A measurement technique to determine the calibration accuracy of an electromagnetic tracking system to radiation isocenter. Issue 8 (22nd July 2013)
- Record Type:
- Journal Article
- Title:
- A measurement technique to determine the calibration accuracy of an electromagnetic tracking system to radiation isocenter. Issue 8 (22nd July 2013)
- Main Title:
- A measurement technique to determine the calibration accuracy of an electromagnetic tracking system to radiation isocenter
- Authors:
- Litzenberg, Dale W.
Gallagher, Ian
Masi, Kathryn J.
Lee, Choonik
Prisciandaro, Joann I.
Hamstra, Daniel A.
Ritter, Timothy
Lam, Kwok L. - Abstract:
- Abstract : Purpose: : To present and characterize a measurement technique to quantify the calibration accuracy of an electromagnetic tracking system to radiation isocenter. Methods: : This technique was developed as a quality assurance method for electromagnetic tracking systems used in a multi‐institutional clinical hypofractionated prostate study. In this technique, the electromagnetic tracking system is calibrated to isocenter with the manufacturers recommended technique, using laser‐based alignment. A test patient is created with a transponder at isocenter whose position is measured electromagnetically. Four portal images of the transponder are taken with collimator rotations of 45° 135°, 225°, and 315°, at each of four gantry angles (0°, 90°, 180°, 270°) using a 3 × 6 cm 2 radiation field. In each image, the center of the copper‐wrapped iron core of the transponder is determined. All measurements are made relative to this transponder position to remove gantry and imager sag effects. For each of the 16 images, the 50% collimation edges are identified and used to find a ray representing the rotational axis of each collimation edge. The 16 collimator rotation rays from four gantry angles pass through and bound the radiation isocenter volume. The center of the bounded region, relative to the transponder, is calculated and then transformed to tracking system coordinates using the transponder position, allowing the tracking system's calibration offset from radiation isocenterAbstract : Purpose: : To present and characterize a measurement technique to quantify the calibration accuracy of an electromagnetic tracking system to radiation isocenter. Methods: : This technique was developed as a quality assurance method for electromagnetic tracking systems used in a multi‐institutional clinical hypofractionated prostate study. In this technique, the electromagnetic tracking system is calibrated to isocenter with the manufacturers recommended technique, using laser‐based alignment. A test patient is created with a transponder at isocenter whose position is measured electromagnetically. Four portal images of the transponder are taken with collimator rotations of 45° 135°, 225°, and 315°, at each of four gantry angles (0°, 90°, 180°, 270°) using a 3 × 6 cm 2 radiation field. In each image, the center of the copper‐wrapped iron core of the transponder is determined. All measurements are made relative to this transponder position to remove gantry and imager sag effects. For each of the 16 images, the 50% collimation edges are identified and used to find a ray representing the rotational axis of each collimation edge. The 16 collimator rotation rays from four gantry angles pass through and bound the radiation isocenter volume. The center of the bounded region, relative to the transponder, is calculated and then transformed to tracking system coordinates using the transponder position, allowing the tracking system's calibration offset from radiation isocenter to be found. All image analysis and calculations are automated with inhouse software for user‐independent accuracy. Three different tracking systems at two different sites were evaluated for this study. Results: : The magnitude of the calibration offset was always less than the manufacturer's stated accuracy of 0.2 cm using their standard clinical calibration procedure, and ranged from 0.014 to 0.175 cm. On three systems in clinical use, the magnitude of the offset was found to be 0.053 ± 0.036, 0.121 ± 0.023, and 0.093 ± 0.013 cm. Conclusions: : The method presented here provides an independent technique to verify the calibration of an electromagnetic tracking system to radiation isocenter. The calibration accuracy of the system was better than the 0.2 cm accuracy stated by the manufacturer. However, it should not be assumed to be zero, especially for stereotactic radiation therapy treatments where planning target volume margins are very small. … (more)
- Is Part Of:
- Medical physics. Volume 40:Issue 8(2013)
- Journal:
- Medical physics
- Issue:
- Volume 40:Issue 8(2013)
- Issue Display:
- Volume 40, Issue 8 (2013)
- Year:
- 2013
- Volume:
- 40
- Issue:
- 8
- Issue Sort Value:
- 2013-0040-0008-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2013-07-22
- Subjects:
- Therapeutic applications, including brachytherapy -- Standards and calibration
biomedical imaging -- calibration -- cancer -- collimators -- quality assurance -- radiation therapy -- tracking -- transponders
electromagnetic tracking -- Winston‐Lutz -- star shot -- calibration -- radiation isocenter
Radiation therapy -- Testing or calibrating of apparatus or arrangements provided for in groups G01D1/00 to G01D15/00 -- Calibrating of instruments or apparatus -- Responders; Transponders
Calibration -- Medical imaging -- Tracking devices -- Collimation -- Collimators -- Electromagnetic radiation -- Inductors -- Radiation treatment -- Rotation measurement -- Radiosurgery
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4813910 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9346.xml