets were representative of being acquired at the same total effective dose. This was achieved partly by applying a simulated dose reduction to the projection images due to restrictions concerning the tube load settings on the VolumeRAD system. Four observers were included in a visual grading study where the reconstructed tomosynthesis section images were rated according to a set of image quality criteria. Image quality was evaluated relative to the default configuration and default tube voltage on the VolumeRAD system. Results: Overall, the image quality decreased with decreasing projection image density. Regarding angular range covered by the projection images, the image quality increased with decreasing angular range for two of the criteria, whereas for a criterion related to the depth resolution in the section images the reduced angular ranges resulted in inferior image quality as compared to the default configuration. The image quality showed little dependence on the tube voltage. Conclusions: At the standard dose level of the VolumeRAD system, the potential benefits from increasing the dose per projection do not fully compensate for the negative effects resulting from a reduction in the number of projection images. Consequently, the default configuration consisting of 60 projection images acquired over 30° is a good alternative. The tube voltage used in tomosynthesis does not have a large impact on the image quality. … (more)
- Is Part Of:
- Medical physics. Volume 42:Issue 3(2015)
- Journal:
- Medical physics
- Issue:
- Volume 42:Issue 3(2015)
- Issue Display:
- Volume 42, Issue 3 (2015)
- Year:
- 2015
- Volume:
- 42
- Issue:
- 3
- Issue Sort Value:
- 2015-0042-0003-0000
- Page Start:
- 1200
- Page End:
- 1212
- Publication Date:
- 2015-02-18
- Subjects:
- diagnostic radiography -- medical image processing -- phantoms
Radiography -- Image quality
Biological material, e.g. blood, urine; Haemocytometers -- Digital computing or data processing equipment or methods, specially adapted for specific applications -- Image data processing or generation, in general
thoracic radiography -- tomosynthesis -- image quality -- optimization -- visual grading
Digital tomosynthesis mammography -- Medical image reconstruction -- Medical image artifacts -- Image sensors -- Radiography -- Image reconstruction -- Medical image contrast
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4907963 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9348.xml